Dosimetric evaluation of a dedicated stereotactic linear accelerator using measurement and Monte Carlo simulation

被引:12
|
作者
Heydarian, M. [1 ,2 ]
Asnaashari, K. [3 ]
Allahverdi, M. [3 ]
Jaffray, D. A. [1 ,2 ,4 ,5 ]
机构
[1] Univ Toronto, Princess Margaret Hosp, Radiat Med Program, Toronto, ON M5G 2M9, Canada
[2] Univ Toronto, Dept Radiat Oncol, Toronto, ON M5G 2M9, Canada
[3] Univ Tehran Med Sci, Dept Med Phys, Tehran, Iran
[4] Univ Toronto, Princess Margaret Hosp, Ontario Canc Inst, Toronto, ON M5G 2M9, Canada
[5] Univ Toronto, Dept Med Biophys, Toronto, ON M5G 2M9, Canada
关键词
dosimetric evaluation; stereotactic radiosurgery; Monte Carlo; IMRT;
D O I
10.1118/1.2964198
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Dosimetric parameters of a dedicated stereotactic linear accelerator have been investigated using measurements and Monte Carlo simulations. This linac has a unique built in multileaf collimation (MLC) system with the maximum opening of 16x21 cm(2) and 4 mm leaf width at the isocenter and has successfully been modeled for the first time using the Monte Carlo simulation. The high resolution MLC, combined with its relatively large maximum field size, opens up a new opportunity for expanding stereotactic radiation treatment techniques from traditionally treating smaller targets to larger ones for both cranial and extracranial lesions. Dosimetric parameters of this linac such as accuracy of leaf positioning and field shaping, leakage and transmission, percentage depth doses, off-axes dose profiles, and dose penumbras were measured and calculated for different field sizes, depths, and source to surface distances. In addition, the ability of the linac in accurate dose delivery of several treatment plans, including intensity modulated radiation therapy (IMRT), performed on phantom and patients was determined. Ionization chamber, photon diode detector, films, several solid water phantoms, and a water tank were used for the measurements. The MLC leaf positioning to any particular point in the maximum aperture was accurate with a standard deviation of 0.29 mm. Maximum and average leakages were 1.7% and 1.1% for the reference field of 10.4x9.6 cm(2). Measured penumbra widths (80%-20%) for this field at source axis distance (SAD) of 100 cm at a depth of 1.5 cm (d(max)) were 3.2 and 4 mm for the leaf-sides and leaf-ends, respectively. The corresponding results at 10 cm depth and SAD=100 cm were 5.4 and 6.3 mm. Monte Carlo results generally agreed with the measurements to within 1% and or 1 mm, with respective uncertainties of 0.5% and 0.2 mm. The linac accuracy in delivering non-IMRT treatment plans was better than 1%. Ionization chamber dosimetry results for a phantom IMRT plan in the high dose and low dose regions were -0.5% and +3.6%, respectively. Dosimetry results at isocenter for three patients' IMRT plans were measured to be within 3% of their corresponding treatment plans. Film dosimetry was also used to compare dose distributions of IMRT treatment plans and delivered cumulative doses at different cross sectional planes. (C) 2008 American Association of Physicists in Medicine.
引用
收藏
页码:3943 / 3954
页数:12
相关论文
共 50 条
  • [21] Radiation dose distribution around a mobile linear accelerator Mobetron using Monte Carlo simulation
    Garcia-Cases, F.
    Patino, G.
    Perez-Calatayud, J.
    Ballester, F.
    Granero, D.
    Vijande, J.
    2016 GLOBAL MEDICAL ENGINEERING PHYSICS EXCHANGES/PAN AMERICAN HEALTH CARE EXCHANGES (GMEPE/PAHCE), 2016,
  • [22] GATE Monte Carlo Simulation of a 6 MV SIEMENS OncorTM Linear Accelerator Photon Beam : Feasibility and Preliminary Dosimetric Study
    Benhalouche, S.
    Edel, S.
    Visvikis, D.
    Pradier, O.
    Boussion, N.
    MEDICAL PHYSICS, 2011, 38 (06)
  • [23] Monte Carlo simulation of electron beams generated by a 12 MeV dedicated mobile IORT accelerator
    Iaccarino, G.
    Strigari, L.
    D'Andrea, M.
    Bellesi, L.
    Felici, G.
    Ciccotelli, A.
    Benassi, M.
    Soriani, A.
    PHYSICS IN MEDICINE AND BIOLOGY, 2011, 56 (14): : 4579 - 4596
  • [24] A Monte Carlo based modelling of a dedicated mobile IOERT accelerator
    Besheli, M. Ghorbanpour
    Fielitz, O.
    Budach, W.
    Simiantonakis, I.
    RADIOTHERAPY AND ONCOLOGY, 2016, 119 : S727 - S727
  • [25] Determining optimization of the initial parameters in Monte Carlo simulation for linear accelerator radiotherapy
    Chang, Kwo-Ping
    Wang, Zhi-Wei
    Shiau, An-Cheng
    RADIATION PHYSICS AND CHEMISTRY, 2014, 95 : 161 - 165
  • [26] Benchmarking of electron beam parameters based on Monte Carlo linear accelerator simulation
    Zhang, Fan
    Zhou, Mi
    Liu, Jing
    Yue, Lu
    Deng, Lihua
    Xu, Zhijian
    Wang, Gang
    TRANSLATIONAL CANCER RESEARCH, 2020, 9 (02) : 577 - 584
  • [27] A dosimetric survey of different mammographic spectra using Monte Carlo simulation
    Delis, H
    Spyrou, G
    Tzanakos, G
    Panayiotakis, G
    MEDICON 2001: PROCEEDINGS OF THE INTERNATIONAL FEDERATION FOR MEDICAL & BIOLOGICAL ENGINEERING, PTS 1 AND 2, 2001, : 799 - 802
  • [28] Modeling 6 MeV Electron Beam From Medical Linear Accelerator Using Monte Carlo Simulation
    Dechsupa, P.
    Tannanonta, C.
    Phaisangittisakul, N.
    MEDICAL PHYSICS, 2009, 36 (06)
  • [29] MEASUREMENT UNCERTAINTY EVALUATION USING MONTE CARLO SIMULATION: APPLICATIONS WITH LOGARITHMIC FUNCTIONS
    Sousa, Joao A.
    Ribeiro, Alvaro S.
    Costa, Carlos O.
    Castro, Manuel P.
    ADVANCED MATHEMATICAL AND COMPUTATIONAL TOOLS IN METROLOGY VII, 2006, 72 : 335 - +
  • [30] Evaluation of measurement uncertainty based on the propagation of distributions using Monte Carlo simulation
    Cox, M
    Harris, P
    Siebert, BPL
    MEASUREMENT TECHNIQUES, 2003, 46 (09) : 824 - 833