Detection of Chemical Warfare Agents with a Miniaturized High-Performance Drift Tube Ion Mobility Spectrometer Using High-Energetic Photons for Ionization

被引:31
作者
Ahrens, Andre [2 ]
Allers, Maria [1 ]
Bock, Henrike [1 ]
Hitzemann, Moritz [2 ]
Ficks, Arne [1 ]
Zimmermann, Stefan [2 ]
机构
[1] Bundeswehr Res Inst Protect Technol & CBRN Protect, D-29633 Munster, Germany
[2] Leibniz Univ Hannover, Inst Elect Engn & Measurement Technol, Dept Sensors & Measurement Technol, D-30167 Hannover, Germany
关键词
D O I
10.1021/acs.analchem.2c03422
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
A growing demand for low-cost gas sensors capable of detecting the smallest amounts of highly toxic substances in air, including chemical warfare agents (CWAs) and toxic industrial chemicals (TICs), has emerged in recent years. Ion mobility spectrometers (IMS) are particularly suitable for this application due to their high sensitivity and fast response times. In view of the preferred mobile use of such devices, miniaturized ion drift tubes are required as the core of IMS-based lightweight, low-cost, hand-held gas detectors. Thus, we evaluate the suitability of a miniaturized ion mobility spectrometer featuring an ion drift tube length of just 40 mm and a high resolving power of R-p = 60 for the detection of various CWAs, such as nerve agents sarin (GB), tabun (GA), soman (GD), and cyclosarin (GF), as well as the blister agent sulfur mustard (HD), the blood agent hydrogen cyanide (AC) and the choking agent chlorine (CL). We report on the limits of detection reaching minimum concentration levels of, for instance, 29 ppt(v) for sarin (GB) within an averaging time of only 1 s. Furthermore, we investigate the effects of precursors, simulants, and other common interfering substances on false positive alarms.
引用
收藏
页码:15440 / 15447
页数:8
相关论文
共 38 条
[1]   Miniaturized high-performance drift tube ion mobility spectrometer [J].
Ahrens, Andre ;
Hitzemann, Moritz ;
Zimmermann, Stefan .
INTERNATIONAL JOURNAL FOR ION MOBILITY SPECTROMETRY, 2019, 22 (02) :77-83
[2]  
Ahrens J., 2020, SZENARIEN EVALUATION
[3]  
AIRSENSE Analytics GmbH, GAS DET ARR PERS GDA
[4]   Detection of Volatile Toxic Industrial Chemicals with Classical Ion Mobility Spectrometry and High-Kinetic Energy Ion Mobility Spectrometry [J].
Allers, Maria ;
Schaefer, Christoph ;
Ahrens, Andre ;
Schlottmann, Florian ;
Hitzemann, Moritz ;
Kobelt, Tim ;
Zimmermann, Stefan ;
Hetzer, Ralf .
ANALYTICAL CHEMISTRY, 2022, 94 (02) :1211-1220
[5]  
Baranoski J. M, 2003, EVALUATION RAID M CH
[6]   The response of a membrane inlet ion mobility spectrometer to chlorine and the effect of water contamination of the drying media on ion mobility spectrometric responses to chlorine [J].
Bocos-Bintintan, V ;
Brittain, A ;
Thomas, CLP .
ANALYST, 2001, 126 (09) :1539-1544
[7]   Fast Sensing of Hydrogen Cyanide (HCN) Vapors Using a Hand-Held Ion Mobility Spectrometer with Nonradioactive Ionization Source [J].
Bocos-Bintintan, Victor ;
Ratiu, Ileana Andreea .
SENSORS, 2021, 21 (15)
[8]   Simulation aided design of a low cost ion mobility spectrometer based on printed circuit boards [J].
Bohnhorst, A. ;
Kirk, A. T. ;
Zimmermann, S. .
INTERNATIONAL JOURNAL FOR ION MOBILITY SPECTROMETRY, 2016, 19 (2-3) :167-174
[9]   Recent Developments in Ion Mobility Spectrometry [J].
Borsdorf, Helko ;
Mayer, Thomas ;
Zarejousheghani, Mashaalah ;
Eiceman, Gary A. .
APPLIED SPECTROSCOPY REVIEWS, 2011, 46 (06) :472-521
[10]  
Bruker Corporation, RAID M100PLUS HANDH