Intensity-Modulated Microbend Fiber Optic Sensor for Respiratory Monitoring and Gating During MRI

被引:64
作者
Lau, Doreen [1 ]
Chen, Zhihao [2 ]
Teo, Ju Teng [2 ]
Ng, Soon Huat [2 ]
Rumpel, Helmut [1 ]
Lian, Yong [3 ]
Yang, Hui [1 ]
Kei, Pin Lin [1 ]
机构
[1] Singapore Gen Hosp, Dept Diagnost Radiol, Singapore 169608, Singapore
[2] Inst Infocomm Res, Neural & Biomed Technol Dept, Singapore 138632, Singapore
[3] Natl Univ Singapore, Dept Elect & Comp Engn, Singapore 117576, Singapore
关键词
Fiber optic sensor (FOS); magnetic resonance imaging (MRI); microbending; respiratory gating; respiratory monitoring; HEART;
D O I
10.1109/TBME.2013.2262150
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
This paper describes a novel microbend fiber optic sensor system for respiratory monitoring and respiratory gating in the MRI environment. The system enables the noninvasive real-time monitoring and measurement of breathing rate and respiratory/body movement pattern of healthy subjects inside the MRI gantry, and has potential application in respiratory-gated image acquisition based on respiratory cues. The working principle behind this sensor is based on the microbending effect of an optical fiber on light transmission. The sensor system comprises of a 1.0-mm-thin graded-index multimode optical fiber-embedded plastic sensormat, a photoelectronic transceiver, and a computer with a digital signal processing algorithm. In vitro testing showed that our sensor has a typical signal-to-noise ratio better than 28 dB. Clinical MRI trials conducted on 20 healthy human subjects showed good and comparable breathing rate detection (with an accuracy of +/-2 bpm) and respiratory-gated image quality produced using the sensor system, with reference to current predicate hospital device/system. The MRI safe, ease of operation characteristics, low fabrication cost, and extra patient comfort offered by this system suggest its good potential in replacing predicate device/system and serve as a dual function in real-time respiratory monitoring and respiratory-gated image acquisition at the same time during MRI.
引用
收藏
页码:2655 / 2662
页数:8
相关论文
共 26 条
[11]   MICROBEND FIBEROPTIC SENSOR [J].
LAGAKOS, N ;
COLE, JH ;
BUCARO, JA .
APPLIED OPTICS, 1987, 26 (11) :2171-2180
[12]   A microbent fiber optic pH sensor [J].
Lee, ST ;
Aneeshkumar, B ;
Radhakrishnan, P ;
Vallabhan, CPG ;
Nampoori, VPN .
OPTICS COMMUNICATIONS, 2002, 205 (4-6) :253-256
[13]   MONITORING OF RESPIRATORY AND HEART-RATES USING A FIBEROPTIC SENSOR [J].
LINDBERG, LG ;
UGNELL, H ;
OBERG, PA .
MEDICAL & BIOLOGICAL ENGINEERING & COMPUTING, 1992, 30 (05) :533-537
[14]   A plastic optical fiber microbend sensor used as a low-cost anti-squeeze detector [J].
Linec, Matjaz ;
Donlagic, Denis .
IEEE SENSORS JOURNAL, 2007, 7 (9-10) :1262-1267
[15]   Fibre-optic sensors in health care [J].
Mignani, AG ;
Baldini, F .
PHYSICS IN MEDICINE AND BIOLOGY, 1997, 42 (05) :967-979
[16]  
Pinet É, 2008, NAT PHOTONICS, V2, P150, DOI 10.1038/nphoton.2008.19
[17]   The AAPM/RSNA physics tutorial for residents - MR imaging safety considerations [J].
Price, RR .
RADIOGRAPHICS, 1999, 19 (06) :1641-1651
[18]   Film-Type Sensor Materials PVDF and EMFi in Measurement of Cardiorespiratory Signals-A Review [J].
Rajala, Satu ;
Lekkala, Jukka .
IEEE SENSORS JOURNAL, 2012, 12 (03) :439-446
[19]  
Roopa G, 2011, IEEE SENSOR, P1788, DOI 10.1109/ICSENS.2011.6127073
[20]  
Shellock F G, 1986, Med Electron, V17, P93