Development of a surgical navigation system based on augmented reality using an optical see-through head-mounted display

被引:153
作者
Chen, Xiaojun [1 ]
Xu, Lu [1 ]
Wang, Yiping [1 ]
Wang, Huixiang [2 ]
Wang, Fang [2 ]
Zeng, Xiangsen [2 ]
Wang, Qiugen [2 ]
Egger, Jan [3 ]
机构
[1] Shanghai Jiao Tong Univ, Inst Biomed Mfg & Life Qual Engn, State Key Lab Mech Syst & Vibrat, Sch Mech Engn, Shanghai 200240, Peoples R China
[2] Shanghai Jiao Tong Univ, Shanghai Peoples Hosp 1, Sch Med, Shanghai 200240, Peoples R China
[3] Graz Univ Technol, Fac Comp Sci & Biomed Engn, Inst Comp Graph & Vis, A-8010 Graz, Austria
关键词
Surgical navigation; Augmented reality; Optical see-through HMD; Intra-operative motion tracking; COMPUTER-AIDED NAVIGATION; ORAL IMPLANTOLOGY; ACCURACY; CALIBRATION; PLACEMENT; ZYGOMA; MAXILLECTOMY; REGISTRATION; TECHNOLOGY;
D O I
10.1016/j.jbi.2015.04.003
中图分类号
TP39 [计算机的应用];
学科分类号
081203 ; 0835 ;
摘要
The surgical navigation system has experienced tremendous development over the past decades for minimizing the risks and improving the precision of the surgery. Nowadays, Augmented Reality (AR)-based surgical navigation is a promising technology for clinical applications. In the AR system, virtual and actual reality are mixed, offering real-time, high-quality visualization of an extensive variety of information to the users (Moussa et al., 2012) [1]. For example, virtual anatomical structures such as soft tissues, blood vessels and nerves can be integrated with the real-world scenario in real time. In this study, an AR-based surgical navigation system (AR-SNS) is developed using an optical see-through HMD (head-mounted display), aiming at improving the safety and reliability of the surgery. With the use of this system, including the calibration of instruments, registration, and the calibration of HMD, the 3D virtual critical anatomical structures in the head-mounted display are aligned with the actual structures of patient in real-world scenario during the intra-operative motion tracking process. The accuracy verification experiment demonstrated that the mean distance and angular errors were respectively 0.809 +/- 0.05 mm and 1.038 degrees +/- 0.05 degrees, which was sufficient to meet the clinical requirements. (C) 2015 Elsevier Inc. All rights reserved.
引用
收藏
页码:124 / 131
页数:8
相关论文
共 28 条
[1]   Recent advances in augmented reality [J].
Azuma, R ;
Baillot, Y ;
Behringer, R ;
Feiner, S ;
Julier, S ;
MacIntyre, B .
IEEE COMPUTER GRAPHICS AND APPLICATIONS, 2001, 21 (06) :34-47
[2]  
Caversaccio M, 2000, AM J OTOL, V21, P863
[3]   Development of an open source software module for enhanced visualization during MR-guided interstitial gynecologic brachytherapy [J].
Chen, Xiaojun ;
Egger, Jan .
SPRINGERPLUS, 2014, 3 :1-9
[4]   Real-time motion tracking in image-guided oral implantology [J].
Chen, Xiaojun ;
Lin, Yanping ;
Wu, Yiqun ;
Wang, Chengtao .
INTERNATIONAL JOURNAL OF MEDICAL ROBOTICS AND COMPUTER ASSISTED SURGERY, 2008, 4 (04) :339-347
[5]   Image guided oral implantology and its application in the placement of zygoma implants [J].
Chen Xiaojun ;
Ye Ming ;
Lin Yanping ;
Wu Yiqun ;
Wang Chengtao .
COMPUTER METHODS AND PROGRAMS IN BIOMEDICINE, 2009, 93 (02) :162-173
[6]   Accuracy of computer navigation in ear, nose, throat surgery - The influence of matching strategy [J].
Claes, J ;
Koekelkoren, E ;
Wuyts, FL ;
Claes, GME ;
Van den Hauwe, L ;
Van de Heyning, PH .
ARCHIVES OF OTOLARYNGOLOGY-HEAD & NECK SURGERY, 2000, 126 (12) :1462-1466
[7]  
Cleary K., 2009, IGSTK THE BOOK, P225
[8]   Image-Guided Interventions: Technology Review and Clinical Applications [J].
Cleary, Kevin ;
Peters, Terry M. .
ANNUAL REVIEW OF BIOMEDICAL ENGINEERING, VOL 12, 2010, 12 :119-142
[9]   Basic research and 12 years of clinical experience in computer-assisted navigation technology: a review [J].
Ewers, R ;
Schicho, K ;
Undt, G ;
Wanschitz, F ;
Truppe, M ;
Seemann, R ;
Wagner, A .
INTERNATIONAL JOURNAL OF ORAL AND MAXILLOFACIAL SURGERY, 2005, 34 (01) :1-8
[10]   Predicting error in rigid-body point-based registration [J].
Fitzpatrick, JM ;
West, JB ;
Maurer, CR .
IEEE TRANSACTIONS ON MEDICAL IMAGING, 1998, 17 (05) :694-702