Dosimetric evaluation of a dedicated stereotactic linear accelerator using measurement and Monte Carlo simulation

被引:12
|
作者
Heydarian, M. [1 ,2 ]
Asnaashari, K. [3 ]
Allahverdi, M. [3 ]
Jaffray, D. A. [1 ,2 ,4 ,5 ]
机构
[1] Univ Toronto, Princess Margaret Hosp, Radiat Med Program, Toronto, ON M5G 2M9, Canada
[2] Univ Toronto, Dept Radiat Oncol, Toronto, ON M5G 2M9, Canada
[3] Univ Tehran Med Sci, Dept Med Phys, Tehran, Iran
[4] Univ Toronto, Princess Margaret Hosp, Ontario Canc Inst, Toronto, ON M5G 2M9, Canada
[5] Univ Toronto, Dept Med Biophys, Toronto, ON M5G 2M9, Canada
关键词
dosimetric evaluation; stereotactic radiosurgery; Monte Carlo; IMRT;
D O I
10.1118/1.2964198
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Dosimetric parameters of a dedicated stereotactic linear accelerator have been investigated using measurements and Monte Carlo simulations. This linac has a unique built in multileaf collimation (MLC) system with the maximum opening of 16x21 cm(2) and 4 mm leaf width at the isocenter and has successfully been modeled for the first time using the Monte Carlo simulation. The high resolution MLC, combined with its relatively large maximum field size, opens up a new opportunity for expanding stereotactic radiation treatment techniques from traditionally treating smaller targets to larger ones for both cranial and extracranial lesions. Dosimetric parameters of this linac such as accuracy of leaf positioning and field shaping, leakage and transmission, percentage depth doses, off-axes dose profiles, and dose penumbras were measured and calculated for different field sizes, depths, and source to surface distances. In addition, the ability of the linac in accurate dose delivery of several treatment plans, including intensity modulated radiation therapy (IMRT), performed on phantom and patients was determined. Ionization chamber, photon diode detector, films, several solid water phantoms, and a water tank were used for the measurements. The MLC leaf positioning to any particular point in the maximum aperture was accurate with a standard deviation of 0.29 mm. Maximum and average leakages were 1.7% and 1.1% for the reference field of 10.4x9.6 cm(2). Measured penumbra widths (80%-20%) for this field at source axis distance (SAD) of 100 cm at a depth of 1.5 cm (d(max)) were 3.2 and 4 mm for the leaf-sides and leaf-ends, respectively. The corresponding results at 10 cm depth and SAD=100 cm were 5.4 and 6.3 mm. Monte Carlo results generally agreed with the measurements to within 1% and or 1 mm, with respective uncertainties of 0.5% and 0.2 mm. The linac accuracy in delivering non-IMRT treatment plans was better than 1%. Ionization chamber dosimetry results for a phantom IMRT plan in the high dose and low dose regions were -0.5% and +3.6%, respectively. Dosimetry results at isocenter for three patients' IMRT plans were measured to be within 3% of their corresponding treatment plans. Film dosimetry was also used to compare dose distributions of IMRT treatment plans and delivered cumulative doses at different cross sectional planes. (C) 2008 American Association of Physicists in Medicine.
引用
收藏
页码:3943 / 3954
页数:12
相关论文
共 50 条
  • [1] Focal spot estimation of an Elekta dedicated stereotactic linear accelerator Monte Carlo model
    Herwiningsih, S.
    Fielding, A.
    13TH SOUTH-EAST ASIAN CONGRESS OF MEDICAL PHYSICS 2015 (SEACOMP), 2016, 694
  • [2] Erratum: Activation evaluation of Siemens linear accelerator using Monte Carlo simulation
    Young-Jae Jang
    Na Hye Kwon
    Seong Hee Park
    Yona Choi
    Hyungjoon Yu
    Kum Bae Kim
    Dong Wook Kim
    Sang Hyoun Choi
    Journal of the Korean Physical Society, 2022, 81 : 1188 - 1188
  • [3] Monte Carlo simulation of a clinical linear accelerator
    Lin, SY
    Chu, TC
    Lin, JP
    APPLIED RADIATION AND ISOTOPES, 2001, 55 (06) : 759 - 765
  • [4] Basic dosimetric comparison of medical linear accelerator using FlLUKA Monte Carlo model
    Kummali, A.
    Thekkadath, S.
    Mm, M.
    Raman, G. R.
    RADIOTHERAPY AND ONCOLOGY, 2014, 111 : S137 - S138
  • [5] Monte Carlo simulation of stereotactic microbeam radiation therapy: evaluation of the usage of a linear accelerator as the x-ray source
    Gokeri, Gurdal
    Kocar, Cemil
    Tombakoglu, Mehmet
    Cecen, Yigit
    PHYSICS IN MEDICINE AND BIOLOGY, 2013, 58 (13): : 4621 - 4642
  • [6] Monte Carlo Simulation of Linear Accelerator for Dosimetry Analysis
    Francis, Seenia
    Suresh, Darshana
    Nath, Shalini
    Lakshmi, Saai D. R.
    Jayaraj, P. B.
    Puzhakkal, Niyas
    Pournami, P. N.
    2021 6TH INTERNATIONAL CONFERENCE FOR CONVERGENCE IN TECHNOLOGY (I2CT), 2021,
  • [7] Monte Carlo model of the Novalis Classic 6MV stereotactic linear accelerator using the GATE simulation platform
    Wiebe, J.
    Ploquin, N.
    MEDICAL PHYSICS, 2014, 41 (08) : 16 - 16
  • [8] Dosimetric Deviation of Misaligned Beams for a 6 MV Photon Linear Accelerator Using Monte Carlo Simulations
    Kim, S.
    MEDICAL PHYSICS, 2015, 42 (06) : 3405 - 3405
  • [9] Monte Carlo simulation of a medical linear accelerator for radiotherapy use
    Serrano, B.
    Hachem, A.
    Franchisseur, E.
    Herault, J.
    Marcie, S.
    Costa, A.
    Bensadoun, R. J.
    Barthe, J.
    Gerard, J. P.
    RADIATION PROTECTION DOSIMETRY, 2006, 119 (1-4) : 506 - 509
  • [10] Monte Carlo simulation of lung shields for breast IORT with a dedicated mobile accelerator
    Ayala, R.
    Sendon, J. R.
    Garcia, R.
    Gallego, P.
    Polo, R.
    San Miguel, F.
    RADIOTHERAPY AND ONCOLOGY, 2014, 111 : S147 - S148