The Position and Number of Screws Influence Screw Perforation of the Humeral Head in Modern Locking Plates: A Cadaver Study

被引:51
作者
Erhardt, Johannes B. [1 ,2 ]
Stoffel, Karl [1 ,2 ]
Kampshoff, Jorg [1 ,2 ]
Badur, Nicole [2 ]
Yates, Piers [2 ]
Kuster, Markus S. [1 ,2 ]
机构
[1] Cantonal Hosp St Gallen, Dept Orthopaed Surg, CH-9007 St Gallen, Switzerland
[2] Univ Western Australia, Fremantle Orthopaed Unit, Fremantle, WA, Australia
关键词
cut out; screw perforation; proximal humerus fracture; polyaxial; locking plate; ANGULAR STABLE PLATE; PROXIMAL HUMERUS; BIOMECHANICAL EVALUATION; INTERNAL-FIXATION; STATISTICAL POWER; OPEN REDUCTION; FRACTURES; MULTICENTER; PHILOS;
D O I
10.1097/BOT.0b013e31823db922
中图分类号
R826.8 [整形外科学]; R782.2 [口腔颌面部整形外科学]; R726.2 [小儿整形外科学]; R62 [整形外科学(修复外科学)];
学科分类号
摘要
Objectives: Screw perforation of the humeral head in locking plate osteosynthesis occurs in up to 30% of cases. The current study compared different fixation possibilities (eg, number and position of screws) to reduce screw perforation in the humeral head. Methods: A humeral head fracture with a missing medial support was created in 30 fresh-frozen cadavers and fixed with a polyaxial locking plate (NCB PH; Zimmer, Warsaw, IN). The constructs were loaded with increasing force and the number of cycles until screw perforation was recorded. Four different fixation methods were tested: group 1 five screws with fixed angle, group 2 five screws in polyaxial position according to bone strength, group 3 three screws, and group 4 five screws with 1 as an inferomedial support screw. Results: More screws in the humeral head significantly increased the number of cycles before screw perforation. An inferomedial support screw further increased the number of cycles. Polyaxial screw placement compared with fixed-angle placement had no effect on the screw perforation phenomenon. Conclusions: We recommend to position an inferomedial support screw, and at least 5 screws in the head fragment, when using a locking plate in proximal humerus fractures with disrupted medial hinge.
引用
收藏
页码:E188 / E192
页数:5
相关论文
共 21 条
  • [1] Open Reduction and Internal Fixation of Proximal Humerus Fractures Using a Proximal Humeral Locked Plate: A Prospective Multicenter Analysis
    Brunner, Felix
    Sommer, Christoph
    Bahrs, Christian
    Hettwinkel, Rainer
    Hafner, Christian
    Rillmann, Paavo
    Kohut, Georges
    Ekelund, Anders
    Muller, Mathias
    Audige, Laurent
    Babst, Reto
    [J]. JOURNAL OF ORTHOPAEDIC TRAUMA, 2009, 23 (03) : 163 - 172
  • [2] Proximal humeral internal locking system (PHILOS) for the treatment of proximal humeral fractures
    Charalambous, C. P.
    Siddique, I.
    Valluripalli, K.
    Kovacevic, M.
    Panose, P.
    Srinivasan, M.
    Marynissen, H.
    [J]. ARCHIVES OF ORTHOPAEDIC AND TRAUMA SURGERY, 2007, 127 (03) : 205 - 210
  • [3] Two-part surgical neck fractures of the proximal part of the humerus - A biomechanical evaluation of two fixation nchniques
    Edwards, Sara L.
    Wilson, Nicole A.
    Zhang, Li-Qun
    Flores, Steven
    Merk, Bradley R.
    [J]. JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 2006, 88A (10) : 2258 - 2264
  • [4] Early results in the treatment of proximal humeral fractures with a polyaxial locking plate
    Erhardt, Johannes B.
    Roderer, G.
    Grob, K.
    Forster, T. N.
    Stoffel, K.
    Kuster, M. S.
    [J]. ARCHIVES OF ORTHOPAEDIC AND TRAUMA SURGERY, 2009, 129 (10) : 1367 - 1374
  • [5] G*Power 3: A flexible statistical power analysis program for the social, behavioral, and biomedical sciences
    Faul, Franz
    Erdfelder, Edgar
    Lang, Albert-Georg
    Buchner, Axel
    [J]. BEHAVIOR RESEARCH METHODS, 2007, 39 (02) : 175 - 191
  • [6] An algorithm for estimation of shoulder muscle forces for clinical use
    Favre, P
    Sheikh, R
    Fucentese, SF
    Jacob, HAC
    [J]. CLINICAL BIOMECHANICS, 2005, 20 (08) : 822 - 833
  • [7] Sample size and statistical power of randomised, controlled trials in orthopaedics
    Freedman, KB
    Back, S
    Bernstein, J
    [J]. JOURNAL OF BONE AND JOINT SURGERY-BRITISH VOLUME, 2001, 83B (03): : 397 - 402
  • [8] The importance of medial support in locked plating of proximal humerus fractures
    Gardner, Michael J.
    Weil, Yoram
    Barker, Joseph U.
    Kelly, Bryan T.
    Helfet, David L.
    Lorich, Dean G.
    [J]. JOURNAL OF ORTHOPAEDIC TRAUMA, 2007, 21 (03) : 185 - 191
  • [9] Hybrid locked plating of osteoporotic fractures of the humerus
    Gardner, Michael J.
    Griffith, Matthew H.
    Demetrakopoulos, Demetris
    Brophy, Robert H.
    Grose, Andrew
    Helfet, David L.
    Lorich, Dean G.
    [J]. JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 2006, 88A (09) : 1962 - 1967
  • [10] Where should implants be anchored in the humeral head?
    Hepp, P
    Lill, H
    Bail, H
    Korner, J
    Niederhagen, M
    Haas, NP
    Josten, C
    Duda, GN
    [J]. CLINICAL ORTHOPAEDICS AND RELATED RESEARCH, 2003, (415) : 139 - 147