Liquid chromatography-isotope dilution tandem mass spectrometry method for the measurement of urea in human serum and assignment of reference values to external quality assessment samples

被引:5
|
作者
Yong, Sharon [1 ]
Liu, Hong [1 ]
Wong, Lingkai [1 ]
Teo, Hui Ling [1 ]
Chen, Yizhao [1 ]
Liu, Qinde [1 ]
Teo, Tang Lin [1 ]
Lee, Tong Kooi [1 ]
机构
[1] Chem Metrol Div, Appl Sci Grp, Hlth Sci Author, Singapore, Singapore
关键词
Urea; Human serum; Isotope dilution mass spectrometry; Liquid chromatography-isotope dilution; tandem mass spectrometry; Reference measurement procedure; External quality assessment;
D O I
10.1016/j.ijms.2017.01.013
中图分类号
O64 [物理化学(理论化学)、化学物理学]; O56 [分子物理学、原子物理学];
学科分类号
070203 ; 070304 ; 081704 ; 1406 ;
摘要
Reference measurement procedures are essential for the accurate measurement of urea in human serum. The widely used gas chromatography-isotope dilution mass spectrometry (GC-IDMS) involves tedious derivatization processes. We developed and validated a simple liquid chromatography-isotope dilution tandem mass spectrometry (LC-IDMS/MS) method for the measurement of urea in human serum. In this method, C-13,N-15(2)-urea was spiked into serum samples as the internal standard, and pure urea certified reference material (CRM) was used as the calibration standard to ensure the traceability of the measurements to the International System of Units (SI). The sample cleanup procedure involved a simple protein precipitation step, followed by centrifugation and filtration. The developed LC-IDMS/MS method was found to be accurate and highly reproducible. The coefficient of variation (CV) was generally below 1%. Good accuracy was demonstrated by measuring serum CRM5 from two National Metrology Institutes (NMIs) with relative deviations from the certified values being close to or below 1%, and measurement uncertainties of 2.0%-2.8%. The method was used to assign the reference values of serum urea in ten cycles of an external quality assessment (EQA) program for clinical laboratories over a period of four years. The robust mean of the reported results from clinical laboratories using automated analyzers were close to the assigned values from LC-IDMS/MS method. With good accuracy and precision, as well as simple sample cleanup procedure, the developed LCIDMS/MS method can be used as a reference measurement procedure for the measurement of urea in human serum. (C) 2017 Elsevier B.V. All rights reserved.
引用
收藏
页码:87 / 93
页数:7
相关论文
共 50 条
  • [21] Automated isotope dilution liquid chromatography-tandem mass spectrometry with on-line dilution and solid phase extraction for the measurement of cortisol in human serum sample
    Kawaguchi, Migaku
    Eyama, Sakae
    Takatsu, Akiko
    JOURNAL OF PHARMACEUTICAL AND BIOMEDICAL ANALYSIS, 2014, 96 : 220 - 223
  • [22] Measurement of Cortisol in Human Serum by Isotope Dilution Liquid Chromatography/Mass Spectrometry and a Comparison through a Proficiency Testing
    Lee, Hwashim
    Shin, Hyesun
    Park, Sangryoul
    Park, Chang Joon
    BULLETIN OF THE KOREAN CHEMICAL SOCIETY, 2010, 31 (05) : 1149 - 1154
  • [23] Quantification of monofluoroacetate and monochloroacetate in human urine by isotope dilution liquid chromatography tandem mass spectrometry
    Hamelin, Elizabeth I.
    Mawhinney, Douglas B.
    Parry, Ritchard
    Kobelski, Robert J.
    JOURNAL OF CHROMATOGRAPHY B-ANALYTICAL TECHNOLOGIES IN THE BIOMEDICAL AND LIFE SCIENCES, 2010, 878 (15-16): : 1045 - 1050
  • [24] Measurement of serum 17-hydroxyprogesterone using isotope dilution liquid chromatography-tandem mass spectrometry candidate reference method and evaluation of the performance for three routine methods
    Long, Qichen
    Zhang, Tianjiao
    Yan, Ying
    Zhao, Haijian
    Zhou, Weiyan
    Zeng, Jie
    Li, Shuijun
    Zhang, Jiangtao
    Zeng, Qingzhang
    Zhao, Beibei
    Zhang, Chuanbao
    Chen, Wenxiang
    CLINICAL CHEMISTRY AND LABORATORY MEDICINE, 2021, 59 (03) : 523 - 532
  • [25] Liquid chromatography-tandem mass spectrometry measurement of 26 steroid hormones in human serum and plasma samples
    Jia, Yongjuan
    Liu, Xingli
    Xu, Li
    Zhao, Jinbao
    Ni, Junjun
    Zhang, Yuanyuan
    JOURNAL OF SEPARATION SCIENCE, 2021, 44 (12) : 2358 - 2370
  • [26] Stable isotope-dilution liquid chromatography/tandem mass spectrometry method for determination of thyroxine in saliva
    Higashi, Tatsuya
    Ichikawa, Takuya
    Shimizu, Chikara
    Nagai, So
    Inagaki, Shinsuke
    Min, Jun Zhe
    Chiba, Hitoshi
    Ikegawa, Shigeo
    Toyo'oka, Toshimasa
    JOURNAL OF CHROMATOGRAPHY B-ANALYTICAL TECHNOLOGIES IN THE BIOMEDICAL AND LIFE SCIENCES, 2011, 879 (13-14): : 1013 - 1017
  • [27] Quantification of Folic Acid in Human Serum Using Isotope Dilution Ultra-High-Pressure Liquid Chromatography/Mass Spectrometry
    Hyung, Seok-Won
    Lee, Joonhee
    Seo, Hye-Young
    Kim, Byungjoo
    BULLETIN OF THE KOREAN CHEMICAL SOCIETY, 2018, 39 (01): : 105 - 110
  • [28] SEPARATION AND QUANTIFICATION OF WATER-SOLUBLE CELLULAR METABOLITES IN CLOSTRIDIUM THERMOCELLUM USING LIQUID CHROMATOGRAPHY-ISOTOPE DILUTION TANDEM MASS SPECTROMETRY
    Cui, Jiatao
    Zhang, Jingtao
    Zhu, Xinshu
    Bai, Fali
    Feng, Yingang
    Guan, Wenna
    Cui, Qiu
    ANALYTICAL LETTERS, 2013, 46 (17) : 2767 - 2786
  • [29] Development of a matrix-based candidate reference material for human glycated albumin using isotope-dilution liquid chromatography/tandem mass spectrometry
    Zhang, Rui
    Zhao, Rui
    Ren, Wenhua
    Wang, Mo
    Fan, Gaowei
    Shi, Jie
    Song, Yichuan
    Zhou, Xiaojie
    Wang, Qingtao
    ANALYTICA CHIMICA ACTA, 2023, 1239
  • [30] Development of isotope dilution-liquid chromatography/tandem mass spectrometry as a candidate reference method for the accurate determination of patulin in apple products
    Gab-Allah, Mohamed A.
    Choi, Kihwan
    Kim, Byungjoo
    ANALYTICAL AND BIOANALYTICAL CHEMISTRY, 2022, 414 (05) : 1867 - 1879