Accuracy of the new ICare rebound tonometer vs. other portable tonometers in healthy eyes

被引:78
作者
García-Resúa, C
González-Meijome, JM
Gilino, J
Yebra-Pimentel, E
机构
[1] Univ Santiago Compostela, Escuela Opt & Optometria, Optometry Grp, Dept Appl Phys, Santiago De Compostela 15706, Galicia, Spain
[2] Univ Minho, Dept Phys Optometry, Braga, Portugal
关键词
intraocular pressure; Perkins tonometry; rebound tonometry; ICare; Tono-Pen XL; portable tonometers; accuracy;
D O I
10.1097/01.opx.0000200673.96758.7b
中图分类号
R77 [眼科学];
学科分类号
100212 ;
摘要
Purpose. The ICare (Tiolat Oy, Helsinki, Finland) is a new portable tonometer that measures intraocular pressure (IOP) with a new rebound method, in which a very light probe is used to make momentary contact with the cornea in slow motion. The purpose of this study is to assess the accuracy of the ICare IOP measurements by comparing them against other portable tonometers: Perkins applanation tonometer and Tono-Pen XL digital tonometer (Medtronic Solan, Jacksonville, FL). Methods. Sixty-five young subjects were assessed with each of the tonometers. ICare tonometry was performed first, followed by Perkins applanation tonometry and Tono-Pen XL in a random order. Regression analysis was used to evaluate the relationship between the Perkins tonometer and the remaining tonometers used in this study. Tonometers were also compared by plotting the difference between the methods against the mean. The hypothesis of zero bias was examined by a paired t test and the 95% limits of agreement (LoA) were calculated. Results. ICare and Tono-Pen XL significantly overestimate IOP when compared with Perkins applanation tonometry. The mean of the difference between Perkins and ICare and Perkins and Tono-Pen XL was (mean standard deviation) -3.35 +/- 2.28 mm Hg and -2.78 +/- 2.53 mm Hg, respectively. The 95% LoA between Perkins tonometry and ICare tonometry were between -7.81 and +1.12 and between Perkins tonometry and Tono-Pen XL tonometry between -7.74 and +2.18. Conclusions. Compared with Perkins tonometry, the ICare tonometer allows clinicians to estimate IOP with a portable, rapid, and noninvasive method with similar reliability to that offered by Tono-Pen XL. Clinicians should be aware of the systematic overestimation of IOP with the ICare. Further research is needed to evaluate the performance of rebound tonometry in populations with higher IOP and assess the reliability of this technique in the early detection and follow up of glaucomatous patients.
引用
收藏
页码:102 / 107
页数:6
相关论文
共 28 条
[11]   Measuring intraocular pressure with the Pulsair 3000 and Rebound tonometers in elderly patients without an anesthetic [J].
Kontiola, A ;
Puska, P .
GRAEFES ARCHIVE FOR CLINICAL AND EXPERIMENTAL OPHTHALMOLOGY, 2004, 242 (01) :3-7
[12]   A new induction-based impact method for measuring intraocular pressure [J].
Kontiola, AI .
ACTA OPHTHALMOLOGICA SCANDINAVICA, 2000, 78 (02) :142-145
[13]   The induction/impact tonometer: A new instrument to measure intraocular pressure in the rat [J].
Kontiola, AI ;
Goldblum, D ;
Mittag, T ;
Danias, J .
EXPERIMENTAL EYE RESEARCH, 2001, 73 (06) :781-785
[14]  
KOONER KS, 1992, ANN OPHTHALMOL, V24, P29
[15]  
KRAKAU CET, 1971, ACTA OPHTHALMOL, V49, P611
[16]  
LAGERLOF O, 1990, ACTA OPHTHALMOL, V68, P221
[17]  
Lim Li, 1997, CLAO Journal, V23, P130
[18]   ACCURACY OF INTRAOCULAR-PRESSURE MEASUREMENTS WITH 2 DIFFERENT TONOMETERS THROUGH BANDAGE CONTACT-LENSES [J].
MARK, LK ;
ASBELL, PA ;
TORRES, MA ;
FAILLA, SJ .
CORNEA, 1992, 11 (04) :277-281
[19]  
MARQUEZ MTA, 2003, ARCH SOC ESP OFTALMO, V78, P189, DOI DOI 10.4321/S0365-66912003000400004
[20]   CLINICAL COMPARISON OF THE PROTON AND TONO-PEN TONOMETERS WITH THE GOLDMANN APPLANATION TONOMETER [J].
MIDELFART, A ;
WIGERS, A .
BRITISH JOURNAL OF OPHTHALMOLOGY, 1994, 78 (12) :895-898