Effect of the High-Pitch Mode in Dual-Source Computed Tomography on the Accuracy of Three-Dimensional Volumetry of Solid Pulmonary Nodules: A Phantom Study

被引:4
作者
Hwang, Sung Ho [1 ]
Oh, Yu-Whan [1 ]
Ham, Soo-Youn [1 ]
Kang, Eun-Young [2 ]
Lee, Ki Yeol [3 ]
机构
[1] Korea Univ, Anam Hosp, Dept Radiol, Seoul 136705, South Korea
[2] Korea Univ, Guro Hosp, Dept Radiol, Seoul 152703, South Korea
[3] Korea Univ, Ansan Hosp, Dept Radiol, Ansan 425707, South Korea
关键词
Pulmonary nodule; Computed tomography; Volumetric imaging; computer generated 3D; Imaging phantom; ITERATIVE RECONSTRUCTION; AUTOMATED VOLUMETRY; GROWTH-RATE; CT;
D O I
10.3348/kjr.2015.16.3.641
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Objective: To evaluate the influence of high-pitch mode (HPM) in dual-source computed tomography (DSCT) on the accuracy of three-dimensional (3D) volumetry for solid pulmonary nodules. Materials and Methods: A Lung phantom implanted with 45 solid pulmonary nodules (n = 15 for each of 4-mm, 6-mm, and 8-mm in diameter) was scanned twice, first in conventional pitch mode (CPM) and then in HPM using DSCT. The relative percentage volume errors (RPEs) of 3D volumetry were compared between the HPM and CPM. In addition, the intermode volume variability (IVV) of 3D volumetry was calculated. Results: In the measurement of the 6-mm and 8-mm nodules, there was no significant difference in RPE (p > 0.05, respectively) between the CPM and HPM (IVVs of 1.2 +/- 0.9%, and 1.7 +/- 1.5%, respectively). In the measurement of the 4-mm nodules, the mean RPE in the HPM (35.1 +/- 7.4%) was significantly greater (p < 0.01) than that in the CPM (18.4 +/- 5.3%), with an IVV of 13.1 +/- 6.6%. However, the IVVs were in an acceptable range (< 25%), regardless of nodule size. Conclusion: The accuracy of 3D volumetry with HPM for solid pulmonary nodule is comparable to that with CPM. However, the use of HPM may adversely affect the accuracy of 3D volumetry for smaller (< 5 mm in diameter) nodule.
引用
收藏
页码:641 / 647
页数:7
相关论文
共 18 条
[1]   Computed Tomography of the Lung in the High-Pitch Mode Is Breath Holding Still Required? [J].
Baumueller, Stephan ;
Alkadhi, Hatem ;
Stolzmann, Paul ;
Frauenfelder, Thomas ;
Goetti, Robert ;
Schertler, Thomas ;
Plass, Andre ;
Falk, Volkmar ;
Feuchtner, Gudrun ;
Scheffel, Hans ;
Desbiolles, Lotus ;
Leschka, Sebastian .
INVESTIGATIVE RADIOLOGY, 2011, 46 (04) :240-245
[2]   CT Screening and Follow-Up of Lung Nodules: Effects of Tube Current-Time Setting and Nodule Size and Density on Detectability and of Tube Current-Time Setting on Apparent Size [J].
Christe, Andreas ;
Torrente, Jaled Charimo ;
Lin, Margaret ;
Yen, Andrew ;
Hallett, Rich ;
Roychoudhury, Kingshuk ;
Schmitzberger, Florian ;
Vock, Peter ;
Roos, Justus .
AMERICAN JOURNAL OF ROENTGENOLOGY, 2011, 197 (03) :623-630
[3]   Computer-assisted solid lung nodule 3D volumetry on CT: influence of scan mode and iterative reconstruction: a CT phantom study [J].
Coenen, Adriaan ;
Honda, Osamu ;
van der Jagt, Eric J. ;
Tomiyama, Noriyuki .
JAPANESE JOURNAL OF RADIOLOGY, 2013, 31 (10) :677-684
[4]   Accuracy of automated volumetry of pulmonary nodules across different multislice CT scanners [J].
Das, Marco ;
Ley-Zaporozhan, Julia ;
Gietema, H. A. ;
Czech, Andre ;
Muehlenbruch, Georg ;
Mahnken, Andreas H. ;
Katoh, Markus ;
Bakai, Annemarie ;
Salganicoff, Marcos ;
Diederich, Stefan ;
Prokop, Mathias ;
Kauczor, Hans-Ulrich ;
Guenther, Rolf W. ;
Wildberger, Joachim E. .
EUROPEAN RADIOLOGY, 2007, 17 (08) :1979-1984
[5]   Automated volumetry of solid pulmonary nodules in a phantom -: Accuracy across different CT scanner technologies [J].
Das, Marco ;
Muehlenbruch, Georg ;
Katoh, Markus ;
Bakai, Annemarie ;
Salganicoff, Marcos ;
Stanzel, Sven ;
Mahnken, Andreas H. ;
Guenther, Rolf W. ;
Wildberger, Joachim E. .
INVESTIGATIVE RADIOLOGY, 2007, 42 (05) :297-302
[6]   Resolving small pulmonary nodules: CT features [J].
Diederich, S ;
Hansen, J ;
Wormanns, D .
EUROPEAN RADIOLOGY, 2005, 15 (10) :2064-2069
[7]   Application of CT-PSF-based computer-simulated lung nodules for evaluating the accuracy of computer-aided volumetry [J].
Ayumu Funaki ;
Masaki Ohkubo ;
Shinichi Wada ;
Kohei Murao ;
Toru Matsumoto ;
Shinji Niizuma .
Radiological Physics and Technology, 2012, 5 (2) :166-171
[8]   Three-dimensional segmentation and growth-rate estimation of small pulmonary nodules in helical CT images [J].
Kostis, WJ ;
Reeves, AP ;
Yankelevitz, DF ;
Henschke, CI .
IEEE TRANSACTIONS ON MEDICAL IMAGING, 2003, 22 (10) :1259-1274
[9]   High-Pitch Electrocardiogram-Triggered Computed Tomography of the Chest Initial Results [J].
Lell, Michael ;
Hinkmann, Fabian ;
Anders, Katharina ;
Deak, Paul ;
Kalender, Willi A. ;
Uder, Michael ;
Achenbach, Stephan .
INVESTIGATIVE RADIOLOGY, 2009, 44 (11) :728-733
[10]   Managing the small pulmonary nodule discovered by CT [J].
Libby, DM ;
Smith, JP ;
Altorki, NK ;
Pasmantier, MW ;
Yankelevitz, D ;
Henschke, CI .
CHEST, 2004, 125 (04) :1522-1529