Participation in the absolute gravity comparison with a compact cold atom gravimeter

被引:28
作者
Fu, Zhijie [1 ,2 ]
Wang, Qiyu [1 ]
Wang, Zhaoying [1 ]
Wu, Bin [2 ]
Cheng, Bing [2 ]
Lin, Qiang [2 ]
机构
[1] Zhejiang Univ, Inst Opt, Dept Phys, Hangzhou 310027, Zhejiang, Peoples R China
[2] Zhejiang Univ Technol, Coll Sci, Ctr Opt & Optoelect Res, Hangzhou 310023, Zhejiang, Peoples R China
基金
中国国家自然科学基金;
关键词
NEW-GENERATION; LASER SYSTEM; PRECISION; VOLCANO;
D O I
10.3788/COL201917.011204
中图分类号
O43 [光学];
学科分类号
070207 ; 0803 ;
摘要
The first Asia-Pacific Comparison of Absolute Gravimeters ( APMP.M.G-K1) was organized by the National Institute of Metrology (NIM) of China from December 21, 2015 to March 25, 2016 in Changping, Beijing. Our compact cold atom gravimeter (CCAG) was transported from Hangzhou to Beijing with a long distance of about 1200 km to participate in this comparison. The CCAG is the only one, to the best of our knowledge, that is based on the principle of atom interferometry among all the instruments. Absolute gravity in the indicated three test sites has been measured as requested by the organizer. The sensitivity of our CCAG is estimated to be 90 mu Gal/root Hz, even when the measurements are carried out without any vibration isolation. Besides, the accuracy of this gravimeter has been evaluated to be about 19 mu Gal by considering the significant system errors. Our results show a good agreement with the given reference value.
引用
收藏
页数:6
相关论文
共 34 条
  • [31] Shift evaluation of the atomic gravimeter NIM-AGRb-1 and its comparison with FG5X
    Wang, Shao-Kai
    Zhao, Yang
    Zhuang, Wei
    Li, Tian-Chu
    Wu, Shu-Qing
    Feng, Jin-Yang
    Li, Chun-Jian
    [J]. METROLOGIA, 2018, 55 (03) : 360 - 365
  • [32] Reaching a few 10-15 long-term stability of integrating sphere cold atom clock
    Wang, Yaning
    Meng, Yanling
    Wan, Jinyin
    Xiao, Ling
    Yu, Mingyuan
    Wang, Xin
    Ouyang, Xinchuan
    Cheng, Huadong
    Liu, Liang
    [J]. CHINESE OPTICS LETTERS, 2018, 16 (07)
  • [33] Velocity-tunable cold Cs atomic beam from a magneto-optical trap
    Yan, Xueshu
    Wu, Chenfei
    Huang, Jiaqiang
    Zhang, Jianwei
    Wang, Lijun
    [J]. CHINESE OPTICS LETTERS, 2017, 15 (04)
  • [34] Test of Equivalence Principle at 10-8 Level by a Dual-Species Double-Diffraction Raman Atom Interferometer
    Zhou, Lin
    Long, Shitong
    Tang, Biao
    Chen, Xi
    Gao, Fen
    Peng, Wencui
    Duan, Weitao
    Zhong, Jiaqi
    Xiong, Zongyuan
    Wang, Jin
    Zhang, Yuanzhong
    Zhan, Mingsheng
    [J]. PHYSICAL REVIEW LETTERS, 2015, 115 (01)