Comparison between intermittent and continuous spectra optia leukapheresis systems for autologous peripheral blood stem cell collection

被引:31
作者
Lisenko, Katharina [1 ]
Pavel, Petra [2 ]
Bruckner, Thomas [3 ]
Puthenparambil, Joe [1 ]
Hundemer, Michael [1 ]
Schmitt, Anita [1 ]
Witzens-Harig, Mathias [1 ]
Ho, Anthony D. [1 ]
Wuchter, Patrick [1 ]
机构
[1] Heidelberg Univ, Dept Med 5, Neuenheimer Feld 410, D-69120 Heidelberg, Germany
[2] IKTZ Heidelberg GmbH, Stem Cell Lab, Heidelberg, Germany
[3] Heidelberg Univ, Inst Med Biometry & Informat, Heidelberg, Germany
关键词
spectra optia apheresis system; collection efficiency; PBSC; autologous transplantation; leukapheresis; APHERESIS SYSTEMS; MULTIPLE-MYELOMA; PROGENITOR CELLS; HEALTHY DONORS; EFFICIENCY; VALIDATION; PREDICTION; TRIAL;
D O I
10.1002/jca.21463
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Terumo BCT recently introduced a new system for mononuclear cell (MNC) collection that uses a Spectra Optia apheresis machine equipped with a redesigned disposable kit and software program (version 11.2). It allows for the continuous collection of MNCs, unlike the original Spectra Optia system (version 7.2), which included a chamber for two-step cell separation. The aim of this study was to compare the two apheresis systems in regard to specific performance parameters. A retrospective data analysis of 150 patients who had undergone peripheral blood stem cell collection between March of 2014 and May of 2015 at our institution was performed. For the matched comparison, patients were divided into two groups by diagnosis and by previous forms of therapy received: a homogeneous group of patients with multiple myeloma (MM) that had received first line therapy (MM group, n=88) and a heterogeneous group that included all of the other patients (other group, n=62). No significant differences in CD34+ collection yields between both collection regimens were found (p(MM)=0.19, p(other)=0.74) in either group. Moreover, similar performance ratios (collected/predicted CD34+ cell number in %) were observed (p(MM)=0.89, p(other)=0.1). No relevant variations in platelet or hemoglobin loss were found between the two systems. We conclude that the new continuous Spectra Optia MNC system is equally efficient in collecting CD34+ cells and can be used without sacrificing collection efficiency levels when treating a broad variety of autologous patients. J. Clin. Apheresis 32:27-34, 2017. (c) 2016 Wiley Periodicals, Inc.
引用
收藏
页码:27 / 34
页数:8
相关论文
共 19 条
[1]   Mobilized allogeneic peripheral stem/progenitor cell apheresis with Spectra Optia v.5.0, a novel, automatic interface-controlled apheresis system: results from the first feasibility trial [J].
Brauninger, S. ;
Bialleck, H. ;
Thorausch, K. ;
Seifried, E. ;
Bonig, H. .
VOX SANGUINIS, 2011, 101 (03) :237-246
[2]   Allogeneic donor peripheral blood "stem cell" apheresis: prospective comparison of two apheresis systems [J].
Brauninger, Susanne ;
Bialleck, Heike ;
Thorausch, Kristin ;
Felt, Tom ;
Seifried, Erhard ;
Bonig, Halvard .
TRANSFUSION, 2012, 52 (05) :1137-1145
[3]   Kinetics of standardized large volume leukapheresis (LVL) in patients do not show a recruitment phenomenon of peripheral blood progenitor cells (PBPC) [J].
Cassens, U ;
Momkvist, PH ;
Zuehlsdorf, M ;
Mohr, M ;
Kienast, J ;
Berdel, WE ;
Sibrowski, W .
BONE MARROW TRANSPLANTATION, 2001, 28 (01) :13-20
[4]   Hematopoietic stem cell donation [J].
Chen, Shu-Huey ;
Wang, Tso-Fu ;
Yang, Kuo-Liang .
INTERNATIONAL JOURNAL OF HEMATOLOGY, 2013, 97 (04) :446-455
[5]   Peripheral blood stem cell collection in low-weight children: retrospective comparison of two apheresis devices [J].
Cherqaoui, Bilade ;
Rouel, Nadege ;
Auvrignon, Anne ;
Defachelles, Anne-Sophie ;
Demeocq, Francois ;
Kanold, Justyna ;
Merlin, Etienne .
TRANSFUSION, 2014, 54 (05) :1371-1378
[6]   HPC-A Dose Prediction on the Optia® Cell Separator Based on a Benchmark CE2 Collection Efficiency: Promoting Clinical Efficiency, Minimizing Toxicity, and Allowing Quality Control [J].
Cousins, Antony F. ;
Sinclair, Joy E. ;
Alcorn, Michael J. ;
Green, Rachel H. A. ;
Douglas, Kenny W. .
JOURNAL OF CLINICAL APHERESIS, 2015, 30 (06) :321-328
[7]  
Douglas KW, 2014, CYTOTHERAPY, V16, P285, DOI [10.1016/j.jcyt.2013.12.001, 10.4161/19336896.2014.971561]
[8]   Collection of peripheral blood progenitor cells on Day 4 is feasible and effective while reducing granulocyte-colony-stimulating factor exposure to healthy donors [J].
Flommersfeld, Sabine ;
Sohlbach, Kristina ;
Jaques, Gabriele ;
Bein, Gregor ;
Hoffmann, Joerg ;
Kostrewa, Philippe ;
Sachs, Ulrich J. .
TRANSFUSION, 2015, 55 (06) :1269-1274
[9]   A single center comparison between three different apheresis systems for autologous and allogeneic stem cell collections [J].
Flommersfeld, Sabine ;
Bakchoul, Tamam ;
Bein, Gregor ;
Wachtel, Almut ;
Loechelt, Cordula ;
Sachs, Ulrich J. .
TRANSFUSION AND APHERESIS SCIENCE, 2013, 49 (03) :428-433
[10]   Change in CD34+cell concentration during peripheral blood progenitor cell collection: effects on collection efficiency and efficacy [J].
Ford, CD ;
Greenwood, J ;
Strupp, A ;
Lehman, CM .
TRANSFUSION, 2002, 42 (07) :904-911