Time Course of Current of Injury Is Related to Acute Stability of Active-Fixation Pacing Leads in Rabbits

被引:6
作者
Shali, Shalaimaiti [1 ]
Wushou, Alimujiang [2 ]
Liu, Entao [3 ]
Jia, Lin [3 ]
Yao, Ruiming [1 ]
Su, Yangang [1 ]
Ge, Junbo [1 ]
机构
[1] Fudan Univ, Zhong Shan Hosp, Shanghai Inst Cardiovasc Dis, Dept Cardiol, Shanghai 200433, Peoples R China
[2] Xi An Jiao Tong Univ, Coll & Hosp Stomatol, Dept Oral & Maxillofacial Head & Neck Surg Oncol, Xian, Peoples R China
[3] Medtron Shanghai Co Ltd, Dept Cardiac Rhythm Dis Management, Shanghai, Peoples R China
关键词
PACEMAKER; PERFORMANCE; ELECTRODES; UNIPOLAR;
D O I
10.1371/journal.pone.0057727
中图分类号
O [数理科学和化学]; P [天文学、地球科学]; Q [生物科学]; N [自然科学总论];
学科分类号
07 ; 0710 ; 09 ;
摘要
Background: Magnitude of current of injury (COI) consequent to pacemaker lead fixation is recognized as a predictor of acute lead stability. It is unclear whether dynamic monitoring of COI after lead fixation provides additional information beyond a single assessment performed at the time of fixation. Objectives: This study was aimed to test the hypothesis that the time course of COI is related to acute lead stability. Methods and Results: Active fixation leads with fixed screw were anchored to either Langendorff-perfused rabbit hearts endocardially or in vivo hearts epicardially in manners of contact the helix with no rotation, half rotation and full rotation, respectively. Intracardiac electrogram (EGM) was monitored dynamically from onset to resolution of COI, and magnitudes of intrinsic R wave and COI, including ST-segment elevation, ST/R and intracardiac EGM duration (IED), were measured. A digital force gauge was applied to assess lead stability. In vitro, COI in contacted leads was significantly smaller than those in half rotated (p < 0.05) and fully rotated leads (p < 0.05), and presented most precipitous recovery to baseline (1.5+/-1.1 min, p < 0.05). Half-rotated and fully rotated leads manifested the same magnitude of COI right after placement. However, the time course of COI was significantly longer in fully rotated leads than that in half rotated leads (26.5+/-2.8 min vs. 5.6+/-2.0 min, p < 0.05). Similar findings were observed in vivo. The time course of COI was significantly correlated with the force needed to detach the lead from myocardium (r = 0. 72, n = 48, p < 0.001). Conclusions: Time course of COI is related to acute lead stability in rabbits. One might be misled by a single assessment of COI magnitude right after lead placement, whereas persistence of COI is likely to be a useful indicator of adequate lead stability.
引用
收藏
页数:7
相关论文
共 15 条
[1]   Selective site right ventricular pacing [J].
Albouaini, K. ;
Alkarmi, A. ;
Mudawi, T. ;
Gammage, M. D. ;
Wright, D. J. .
HEART, 2009, 95 (24) :2030-2039
[2]  
Avramovitch N, 2004, HEART RHYTHM, V71, P219
[3]   EARLY COMPLICATIONS AFTER DUAL-CHAMBER VERSUS SINGLE CHAMBER PACEMAKER IMPLANTATION [J].
CHAUHAN, A ;
GRACE, AA ;
NEWELL, SA ;
STONE, DL ;
SHAPIRO, LM ;
SCHOFIELD, PM ;
PETCH, MC .
PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 1994, 17 (11) :2012-2015
[4]   Acute Lead Dislodgements and In-Hospital Mortality in Patients Enrolled in the National Cardiovascular Data Registry Implantable Cardioverter Defibrillator Registry [J].
Cheng, Alan ;
Wang, Yongfei ;
Curtis, Jeptha P. ;
Varosy, Paul D. .
JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 2010, 56 (20) :1651-1656
[5]   ACUTE CHANGES IN PACING THRESHOLD AND R-WAVE OR P-WAVE AMPLITUDE DURING PERMANENT PACEMAKER IMPLANTATION [J].
DEBUITLEIR, M ;
KOU, WH ;
SCHMALTZ, S ;
MORADY, F .
AMERICAN JOURNAL OF CARDIOLOGY, 1990, 65 (15) :999-1003
[6]   COMPARISON OF UNIPOLAR AND BIPOLAR ELECTROGRAMS FOR CARDIAC-PACEMAKER SENSING [J].
DECAPRIO, V ;
HURZELER, P ;
FURMAN, S .
CIRCULATION, 1977, 56 (05) :750-755
[7]   Transmembrane voltage changes produced by real and virtual electrodes during monophasic defibrillation shock delivered by an implantable electrode [J].
Efimov, IR ;
Cheng, YN ;
Biermann, M ;
VanWagoner, DR ;
Mazgalev, TN ;
Tchou, PJ .
JOURNAL OF CARDIOVASCULAR ELECTROPHYSIOLOGY, 1997, 8 (09) :1031-1045
[8]   DIRECT-CURRENT INJURY FROM EXTERNAL PACEMAKER RESULTS IN TISSUE ELECTROLYSIS [J].
GROSSI, EA ;
PARISH, MA ;
KRALIK, MR ;
GLASSMAN, LR ;
ESPOSITO, RA ;
RIBAKOVE, GH ;
GALLOWAY, AC ;
COLVIN, SB .
ANNALS OF THORACIC SURGERY, 1993, 56 (01) :156-157
[9]   CHARACTERISTICS OF INTRA-CARDIAC ELECTROGRAMS [J].
MYERS, GH ;
KRESH, YM ;
PARSONNET, V .
PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 1978, 1 (01) :90-103
[10]   EARLY MALFUNCTION OF TRANSVENOUS PACEMAKER ELECTRODES - 3-CENTER STUDY [J].
PARSONNET, V ;
BILITCH, M ;
FURMAN, S ;
FISHER, JD ;
ESCHER, DJW ;
MYERS, G ;
CASSADY, E .
CIRCULATION, 1979, 60 (03) :590-596