Comparison of pinhole and SPECT 99mTc-MIBI imaging in primary hyperparathyroidism

被引:25
|
作者
Shon, Ivan A. Ho [1 ]
Yan, Warren [1 ]
Roach, Paul J. [1 ,3 ]
Bernard, Elizabeth J. [1 ,3 ]
Shields, Melissa [3 ]
Sywak, Mark [2 ]
Sidhu, Stanley [2 ]
Delbridge, Leigh W. [2 ]
机构
[1] Royal N Shore Hosp, Dept Nucl Med, St Leonards, NSW 2065, Australia
[2] Royal N Shore Hosp, Dept Surg, St Leonards, NSW 2065, Australia
[3] Dept N Shore Radiol & Nucl Med, St Leonards, NSW, Australia
关键词
hyperparathyroidism; parathyroid adenoma; primary; radionuclide imaging; single photon emission computed tomography; technetium-99m-methoxy-2-isobutylisonitrile;
D O I
10.1097/MNM.0b013e328309789e
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Objective This study aims to compare dual tracer, dual phase pinhole technetium-99m labelled 2-methoxyisobutylisonitrile (Tc-99m-MIBI) imaging (including oblique imaging), with single photon emission computed tomography (SPECT) and dual phase planar Tc-99m-MIBI images, and combined SPECT, dual phase planar Tc-99m-MIBI images and anterior pinhole thyroid images for the localization of parathyroid adenomas in the neck in primary hyperparathyroidism. Methods Sixty-two patients underwent Tc-99m-MIBI dual phase, anterior and anterior oblique pinhole images of the neck, anterior planar images of the neck and chest and early phase neck/chest SPECT followed by [Tc-99m] pertechnetate anterior and anterior oblique pinhole thyroid images. Images were reviewed by consensus in three combinations - dual phase anterior and anterior oblique pinhole Tc-99m-MIBI images and pinhole thyroid images; SPECT and dual phase planar Tc-99m-MIBI images and combined SPECT, dual phase planar Tc-99m-MIBI images and anterior pinhole thyroid images. Results For 52 parathyroid adenomas in 50 patients, the sensitivity of dual phase anterior and anterior oblique pinhole Tc-99m-images and pinhole thyroid images was 81%. Significantly lower sensitivities were observed with SPECT and dual phase planar Tc-99m-MIBI images (54%, P=0.0005) and combined SPECT, dual phase planar Tc-99m-MIBI images and anterior pinhole thyroid images (65%, P=0.0209). The positive predictive value for all imaging combinations was 88-92%. Conclusion Dual phase anterior and anterior oblique pinhole 99mTc-MIBI images and pinhole thyroid images are significantly more sensitive than imaging combinations that included SPECT and remains the optimal imaging protocol for the localization of parathyroid adenomas in the neck in primary hyperparathyroidism. Nucl Med Commun 29:949-955 (C) 2008 Wolters Kluwer Health vertical bar Lippincott Williams & Wilkins.
引用
收藏
页码:949 / 955
页数:7
相关论文
共 50 条
  • [31] Factors associated with negative 99mTc-MIBI scanning in patients with primary hyperparathyroidism
    Carral, F.
    Jimenez, A., I
    Tome, M.
    Alvarez, J.
    Diez, A.
    Partida, F.
    Ayala, C.
    REVISTA ESPANOLA DE MEDICINA NUCLEAR E IMAGEN MOLECULAR, 2021, 40 (04): : 222 - 228
  • [32] Cystic parathyroid adenoma: Primary hyperparathyroidism without 99mTc-MIBI uptake
    Oliver Guillen, J. R.
    de la Cueva Barrao, L.
    Miguelena Bobadilla, J. M.
    Gimenez Maurel, T.
    Hernando Almudi, E.
    Casamayor Franco, M. C.
    REVISTA ESPANOLA DE MEDICINA NUCLEAR E IMAGEN MOLECULAR, 2017, 36 (04): : 254 - 256
  • [33] 99mTc-MIBI SPECT in the study of brain lymphoma
    Mena, L. M.
    Martin, F.
    Bravo, C.
    Santos, A. B.
    Escabia, M. V.
    Jimenez, I. R.
    REVISTA ESPANOLA DE MEDICINA NUCLEAR, 2011, 30 (06): : 380 - 381
  • [34] Contrast Enhanced [99mTc] MIBI SPECT/CT in Primary Hyperparathyroidism
    Komek, Halil
    Yilmaz, Edip Erdal
    Cakabay, Bahri
    Altindag, Serdar
    Ozdemir, Nilufer
    Can, Canan
    Kepenek, Ferat
    ANNALI ITALIANI DI CHIRURGIA, 2018, 89 (05) : 379 - 384
  • [35] The role of 99mTc-MIBI SPECT-CT in reoperation therapy of persistent hyperparathyroidism patients
    Yin, Lijie
    Guo, Dandan
    Liu, Jie
    Yan, Jue
    OPEN MEDICINE, 2015, 10 (01): : 462 - 467
  • [36] Strategy of operative treatment of hyperparathyroidism using US scan and 99mTc-MIBI SPECT/CT
    Noda, Satoru
    Onoda, Naoyoshi
    Kashiwagi, Shinichiro
    Kawajiri, Hidemi
    Takashima, Tsutomu
    Ishikawa, Tetsuro
    Yoshida, Atsushi
    Higashiyama, Shigeaki
    Kawabe, Joji
    Imanishi, Yasuo
    Tahara, Hideki
    Inaba, Masaaki
    Osawa, Masahiko
    Hirakawa, Kosei
    ENDOCRINE JOURNAL, 2014, 61 (03) : 225 - 230
  • [37] Comparison of 99mTc-MIBI mammoscintigraphy findings obtained on planar and SPECT images
    Titskaya, A. A.
    Chernov, V. I.
    Sinilkin, I. G.
    Slonimskaya, E. M.
    EUROPEAN JOURNAL OF NUCLEAR MEDICINE AND MOLECULAR IMAGING, 2008, 35 : S259 - S259
  • [38] Effect of calcium channel blockers on the sensitivity of preoperative 99mTc-MIBI SPECT for hyperparathyroidism -: Discussion
    Heller, KS
    Singh, B
    Irvin, GL
    Mitchell, B
    Friedman
    SURGERY, 2004, 136 (06) : 1204 - 1204
  • [39] Primary hyperparathyroidism regression associated to cinacalcet therapy proved by 99mTc-MIBI scintigraphy
    Coloma, A.
    Hernandez-Estrada, S.
    Bowles, H.
    Sanchez, N.
    Fuster, D.
    Torregrosa, J.
    REVISTA ESPANOLA DE MEDICINA NUCLEAR E IMAGEN MOLECULAR, 2017, 36 (02): : 120 - 121
  • [40] 99mTc-MIBI scintigraphy and cell cycle in secondary hyperparathyroidism
    Fuster, D
    Pons, F
    Torregrosa, JV
    Laterza, C
    Mateos, JJ
    Martin, F
    Vidal, S
    Herranz, R
    EUROPEAN JOURNAL OF NUCLEAR MEDICINE, 1999, 26 (09): : 1084 - 1084