ISFET glucose sensor system with fast recovery characteristics by employing electrolysis

被引:54
作者
Park, KY
Choi, SB
Lee, M
Sohn, BK
Choi, SY
机构
[1] Kyungpook Natl Univ, Sch Elect Engn & Comp Sci, Puk Gu, Taegu 702701, South Korea
[2] Kyungpook Natl Univ, Dept Sensor Engn, Taegu 702701, South Korea
关键词
enzyme field-effect transistor; glucose sensor; ISFET; electrolysis;
D O I
10.1016/S0925-4005(01)01049-8
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
The long recovery-time involved with an ion-sensitive field-effect transistor (ISFET)-based glucose sensor, in relation to long-term monitoring is one of the main problems preventing widespread commercialization. Accordingly, this paper proposes a new ISFET glucose sensor system with rapid recovery characteristics by an electrolysis method. In addition, since practical application requires a detailed understanding of the influence of electrolysis, the current article also investigates the recovery characteristics of an ISFET glucose sensor including the long-term stability, and the recovery time when using electrolysis. As a result, it was found that the recovery time of an ISFET glucose sensor was reduced to less than two minutes when using electrolysis in contrast to the 10 or 20 min required using a conventional method. Furthermore, no variation in the long-term stability at biasing the reduction potential was observed for 15 days. (C) 2002 Elsevier Science B.V. All rights reserved.
引用
收藏
页码:90 / 97
页数:8
相关论文
共 13 条
[2]  
Bergveld P., 1988, Analytical and Biomedical Applications of Ion-Selective Field-Effect Transistors
[3]  
HAL REG, 1995, SENSOR ACTUAT B-CHEM, V23, P17
[4]   AFFINITY BIOSENSORS [J].
LEECH, D .
CHEMICAL SOCIETY REVIEWS, 1994, 23 (03) :205-213
[5]   A micro-planar amperometric glucose sensor unsusceptible to interference species [J].
Matsumoto, T ;
Furusawa, M ;
Fujiwara, H ;
Matsumoto, Y ;
Ito, N .
SENSORS AND ACTUATORS B-CHEMICAL, 1998, 49 (1-2) :68-72
[6]  
PARK KY, 2000, J KOREAN SENSORS SOC, V9, P203
[7]   COMBINATION OF AMPEROMETRIC AND POTENTIOMETRIC SENSOR PRINCIPLES FOR ONLINE BLOOD MONITORING [J].
SCHELTER, W ;
GUMBRECHT, W ;
MONTAG, B ;
SYKORA, U ;
ERHARDT, W .
SENSORS AND ACTUATORS B-CHEMICAL, 1992, 6 (1-3) :91-95
[8]   ISFET glucose sensor based on a new principle using the electrolysis of hydrogen peroxide [J].
Seo, HI ;
Kim, CS ;
Sohn, BK ;
Yeow, T ;
Son, MT ;
Haskard, M .
SENSORS AND ACTUATORS B-CHEMICAL, 1997, 40 (01) :1-5
[9]   ISFET glucose and sucrose sensors by using platinum electrode and photo-crosslinkable polymers [J].
Sohn, BK ;
Cho, BW ;
Kim, CS ;
Kwon, DH .
SENSORS AND ACTUATORS B-CHEMICAL, 1997, 41 (1-3) :7-11
[10]   A new pH-ISFET based dissolved oxygen sensor by employing electrolysis of oxygen [J].
Sohn, BK ;
Kim, CS .