Metal-backed glenoid implant with polyethylene insert is not a viable long-term therapeutic option

被引:48
作者
Boileau, Pascal [1 ]
Moineau, Gregory [1 ]
Morin-Salvo, Nicolas [1 ]
Avidor, Cyrille [1 ]
Godeneche, Arnaud [2 ]
Levigne, Christophe [3 ]
Baba, Mohamed [1 ]
Walch, Gilles [2 ]
机构
[1] Univ Nice Sophia Antipolis, Hop Archet 2, F-06200 Nice, France
[2] Ctr Med Santy, Lyon, France
[3] Clin Parc, Lyon, France
关键词
Total shoulder arthroplasty; uncemented glenoid; glenoid survival; revision; TOTAL SHOULDER ARTHROPLASTY; PRIMARY GLENOHUMERAL OSTEOARTHRITIS; FOLLOW-UP; COMPONENT; PROSTHESIS; ARTHRITIS; OUTCOMES; DESIGN;
D O I
10.1016/j.jse.2015.02.012
中图分类号
R826.8 [整形外科学]; R782.2 [口腔颌面部整形外科学]; R726.2 [小儿整形外科学]; R62 [整形外科学(修复外科学)];
学科分类号
摘要
Background: The aims of this study were to determine the survival of anatomic total shoulder arthroplasty with uncemented metal-backed (MB) glenoid components with a polyethylene (PE) insert in primary osteoarthritis, to assess the reasons for revision surgery, and to identify patients and diagnostic factors that influence failure rates. Methods: Between 1994 and 1999, 165 patients (mean age, 68 years) with primary osteoarthritis were treated with anatomic total shoulder arthroplasty using an uncemented MB/PE glenoid component. Outcomes were assessed both clinically and radiologically with a minimum of 2 years of follow-up. A prosthetic survival curve was constructed with the end point defined as either partial or complete revision, using 100% confidence intervals. Results: Survival rate free of revision was 46% (100% confidence interval, 32%-54%) at 12 years. At a mean follow-up of 8.5 years (range, 2-16 years), revision was required in 61 patients (37%); 80% of shoulders undergoing revision (49 of 61) had evidence of PE wear. Glenoid loosening (because of osteolysis secondary to wear debris), soft tissue deficiency, and prosthetic instability were the most common modes of failure. Younger patients and biconcave glenoids (with posterior humeral subluxation) have a negative effect on implant survival. Proximal humerus osteolysis was significantly more frequent in shoulders with PE wear. Exchange of the PE insert (with conservation of the MB tray) was possible in only 3% of the revised shoulders. Conclusion: Uncemented MB glenoid resurfacing is not a viable long-term therapeutic option because of accelerated PE wear leading to early revision surgery. Conservation of the MB tray with reinsertion of a new PE insert is rarely possible because of glenoid bone loss, implant loosening, soft tissue deficiency, and prosthetic instability. Younger patients and biconcave glenoids have a negative effect on implant survival. (C) 2015 The Authors. Published by Elsevier Inc. on behalf of Journal of Shoulder and Elbow Surgery Board of Trustees.
引用
收藏
页码:1534 / 1543
页数:10
相关论文
共 31 条
[1]  
[Anonymous], BMC MUSCULOSKELET DI
[2]   The three-dimensional geometry of the proximal humerus - Implications for surgical technique and prosthetic design [J].
Boileau, P ;
Walch, G .
JOURNAL OF BONE AND JOINT SURGERY-BRITISH VOLUME, 1997, 79B (05) :857-865
[3]   Grammont reverse prosthesis: Design, rationale, and biomechanics [J].
Boileau, P ;
Watkinson, DJ ;
Hatzidakis, AM ;
Balg, F .
JOURNAL OF SHOULDER AND ELBOW SURGERY, 2005, 14 (01) :147S-161S
[4]   Cemented polyethylene versus uncemented metal-backed glenoid components in total shoulder arthroplasty:: A prospective, double-blind, randomized study [J].
Boileau, P ;
Avidor, C ;
Krishnan, SG ;
Walch, G ;
Kempf, JF ;
Molé, D .
JOURNAL OF SHOULDER AND ELBOW SURGERY, 2002, 11 (04) :351-359
[5]  
Boileau P, 2001, 2000 PROTHESES EPAUL, P489
[6]   Results of total shoulder arthroplasty with a monoblock porous tantalum glenoid component: a prospective minimum 2-year follow-up study [J].
Budge, Matthew D. ;
Nolan, Elizabeth M. ;
Heisey, Meredith H. ;
Baker, Kevin ;
Wiater, J. Michael .
JOURNAL OF SHOULDER AND ELBOW SURGERY, 2013, 22 (04) :535-541
[7]   Mid-term results of a metal-backed glenoid component in total shoulder replacement [J].
Castagna, A. ;
Randelli, M. ;
Garofalo, R. ;
Maradei, L. ;
Giardella, A. ;
Borroni, M. .
JOURNAL OF BONE AND JOINT SURGERY-BRITISH VOLUME, 2010, 92B (10) :1410-1415
[8]   Polyethylene insert exchange for wear after toad shoulder arthroplasty [J].
Cheung, Emilie V. ;
Sperling, John W. ;
Cofield, Robert H. .
JOURNAL OF SHOULDER AND ELBOW SURGERY, 2007, 16 (05) :574-578
[9]   Effect of glenoid cementation on total shoulder arthroplasty for degenerative arthritis of the shoulder: a review of the New Zealand National Joint Registry [J].
Clitherow, Harry D. S. ;
Frampton, Christopher M. A. ;
Astley, Timothy M. .
JOURNAL OF SHOULDER AND ELBOW SURGERY, 2014, 23 (06) :775-781
[10]  
Cofield R H, 1992, J Shoulder Elbow Surg, V1, P77, DOI 10.1016/S1058-2746(09)80124-9