Simultaneous ultratrace determination of Zr and Nb in chromium matrixes with ICP-dynamic reaction cell MS

被引:42
作者
Hattendorf, B [1 ]
Günther, D [1 ]
Schönbächler, M [1 ]
Halliday, A [1 ]
机构
[1] Swiss Fed Inst Technol, CH-8092 Zurich, Switzerland
关键词
D O I
10.1021/ac015549a
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
A dynamic reaction cell (DRC) has been used to minimize the formation of metal-argide ions in inductively coupled plasma mass spectrometry and applied to the determination of Zr and Nb in Cr-rich samples. The formation of ArCr+ species from the plasma gas and the sample matrix was reduced by ion molecule reactions inside a DRC of the ICPMS used. Hydrogen was used as reaction gas, and the efficiency in the reduction of ArCr+ was similar to that of other plasma-based polyatomic ions as reported in an earlier study. The formation of CrOx+ ions is enhanced when the DRC is operated in pressurized mode. Adjustment of the transmission properties of the band-pass quadrupole to reject precursor ions can be achieved without dramatic decrease of sensitivity but with a significant improvement in the signal/background ratio. Measurements in solutions containing concentrations of up to 2 g/L Cr showed that the determination of Nb and Zr is possible in the nanogram per liter range in such a matrix. The limits of detection for Nb and Zr in pure Cr metal have been estimated at 2 ng/g for Nb and 5 ng/g for Zr. Analysis of basaltic reference samples resulted in very good agreement with previously published data.
引用
收藏
页码:5494 / 5498
页数:5
相关论文
共 19 条
[1]  
[Anonymous], 1989, GEOSTANDARD NEWSLETT, DOI DOI 10.1111/J.1751-908X.1989.TB00476.X
[2]   Investigations on cluster and molecular ion formation by plasma mass spectrometry [J].
Becker, JS ;
Dietze, HJ .
FRESENIUS JOURNAL OF ANALYTICAL CHEMISTRY, 1997, 359 (4-5) :338-345
[3]  
Bollinger DS, 1999, ATOM SPECTROSC, V20, P60
[4]  
Denoyer ER, 1999, SPECTROSCOPY, V14, P43
[5]   Analytical performance of the plasma source RF quadrupole ion trap in elemental and isotopic MS [J].
Eiden, GC ;
Barinaga, CJ ;
Koppenaal, DW .
JOURNAL OF ANALYTICAL ATOMIC SPECTROMETRY, 1999, 14 (08) :1129-1132
[6]   Capabilities and limits of ICP-MS for direct determination of element traces in saline solutions [J].
Falk, H ;
Geerling, R ;
Hattendorf, B ;
KrengelRothensee, K ;
Schmidt, KP .
FRESENIUS JOURNAL OF ANALYTICAL CHEMISTRY, 1997, 359 (4-5) :352-356
[7]   Application of a hexapole collision and reaction cell in ICP-MS Part I: Instrumental aspects and operational optimization [J].
Feldmann, I ;
Jakubowski, N ;
Stuewer, D .
FRESENIUS JOURNAL OF ANALYTICAL CHEMISTRY, 1999, 365 (05) :415-421
[8]  
GERLICH D, 1992, ADV CHEM PHYS, V82, P1
[9]   Experimental evidence for the formation of doubly charged oxide and hydroxide ions in inductively coupled plasma mass spectrometry [J].
Hattendorf, B ;
Günther, D .
FRESENIUS JOURNAL OF ANALYTICAL CHEMISTRY, 2001, 370 (05) :483-487
[10]   Characteristics and capabilities of an ICP-MS with a dynamic reaction cell for dry aerosols and laser ablation [J].
Hattendorf, B ;
Günther, D .
JOURNAL OF ANALYTICAL ATOMIC SPECTROMETRY, 2000, 15 (09) :1125-1131