Predictability and limitations of non-invasive murine tonometry: Comparison of two devices

被引:32
作者
Filippopoulos, Theodoros
Matsubara, Akihisa
Danias, John
Huang, Wei
Dobberfuhl, Adam
Ren, Lizhen
Mittag, Thomas
Miller, Joan W.
Grosskreutz, Cynthia L.
机构
[1] Harvard Univ, Sch Med, Massachusetts Eye & Ear Infirm, Howe Lab,Glaucoma Serv, Boston, MA 02114 USA
[2] Harvard Univ, Sch Med, Massachusetts Eye & Ear Infirm, Angiogenesis Lab,Retina Serv, Boston, MA 02114 USA
[3] Mt Sinai Sch Med, Dept Ophthalmol, New York, NY USA
关键词
glaucoma; mouse; non-invasive; tonometry; intraocular pressure (IOP); model; DBA/2J;
D O I
10.1016/j.exer.2005.12.005
中图分类号
R77 [眼科学];
学科分类号
100212 ;
摘要
Our purpose was to evaluate the accuracy; reproducibility and predictive ability of two non-invasive tonometers developed for intraocular pressure (IOP) measurements in the mouse. The prototype impact-rebound tonometer (I-R) and a prototype optical interferometry tonometer (OIT) utilizing a fiberoptic pressure sensor, were compared. Enucleated eyes from C57/BL6 mice were used for the calibration. The anterior chamber was cannulated and the IOP was adjusted in increments of 5 cm of H2O (open stopcock method). A calibration curve was generated for each individual eye along with a master calibration curve for all eyes. Two operators measured the IOP. The instruments were then used in alternating order to measure the IOP in C57/BL6 and in DBA2/J animals. The same eyes were subsequently cannulated and the error of the non-invasive tonometers was determined. Both tonometers yielded almost equivalent ex vivo calibration curves with individual R-2 of 0.9878 and 0.9902 respectively. Both instruments were highly reproducible. In vivo the I-R tonometer underestimated while the OIT overestimated the IOP. This error was systematic and therefore predictable. The confidence intervals of this error were determined by comparing the IOP estimates provided by each tonometer with the measurements obtained invasively by cannulation in vivo. The 95% CI of the error were 2.36 mmHg for the I-R and 2.62 mmHg for the OIT respectively. Non-invasive tonometry in the mouse is feasible. Both non-invasive instruments provide accurate and reproducible measurements with the OIT requiring calibration curves for each individual. investigator. (c) 2006 Elsevier Ltd. All rights reserved.
引用
收藏
页码:194 / 201
页数:8
相关论文
共 28 条
[1]  
AHMED E, 2003, INVEST OPHTH VIS SCI, V44, P3336
[2]   Experimental mouse ocular hypertension: Establishment of the model [J].
Aihara, M ;
Lindsey, JD ;
Weinreb, RN .
INVESTIGATIVE OPHTHALMOLOGY & VISUAL SCIENCE, 2003, 44 (10) :4314-4320
[3]   Ocular hypertension in mice with a targeted type I collagen mutation [J].
Aihara, M ;
Lindsey, JD ;
Weinreb, RN .
INVESTIGATIVE OPHTHALMOLOGY & VISUAL SCIENCE, 2003, 44 (04) :1581-1585
[4]  
Aihara M, 2002, INVEST OPHTH VIS SCI, V43, P146
[5]  
Avila MY, 2001, INVEST OPHTH VIS SCI, V42, P1841
[6]   Measuring agreement in method comparison studies [J].
Bland, JM ;
Altman, DG .
STATISTICAL METHODS IN MEDICAL RESEARCH, 1999, 8 (02) :135-160
[7]  
Cohan BE, 2001, INVEST OPHTH VIS SCI, V42, P340
[8]  
Cohan BE, 2001, INVEST OPHTH VIS SCI, V42, P2560
[9]   Quantitative analysis of retinal ganglion cell (RGC) loss in aging DBA/2NNia glaucomatous mice: Comparison with RGC loss in aging C57/BL6 mice [J].
Danias, J ;
Lee, KC ;
Zamora, MF ;
Chen, B ;
Shen, F ;
Filippopoulos, T ;
Su, YL ;
Goldblum, D ;
Podos, SM ;
Mittag, T .
INVESTIGATIVE OPHTHALMOLOGY & VISUAL SCIENCE, 2003, 44 (12) :5151-5162
[10]   Method for the noninvasive measurement of intraocular pressure in mice [J].
Danias, J ;
Kontiola, AI ;
Filippopoulos, T ;
Mittag, T .
INVESTIGATIVE OPHTHALMOLOGY & VISUAL SCIENCE, 2003, 44 (03) :1138-1141