Liquid chromatography-isotope dilution tandem mass spectrometry method for the measurement of urea in human serum and assignment of reference values to external quality assessment samples

被引:5
|
作者
Yong, Sharon [1 ]
Liu, Hong [1 ]
Wong, Lingkai [1 ]
Teo, Hui Ling [1 ]
Chen, Yizhao [1 ]
Liu, Qinde [1 ]
Teo, Tang Lin [1 ]
Lee, Tong Kooi [1 ]
机构
[1] Chem Metrol Div, Appl Sci Grp, Hlth Sci Author, Singapore, Singapore
关键词
Urea; Human serum; Isotope dilution mass spectrometry; Liquid chromatography-isotope dilution; tandem mass spectrometry; Reference measurement procedure; External quality assessment;
D O I
10.1016/j.ijms.2017.01.013
中图分类号
O64 [物理化学(理论化学)、化学物理学]; O56 [分子物理学、原子物理学];
学科分类号
070203 ; 070304 ; 081704 ; 1406 ;
摘要
Reference measurement procedures are essential for the accurate measurement of urea in human serum. The widely used gas chromatography-isotope dilution mass spectrometry (GC-IDMS) involves tedious derivatization processes. We developed and validated a simple liquid chromatography-isotope dilution tandem mass spectrometry (LC-IDMS/MS) method for the measurement of urea in human serum. In this method, C-13,N-15(2)-urea was spiked into serum samples as the internal standard, and pure urea certified reference material (CRM) was used as the calibration standard to ensure the traceability of the measurements to the International System of Units (SI). The sample cleanup procedure involved a simple protein precipitation step, followed by centrifugation and filtration. The developed LC-IDMS/MS method was found to be accurate and highly reproducible. The coefficient of variation (CV) was generally below 1%. Good accuracy was demonstrated by measuring serum CRM5 from two National Metrology Institutes (NMIs) with relative deviations from the certified values being close to or below 1%, and measurement uncertainties of 2.0%-2.8%. The method was used to assign the reference values of serum urea in ten cycles of an external quality assessment (EQA) program for clinical laboratories over a period of four years. The robust mean of the reported results from clinical laboratories using automated analyzers were close to the assigned values from LC-IDMS/MS method. With good accuracy and precision, as well as simple sample cleanup procedure, the developed LCIDMS/MS method can be used as a reference measurement procedure for the measurement of urea in human serum. (C) 2017 Elsevier B.V. All rights reserved.
引用
收藏
页码:87 / 93
页数:7
相关论文
共 50 条
  • [1] Measurement of urine albumin by liquid chromatography-isotope dilution tandem mass spectrometry and its application to value assignment of external quality assessment samples and certification of reference materials
    Chen, Yizhao
    Liu, Hong
    Loh, Tze Ping
    Liu, Qinde
    Teo, Tang Lin
    Lee, Tong Kooi
    Sethi, Sunil Kumar
    CLINICAL CHEMISTRY AND LABORATORY MEDICINE, 2021, 59 (04) : 711 - 720
  • [2] DETERMINATION OF UREA IN MILK BY LIQUID CHROMATOGRAPHY-ISOTOPE DILUTION MASS SPECTROMETRY
    Dai, Xinhua
    Zhao, Yingchen
    Li, Ming
    Fang, Xiang
    Li, Xiaomin
    Li, Hongmei
    Xu, Bei
    ANALYTICAL LETTERS, 2012, 45 (12) : 1557 - 1565
  • [3] Developing a reference measurement procedure for free glycerol in human serum by two-step gas chromatography-isotope dilution mass spectrometry
    Chen, Yizhao
    Teo, Hui Ling
    Liu, Qinde
    Lee, Tong Kooi
    CLINICAL BIOCHEMISTRY, 2015, 48 (13-14) : 897 - 903
  • [4] An isotope dilution liquid chromatography-tandem mass spectrometry candidate reference measurement procedure for aldosterone measurement in human plasma
    Qiaoxuan Zhang
    Liqiao Han
    Songbai Zheng
    Fen Ouyang
    Xiaobin Wu
    Jun Yan
    Min Zhan
    Peifeng Ke
    Junhua Zhuang
    Xianzhang Huang
    Analytical and Bioanalytical Chemistry, 2021, 413 : 4471 - 4481
  • [5] An isotope dilution liquid chromatography-tandem mass spectrometry candidate reference measurement procedure for aldosterone measurement in human plasma
    Zhang, Qiaoxuan
    Han, Liqiao
    Zheng, Songbai
    Ouyang, Fen
    Wu, Xiaobin
    Yan, Jun
    Zhan, Min
    Ke, Peifeng
    Zhuang, Junhua
    Huang, Xianzhang
    ANALYTICAL AND BIOANALYTICAL CHEMISTRY, 2021, 413 (17) : 4471 - 4481
  • [6] Determination of serum glucose by isotope dilution liquid chromatography-tandem mass spectrometry: a candidate reference measurement procedure
    Zhang, Tianjiao
    Zhang, Chuanbao
    Zhao, Haijian
    Zeng, Jie
    Zhang, Jiangtao
    Zhou, Weiyan
    Yan, Ying
    Wang, Yufei
    Wang, Mo
    Chen, Wenxiang
    ANALYTICAL AND BIOANALYTICAL CHEMISTRY, 2016, 408 (26) : 7403 - 7411
  • [7] Candidate reference measurement procedure for determination of urea in serum by liquid chromatography-tandem mass spectrometry
    Han, Liqiao
    Huang, Xiaoting
    Zhang, Lu
    Zhang, Qiaoxuan
    Wang, Jianbing
    Lin, Haibiao
    Yan, Jun
    Zhuang, Junhua
    Huang, Xianzhang
    JOURNAL OF PHARMACEUTICAL AND BIOMEDICAL ANALYSIS, 2019, 162 : 124 - 129
  • [8] Development and evaluation of a candidate reference measurement procedure for the determination of testosterone in human serum using isotope dilution liquid chromatography/tandem mass spectrometry
    Susan S.-C. Tai
    Bei Xu
    Michael J. Welch
    Karen W. Phinney
    Analytical and Bioanalytical Chemistry, 2007, 388 : 1087 - 1094
  • [9] Development and evaluation of a candidate reference measurement procedure for the determination of testosterone in human serum using isotope dilution liquid chromatography/tandem mass spectrometry
    Tai, Susan S. -C.
    Xu, Bei
    Welch, Michael J.
    Phinney, Karen W.
    ANALYTICAL AND BIOANALYTICAL CHEMISTRY, 2007, 388 (5-6) : 1087 - 1094
  • [10] An improved reference measurement procedure for triglycerides and total glycerides in human serum by isotope dilution gas chromatography-mass spectrometry
    Chen, Yizhao
    Liu, Qinde
    Yong, Sharon
    Teo, Hui Ling
    Lee, Tong Kooi
    CLINICA CHIMICA ACTA, 2014, 428 : 20 - 25