Computational fluid dynamics simulation of blood flow profile and shear stresses in bileaflet mechanical heart valve by using monolithic approach

被引:19
作者
Kadhim, Saleem Khalefa [1 ]
Nasif, Mohammad Shakir [1 ]
Al-Kayiem, Hussain H. [1 ]
Al-Waked, Rafat [2 ]
机构
[1] Univ Teknol PETRONAS, Mech Engn Dept, Block 17, Seri Iskandar 32610, Perak, Malaysia
[2] German Jordanian Univ, Sch Appl Tech Sci, Amman, Jordan
来源
SIMULATION-TRANSACTIONS OF THE SOCIETY FOR MODELING AND SIMULATION INTERNATIONAL | 2018年 / 94卷 / 02期
关键词
Bileaflet mechanical heart valve; computational fluid dynamics; blood flow shear stresses; damage to blood cells; NUMERICAL-ANALYSIS; VALIDATION;
D O I
10.1177/0037549717712603
中图分类号
TP39 [计算机的应用];
学科分类号
081203 ; 0835 ;
摘要
Bileaflet mechanical heart valves (BMHVs) are widely used to replace diseased heart valves. However, patients may suffer from implant complications, such as platelet aggregation and damage to blood cells, which could lead to BMHV failure. These complications are related to the blood flow patterns in the BMHV. A three-dimensional computational fluid dynamic (CFD) model was developed to investigate blood hydrodynamics and shear stresses at different cardiac cycles. A user-defined function (UDF) code was developed to model the valve leaflet motion. This UDF updates the tetrahedral mesh according to the location of the valve leaflet, which enables modeling of complicated moving geometries and achieves solution convergence with ease without the need to adjust the relaxation factor values. The agreement between the experimental and numerical results indicates that the developed model could be used with confidence to simulate BMHV motion and blood flow. Furthermore, valve leaflet and valve pivot were found to be continuously exposed to shear stresses higher than 52.3 Pa which according to previous research findings may cause damage to blood platelets.
引用
收藏
页码:93 / 104
页数:12
相关论文
共 23 条
  • [1] Investigating the Use of Turbulence Models for Flow Investigations in a Positive Displacement Ventricular Assist Device
    Al-Azawy, Mohammed G.
    Turan, Ali
    Revell, Alistair
    [J]. 6TH EUROPEAN CONFERENCE OF THE INTERNATIONAL FEDERATION FOR MEDICAL AND BIOLOGICAL ENGINEERING, 2015, 45 : 395 - 398
  • [2] Validation of a numerical FSI simulation of an aortic BMHV by in vitro PIV experiments
    Annerel, S.
    Claessens, T.
    Degroote, J.
    Segers, P.
    Vierendeels, J.
    [J]. MEDICAL ENGINEERING & PHYSICS, 2014, 36 (08) : 1014 - 1023
  • [3] [Anonymous], 2012, Cardiovasc. Eng. Technol., DOI DOI 10.1007/S13239-012-0087-5
  • [4] Cefalu J, P 2011 MIDW SECT C A, P1
  • [5] Pulsatile blood flows through a bileaflet mechanical heart valve with different approach methods of numerical analysis; Pulsatile flows with fixed leaflets and interacted with moving leaflets
    Choeng Ryul Choi
    Chang Nyung Kim
    Young Joo Kwon
    Jae Won Lee
    [J]. KSME International Journal, 2003, 17 (7): : 1073 - 1082
  • [6] Vorticity dynamics of a bileaflet mechanical heart valve in an axisymmetric aorta
    Dasi, L. P.
    Ge, L.
    Simon, H. A.
    Sotiropoulos, F.
    Yoganathan, A. P.
    [J]. PHYSICS OF FLUIDS, 2007, 19 (06)
  • [7] On the effect of aortic root geometry on the coronary entry-flow after a bileaflet mechanical heart valve implant: a numerical study
    de Tullio, M. D.
    Pedrizzetti, G.
    Verzicco, R.
    [J]. ACTA MECHANICA, 2011, 216 (1-4) : 147 - 163
  • [8] Validation of a fluid-structure interaction model for a bileaflet mechanical heart valve
    Forsythe, Neil
    Mueller, Jens-Dominik
    [J]. INTERNATIONAL JOURNAL OF COMPUTATIONAL FLUID DYNAMICS, 2008, 22 (08) : 541 - 553
  • [9] Fortini S, 16 INT S APPL LAS TE
  • [10] A time-dependent numerical analysis of flow in a mechanical heart valve: Comparison with experimental results
    Gkanis, Vasileios
    Housiadas, Christos
    [J]. INTERNATIONAL JOURNAL OF COMPUTATIONAL FLUID DYNAMICS, 2010, 24 (05) : 157 - 168