Spatially resolved measurements to improve analytical performance of solution-cathode glow discharge optical-emission spectrometry

被引:29
|
作者
Schwartz, Andrew J. [1 ]
Ray, Steven J. [1 ,2 ]
Chan, George C. -Y. [1 ,3 ]
Hieftje, Gary M. [1 ]
机构
[1] Indiana Univ, Dept Chem, Bloomington, IN 47405 USA
[2] SUNY Buffalo, Dept Chem, Buffalo, NY 14260 USA
[3] Lawrence Berkeley Natl Lab, Berkeley, CA 94720 USA
基金
美国能源部;
关键词
Solution-cathode glow discharge; Optical-emission spectrometry; Instrumentation; Spatial discrimination; FLAGGING MATRIX INTERFERENCES; ATMOSPHERIC-PRESSURE; ELEMENTAL ANALYSIS; ONLINE SEPARATION; AQUEOUS SAMPLES; SYSTEM DRIFT; PRECONCENTRATION; PROFILES; COMPACT; SILICA;
D O I
10.1016/j.sab.2016.10.004
中图分类号
O433 [光谱学];
学科分类号
0703 ; 070302 ;
摘要
Past studies of the solution-cathode glow discharge (SCGD) revealed that elemental and molecular emission are not spatially homogenous throughout the source, but rather conform to specific zones within the discharge. Exploiting this inhomogeneity can lead to improved analytical performance if emission is collected only from regions of the discharge where analyte species emit strongly and background emission (from continuum, elemental and/or molecular sources) is lower. Effects of this form of spatial discrimination on the analytical performance of SCGD optical emission spectrometry (OES) have been investigated with an imaging spectrograph for fourteen atomic lines, with emphasis on detection limits and precision. Vertical profiles of the emission intensity, signal to -background ratio, and signal-to-noise ratio were collected and used to determine the optimal region to view the SCGD on a per-element basis. With optimized spatial filtering, detection limits ranged from 0.09-360 ppb, a 1.4-13.6 fold improvement over those obtained when emission is collected from the full vertical profile (1.1-840 ppb), with a 4.2-fold average improvement. Precision was found to be unaffected by spatial filtering, ranging from 0.5-2.6% relative standard deviation (RSD) for all elements investigated, closely comparable to the 0.4-2.4% RSD observed when no spatial filtering is used. Spatial profiles also appear useful for identifying optimal line pairs for internal standardization and for flagging the presence of matrix interferences in SCGD-OES. (C) 2016 Elsevier B.V. All rights reserved.
引用
收藏
页码:168 / 176
页数:9
相关论文
共 50 条
  • [21] Analysis of metal elements by solution cathode glow discharge-atomic emission spectrometry with a modified pulsation damper
    Wang, Jinmei
    Tang, Pengfei
    Zheng, Peichao
    Zhai, Xiang
    JOURNAL OF ANALYTICAL ATOMIC SPECTROMETRY, 2017, 32 (10) : 1925 - 1931
  • [22] Determination of gallium and indium by solution cathode glow discharge as an excitation source for atomic emission spectrometry
    Yu, Jie
    Kang, Yuejing
    Lu, Quanfang
    Luo, Hui
    Cui, Lijuan
    Lu, Zhuanhong
    Li, Juanlong
    SPECTROCHIMICA ACTA PART B-ATOMIC SPECTROSCOPY, 2020, 172
  • [23] Determination of trace sodium, lithium, magnesium, and potassium impurities in colloidal silica by slurry introduction into an atmospheric-pressure solution-cathode glow discharge and atomic emission spectrometry
    Wang, Zheng
    Schwartz, Andrew J.
    Ray, Steven J.
    Hieftje, Gary M.
    JOURNAL OF ANALYTICAL ATOMIC SPECTROMETRY, 2013, 28 (02) : 234 - 240
  • [24] Comprehensive evaluation of matrix effect in the detection of Zn, Cd, Ni, Cu, Cr and Pb by solution cathode glow discharge optical emission spectrometry
    Cai, Zhaoqing
    Ni, Zhijun
    Gao, Meng
    Wang, Zheng
    MICROCHEMICAL JOURNAL, 2023, 195
  • [25] Determination of gallium in water samples by atomic emission spectrometry based on solution cathode glow discharge
    Li, Hongwei
    Zu, Wenchuan
    Liu, Fengkui
    Wang, Yu
    Yang, Yin
    Yang, Xiaotao
    Liu, Cong
    SPECTROCHIMICA ACTA PART B-ATOMIC SPECTROSCOPY, 2019, 152 : 25 - 29
  • [26] ANALYSIS OF PIGMENTED POLYMER-COATINGS WITH RADIOFREQUENCY GLOW-DISCHARGE OPTICAL-EMISSION SPECTROMETRY
    JONES, DG
    PAYLING, R
    GOWER, SA
    BOGE, EM
    JOURNAL OF ANALYTICAL ATOMIC SPECTROMETRY, 1994, 9 (03) : 369 - 373
  • [27] FURTHER IMPROVEMENTS IN CALIBRATION TECHNIQUES FOR DEPTH PROFILING WITH GLOW-DISCHARGE OPTICAL-EMISSION SPECTROMETRY
    BENGTSON, A
    EKLUND, A
    LUNDHOLM, M
    SARIC, A
    JOURNAL OF ANALYTICAL ATOMIC SPECTROMETRY, 1990, 5 (06) : 563 - 567
  • [28] CORRECTING FOR BACKGROUND AND INTER-ELEMENT EFFECTS IN GLOW-DISCHARGE OPTICAL-EMISSION SPECTROMETRY
    PAYLING, R
    BROWN, NV
    GOWER, SA
    JOURNAL OF ANALYTICAL ATOMIC SPECTROMETRY, 1994, 9 (03) : 363 - 368
  • [29] Ultra-sensitive determination of antimony valence by solution cathode glow discharge optical emission spectrometry coupled with hydride generation
    Zhao, Mingyue
    Peng, Xiaoxu
    Yang, Bingcheng
    Wang, Zheng
    JOURNAL OF ANALYTICAL ATOMIC SPECTROMETRY, 2020, 35 (06) : 1148 - 1155
  • [30] Classification of bottled mineral waters using solution cathode glow discharge optical emission spectroscopy and chemometrics methods
    Zheng, Peichao
    Wang, Ningshen
    Wang, Jinmei
    Mao, Xuefeng
    Lai, Chunhong
    Zhong, Chao
    Li, Weiqi
    Luo, Yuanjiang
    MICROCHEMICAL JOURNAL, 2019, 151