Effect of the Configurations of Fiducial Markers on the Accuracy of Surgical Navigation in Zygomatic Implant Placement: An In Vitro Study

被引:33
|
作者
Fan, Shengchi [1 ,2 ,3 ]
Hung, Kuofeng [1 ,2 ,3 ,4 ]
Bornstein, Michael M. [4 ]
Huang, Wei [2 ,3 ,5 ]
Wang, Feng [2 ,3 ,5 ]
Wu, Yiqun [1 ,2 ,3 ]
机构
[1] Shanghai Jiao Tong Univ, Shanghai Peoples Hosp 9, Natl Clin Res Ctr Oral Dis, Dent Clin 2,Dept Oral Implantol,Coll Stomatol,Sch, Shanghai, Peoples R China
[2] Shanghai Key Lab Stomatol, Shanghai, Peoples R China
[3] Shanghai Res Inst Stomatol, Shanghai, Peoples R China
[4] Univ Hong Kong, Fac Dent, Oral & Maxillofacial Radiol Appl Oral Sci, Hong Kong, Peoples R China
[5] Shanghai Jiao Tong Univ, Shanghai Peoples Hosp 9, Natl Clin Res Ctr Oral Dis, Dept Oral Implantol,Coll Stomatol,Sch Med, Shanghai, Peoples R China
基金
国家重点研发计划;
关键词
accuracy; edentulous; navigation system; registration; zygomatic implant; TARGET REGISTRATION; SURGERY; ERROR; RECONSTRUCTION; MANAGEMENT; MODALITIES;
D O I
10.11607/jomi.6821
中图分类号
R78 [口腔科学];
学科分类号
1003 ;
摘要
Purpose: Real-time surgical navigation has been increasingly applied in implant placement. The initial registration procedures were found to substantially affect the overall accuracy, but the number and distribution of fiducial markers are yet undetermined. This study aimed to determine the minimal number and optimal distribution of fiducial markers to achieve clinically acceptable accuracy in surgical navigation for zygomatic implant placement by systematically analyzing the effects of different setups of fiducial markers on target registration error (TRE). Materials and Methods: A maxillary phantom with bone-anchored fiducial markers was scanned using cone beam computed tomography, followed by data processing on Brainlab, a commercially available navigation system. A total of 10 miniscrews were inserted in the edentulous maxilla for the configuration of the fiducial markers, with another two miniscrews as implant targets to assess the TRE in zygomatic bone. Data were then collected in nine configurations with distinct fiducial numbers and positions. Statistical analyses were performed with SPSS. Results: The accuracy of the surgical navigation system was found to depend on both the number and the position of fiducial markers. No significant difference was observed in accuracy among groups with eight fiducials and with a polygon span distribution (P > .05). When the fiducial number decreased to less than six, the markers inserted in a regular triangle were more precise than those in an inverse triangle configuration. When the number of fiducials was five with a polygonal distribution, a low TRE value of 0.59 mm was detected, which was comparable to the accuracy with more than eight fiducials in this study. Conclusion: A scattered distribution with a polygon span with at least five fiducial markers in the edentulous maxilla for registration seems to achieve acceptable TRE values with high accuracy for navigation in zygomatic implant placement.
引用
收藏
页码:85 / 90
页数:6
相关论文
共 50 条
  • [41] Impact of Surgical Template on the Accuracy of Implant Placement
    Xu, Liang-wei
    You, Jia
    Zhang, Jian-xing
    Liu, Yun-feng
    Peng, Wei
    JOURNAL OF PROSTHODONTICS-IMPLANT ESTHETIC AND RECONSTRUCTIVE DENTISTRY, 2016, 25 (08): : 641 - 646
  • [42] Accuracy of Implant Placement with a Stereolithographic Surgical Template
    Cassetta, Michele
    Stefanelli, Luigi V.
    Giansanti, Matteo
    Calasso, Sabrina
    INTERNATIONAL JOURNAL OF ORAL & MAXILLOFACIAL IMPLANTS, 2012, 27 (03) : 655 - 663
  • [43] In Vitro Comparison of Surgical Implant Placement Accuracy Using Guides Fabricated by Three Different Additive Technologies
    Anunmana, Chuchai
    Ueawitthayasuporn, Chananchida
    Kiattavorncharoen, Sirichai
    Thanasrisuebwong, Prakan
    APPLIED SCIENCES-BASEL, 2020, 10 (21): : 1 - 11
  • [44] Comparison of the accuracy of dental implant placement using dynamic and augmented reality-based dynamic navigation: An in vitro study
    Tao, Baoxin
    Fan, Xingqi
    Wang, Feng
    Chen, Xiaojun
    Shen, Yihan
    Wu, Yiqun
    JOURNAL OF DENTAL SCIENCES, 2024, 19 (01) : 196 - 202
  • [45] Accuracy of partially and fully guided surgical techniques for immediate implant placement: An in vitro assessment
    Nicchio, Nicolas
    Goncalves, Victor
    Mendonca, Gustavo
    Sales e Pessoa, Roberto
    Frizzera, Fausto
    Zandim-Barcelos, Daniela Leal
    JOURNAL OF PROSTHETIC DENTISTRY, 2023, 129 (02): : 363.e1 - 363.e7
  • [46] Accuracy of a flapless protocol for computer-guided zygomatic implant placement in human cadavers: expectations and reality
    Schiroli, Guido
    Angiero, Francesca
    Zangerl, Antoniettel
    Benedicenti, Stefano
    Ferrante, Franco
    Widmann, Gerlig
    INTERNATIONAL JOURNAL OF MEDICAL ROBOTICS AND COMPUTER ASSISTED SURGERY, 2016, 12 (01): : 102 - 108
  • [47] Accuracy of a milled digital implant surgical guide: An in vitro study
    Liu, Xiaoqian
    Liu, Jianzhang
    Feng, Hailan
    Pan, Shaoxia
    JOURNAL OF PROSTHETIC DENTISTRY, 2022, 127 (03): : 453 - 461
  • [48] The Effects of Distribution of Image Matched Fiducial Markers on Accuracy of Computer-Guided Implant Surgery
    Choi, Yong-Do
    Mai, Hang-Nga
    Mai, Hai Yen
    Ha, Jung-Hong
    Li, Lin-Jie
    Lee, Du-Hyeong
    JOURNAL OF PROSTHODONTICS-IMPLANT ESTHETIC AND RECONSTRUCTIVE DENTISTRY, 2020, 29 (05): : 409 - 414
  • [49] Application of Real-Time Surgical Navigation for Zygomatic Implant Insertion in Patients With Severely Atrophic Maxilla
    Wang, Feng
    Bornstein, Michael M.
    Hung, Kuofeng
    Fan, Shengchi
    Chen, Xiaojun
    Huang, Wei
    Wu, Yiqun
    JOURNAL OF ORAL AND MAXILLOFACIAL SURGERY, 2018, 76 (01) : 80 - 87
  • [50] Comparison of the accuracy of dynamic navigation and the free hand approaches in the placement of pterygoid implants in the completely edentulous maxilla: An in vitro study
    Tao, Baoxin
    Wang, Ningtao
    Ling, Xiaowan
    Ye, Lijuan
    Wu, Yiqun
    JOURNAL OF DENTAL SCIENCES, 2024, 19 (04) : 2341 - 2347