Effect of the Configurations of Fiducial Markers on the Accuracy of Surgical Navigation in Zygomatic Implant Placement: An In Vitro Study

被引:33
|
作者
Fan, Shengchi [1 ,2 ,3 ]
Hung, Kuofeng [1 ,2 ,3 ,4 ]
Bornstein, Michael M. [4 ]
Huang, Wei [2 ,3 ,5 ]
Wang, Feng [2 ,3 ,5 ]
Wu, Yiqun [1 ,2 ,3 ]
机构
[1] Shanghai Jiao Tong Univ, Shanghai Peoples Hosp 9, Natl Clin Res Ctr Oral Dis, Dent Clin 2,Dept Oral Implantol,Coll Stomatol,Sch, Shanghai, Peoples R China
[2] Shanghai Key Lab Stomatol, Shanghai, Peoples R China
[3] Shanghai Res Inst Stomatol, Shanghai, Peoples R China
[4] Univ Hong Kong, Fac Dent, Oral & Maxillofacial Radiol Appl Oral Sci, Hong Kong, Peoples R China
[5] Shanghai Jiao Tong Univ, Shanghai Peoples Hosp 9, Natl Clin Res Ctr Oral Dis, Dept Oral Implantol,Coll Stomatol,Sch Med, Shanghai, Peoples R China
基金
国家重点研发计划;
关键词
accuracy; edentulous; navigation system; registration; zygomatic implant; TARGET REGISTRATION; SURGERY; ERROR; RECONSTRUCTION; MANAGEMENT; MODALITIES;
D O I
10.11607/jomi.6821
中图分类号
R78 [口腔科学];
学科分类号
1003 ;
摘要
Purpose: Real-time surgical navigation has been increasingly applied in implant placement. The initial registration procedures were found to substantially affect the overall accuracy, but the number and distribution of fiducial markers are yet undetermined. This study aimed to determine the minimal number and optimal distribution of fiducial markers to achieve clinically acceptable accuracy in surgical navigation for zygomatic implant placement by systematically analyzing the effects of different setups of fiducial markers on target registration error (TRE). Materials and Methods: A maxillary phantom with bone-anchored fiducial markers was scanned using cone beam computed tomography, followed by data processing on Brainlab, a commercially available navigation system. A total of 10 miniscrews were inserted in the edentulous maxilla for the configuration of the fiducial markers, with another two miniscrews as implant targets to assess the TRE in zygomatic bone. Data were then collected in nine configurations with distinct fiducial numbers and positions. Statistical analyses were performed with SPSS. Results: The accuracy of the surgical navigation system was found to depend on both the number and the position of fiducial markers. No significant difference was observed in accuracy among groups with eight fiducials and with a polygon span distribution (P > .05). When the fiducial number decreased to less than six, the markers inserted in a regular triangle were more precise than those in an inverse triangle configuration. When the number of fiducials was five with a polygonal distribution, a low TRE value of 0.59 mm was detected, which was comparable to the accuracy with more than eight fiducials in this study. Conclusion: A scattered distribution with a polygon span with at least five fiducial markers in the edentulous maxilla for registration seems to achieve acceptable TRE values with high accuracy for navigation in zygomatic implant placement.
引用
收藏
页码:85 / 90
页数:6
相关论文
共 50 条
  • [1] Reliability and accuracy of dynamic navigation for zygomatic implant placement
    Wu, Yiqun
    Tao, Baoxin
    Lan, Kengliang
    Shen, Yihan
    Huang, Wei
    Wang, Feng
    CLINICAL ORAL IMPLANTS RESEARCH, 2022, 33 (04) : 362 - 376
  • [2] Comparison of robotic system and dynamic navigation for zygomatic implant placement: An in vitro study
    Guo, Houzuo
    Jiang, Xi
    Lin, Ye
    JOURNAL OF DENTISTRY, 2024, 150
  • [3] Minimally Invasive Navigation-Guided Quad Zygomatic Implant Placement: A Comparative In Vitro Study
    Fan, Shengchi
    Gielisch, Matthias W.
    Diaz, Leonardo
    Thiem, Daniel G. E.
    Al-Nawas, Bilal
    Kaemmerer, Peer W.
    INTERNATIONAL JOURNAL OF ORAL & MAXILLOFACIAL IMPLANTS, 2023, 38 (02) : 367 - 373
  • [4] Accuracy of Zygomatic Implant Placement Using Task-Autonomous Robotic System or Dynamic Navigation: An In Vitro Study
    Chen, Jinyan
    Tao, Baoxin
    Yu, Xinbo
    Wu, Yiqun
    Wang, Feng
    CLINICAL ORAL IMPLANTS RESEARCH, 2025, 36 (02) : 178 - 190
  • [5] Effect of a dynamic navigation device on the accuracy of implant placement in the completely edentulous mandible: An in vitro study
    Feng, Yuzhang
    Yao, Yufei
    Yang, Xingmei
    JOURNAL OF PROSTHETIC DENTISTRY, 2023, 130 (05) : 731 - 737
  • [6] Real-Time Navigation in Zygomatic Implant Placement: Workflow
    Wu, Yiqun
    Wang, Feng
    Huang, Wei
    Fan, Shengchi
    ORAL AND MAXILLOFACIAL SURGERY CLINICS OF NORTH AMERICA, 2019, 31 (03) : 357 - +
  • [7] Pilot study of a surgical robot system for zygomatic implant placement
    Cao, Zhenggang
    Qin, Chunxia
    Fan, Shengchi
    Yu, Dedong
    Wu, Yiqun
    Qin, Jing
    Chen, Xiaojun
    MEDICAL ENGINEERING & PHYSICS, 2020, 75 (75) : 72 - 78
  • [8] Accuracy and efficiency of a calibration approach in dynamic navigation for implant placement: An in vitro study
    Pei, Xiyan
    Weng, Jinlong
    Sun, Feng
    Ma, Yu
    Iao, Siniong
    Liu, Xiaoqiang
    JOURNAL OF DENTAL SCIENCES, 2024, 19 (01) : 51 - 57
  • [9] Accuracy of a Computer-Aided Dynamic Navigation System in the Placement of Zygomatic Dental Implants: An In Vitro Study
    Gonzalez Rueda, Juan Ramon
    Garcia Avila, Irene
    de Paz Hermoso, Victor Manuel
    Riad Deglow, Elena
    Zubizarreta-Macho, Alvaro
    Pato Mourelo, Jesus
    Montero Martin, Javier
    Hernandez Montero, Sofia
    JOURNAL OF CLINICAL MEDICINE, 2022, 11 (05)
  • [10] Accuracy of a real-time surgical navigation system for the placement of quad zygomatic implants in the severe atrophic maxilla: A pilot clinical study
    Hung, Kuo-feng
    Wang, Feng
    Wang, Hao-wei
    Zhou, Wen-jie
    Huang, Wei
    Wu, Yi-qun
    CLINICAL IMPLANT DENTISTRY AND RELATED RESEARCH, 2017, 19 (03) : 458 - 465