Dosimetric performance of a multipoint plastic scintillator dosimeter as a tool for real-time source tracking in high dose rate192Ir brachytherapy

被引:21
|
作者
Rosales, Haydee M. Linares [1 ,2 ,3 ]
Archambault, Louis [1 ,2 ,3 ]
Beddar, Sam [4 ,5 ]
Beaulieu, Luc [1 ,2 ,3 ]
机构
[1] Univ Laval, Dept Phys Genie Phys & Opt, Quebec City, PQ, Canada
[2] Univ Laval, Ctr Rech Sur Canc, Quebec City, PQ, Canada
[3] Univ Laval, Dept Radiooncol & Axe Oncol, CRCHU Quebec, CHU Quebec, Quebec City, PQ, Canada
[4] Univ Texas MD Anderson Canc Ctr, Dept Radiat Phys, Houston, TX 77030 USA
[5] Univ Texas MD Anderson, UTHlth, Grad Sch Biomed Sci, Houston, TX USA
基金
加拿大自然科学与工程研究理事会;
关键词
HDR brachytherapy; in vivodosimetry; multipoint plastic scintillator detector; plastic scintillator; source tracking; ENERGY BEAM DOSIMETRY; IN-VIVO DOSIMETRY; THERMOLUMINESCENCE DOSIMETRY; TEMPERATURE-DEPENDENCE; MONTE-CARLO; DETECTORS; PHOTON; VERIFICATION; ELECTRON; LIGHT;
D O I
10.1002/mp.14246
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Purpose This study aims to present the performance of a multipoint plastic scintillation detector (mPSD) as a tool for real-time dose measurements (covering three orders of magnitude in dose rate), source-position triangulation, and dwell time assessment in high dose rate (HDR) brachytherapy. Methods A previously characterized and optimized three-point sensor system was used for HDR brachytherapy measurements. The detector was composed of three scintillators: BCF60, BCF12, and BCF10. Scintillation light was transmitted through a single 1-mm-diameter clear optical fiber and read by a compact assembly of photomultiplier tubes (PMTs). Each component was numerically optimized to allow for signal deconvolution using a multispectral approach, taking care of the Cerenkov stem effect as well as extracting the dose from each scintillator. The PMTs were read simultaneously using a data acquisition board at a rate of 100 KHz and controlled with in-house software based on Python. An192Irsource (Flexitron, Elekta-Brachy) was remotely controlled and sent to various positions in a in-house PMMA holder, ensuring 0.1 mm positional accuracy. Dose measurements covering a range of 10 cm of source movement were carried out according to TG-43 U1 recommendations. Water measurements were performed in order to: (a) characterize the system's response in terms of angular dependence; (b) obtain the relative contribution of positioning and measurement uncertainties to the total system uncertainty; (c) assess the system's temporal resolution; and (d) track the source position in real time. The triangulation principle was applied to report the source position in three-dimensional space. Results As expected, the positioning uncertainty dominated close to the source, whereas the measurement uncertainty dominated at larger distances. A maximum measurement uncertainty of 17 % was observed for the BCF60 scintillator at 10 cm from the source. Based on the uncertainty chain, the best compromises between positioning and measurement uncertainties were reached at 17.2, 17.4, and 17.5 mm for the BCF10, BCF12, and BCF60 scintillators, respectively, which also corresponded to the recommended optimal distances to the source for calibration purposes. The detector further exhibited no angular dependence. All dose values were found to be within 2% of the dose value at90 circle. In the experiments performed for source-position determination, the system provided an average location with a standard deviation under 1.7 mm. The maximum observed differences between measured and expected values were 1.82 and 1.8 mm in the x- and z-directions, respectively. Deviations between the mPSD measurements and expected TG-43 values were below 5% in all the explored measurement conditions. With regard to dwell time measurement accuracy, the maximum deviation observed at all distances was 0.56 +/- 0.25 s, with a weighted average of the three scintillators below 0.33 +/- 0.37 s at all distances covered in this study. Conclusions Real-time HDR brachytherapy measurements were performed with an optimized mPSD system. The performance of the system demonstrated that it could be used for simultaneous, in vivo, real-time reporting of dose, dwell time, and source position during HDR brachytherapy.
引用
收藏
页码:4477 / 4490
页数:14
相关论文
共 44 条
  • [21] A system for localization of high dose rate 192Ir source during brachytherapy treatment with silicon detectors
    Batic, Matej
    Burger, Janez
    Cindro, Vladimir
    Kramberger, Gregor
    Mandic, Igor
    Mikuz, Marko
    Studen, Andrej
    Zavrtanik, Marko
    2009 IEEE NUCLEAR SCIENCE SYMPOSIUM CONFERENCE RECORD, VOLS 1-5, 2009, : 3794 - +
  • [22] Development of a low-sensitivity high resolution YAP(Ce) scintillation camera system toward the real-time imaging of an 192Ir source during high-dose-rate brachytherapy
    Nagata, J.
    Yamamoto, S.
    Noguchi, Y.
    Nakaya, T.
    Okudaira, K.
    Kamada, K.
    Yoshikawa, A.
    JOURNAL OF INSTRUMENTATION, 2020, 15 (12):
  • [23] Effect of inhomogeneities and source position on dose distribution of nucletron high dose rate Ir-192 brachytherapy source by Monte Carlo simulation
    Chandola, R. M.
    Tiwari, S.
    Kowar, M. K.
    Choudhary, V.
    JOURNAL OF CANCER RESEARCH AND THERAPEUTICS, 2010, 6 (01) : 54 - 57
  • [24] First experience of 192Ir source stuck event during high-dose-rate brachytherapy in Japan
    Kumagai, Shinobu
    Arai, Norikazu
    Takata, Takeshi
    Kon, Daisuke
    Saitoh, Toshiya
    Oba, Hiroshi
    Furui, Shigeru
    Kotoku, Jun'ichi
    Shiraishi, Kenshiro
    JOURNAL OF CONTEMPORARY BRACHYTHERAPY, 2020, 12 (01) : 53 - 60
  • [25] Evaluation of a real-time BeO ceramic fiber-coupled luminescence dosimetry system for dose verification of high dose rate brachytherapy
    Santos, Alexandre M. Caraca
    Mohammadi, Mohammad
    Afshar, Shahraam, V
    MEDICAL PHYSICS, 2015, 42 (11) : 6349 - 6356
  • [26] Development and implementation of a remote audit tool for high dose rate (HDR) Ir-192 brachytherapy using optically stimulated luminescence dosimetry
    Casey, Kevin E.
    Alvarez, Paola
    Kry, Stephen F.
    Howell, Rebecca M.
    Lawyer, Ann
    Followill, David
    MEDICAL PHYSICS, 2013, 40 (11)
  • [27] A surrogate urethra for real-time planning of high-dose-rate prostate brachytherapy
    Halperin, Heather
    Hilts, Michelle
    Crook, Juanita
    Batchelar, Deidre
    Tisseverasinghe, Steven
    Tetreault-LaFlamme, Audrey
    Bachand, Francois
    BRACHYTHERAPY, 2019, 18 (05) : 675 - 682
  • [28] Verification of High Dose Rate 192Ir Source Position During Brachytherapy Treatment Using Silicon Pixel Detectors
    Batic, Matej
    Burger, Janez
    Cindro, Vladimir
    Kramberger, Gregor
    Mandic, Igor
    Mikuz, Marko
    Studen, Andrej
    Zavrtanik, Marko
    IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 2011, 58 (05) : 2250 - 2256
  • [29] A generic high-dose rate 192Ir brachytherapy source for evaluation of model-based dose calculations beyond the TG-43 formalism
    Ballester, Facundo
    Tedgren, Asa Carlsson
    Granero, Domingo
    Haworth, Annette
    Mourtada, Firas
    Fonseca, Gabriel Paiva
    Zourari, Kyveli
    Papagiannis, Panagiotis
    Rivard, Mark J.
    Siebert, Frank-Andre
    Sloboda, Ron S.
    Smith, Ryan L.
    Thomson, Rowan M.
    Verhaegen, Frank
    Vijande, Javier
    Ma, Yunzhi
    Beaulieu, Luc
    MEDICAL PHYSICS, 2015, 42 (06) : 3048 - 3062
  • [30] Dosimetric characterization of the M-15 high-dose-rate Iridium-192 brachytherapy source using the AAPM and ESTRO formalism
    Minh-Tri Ho Thanh
    Munro, John J., III
    Medich, David C.
    JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2015, 16 (03): : 305 - 317