Graphene ultrathin film electrodes modified with bismuth nanoparticles and polyaniline porous layers for detection of lead and cadmium ions in acetate buffer solutions

被引:43
作者
Wang, Zhaomeng [1 ]
Li, Lin [1 ]
Liu, Erjia [1 ]
机构
[1] Nanyang Technol Univ, Sch Mech & Aerosp Engn, Singapore 639798, Singapore
关键词
Graphene; Solid-state carbon diffusion; Bismuth nanoparticle; Polyaniline porous layer; Square wave anodic stripping voltammetry; Trace lead and cadmium; ANODIC-STRIPPING VOLTAMMETRY; CONDUCTING POLYMERS; CARBON ELECTRODES; SENSING PLATFORM; COMPOSITE FILM; METAL-IONS; OXIDE; NANOCOMPOSITES; CORROSION; TOXICITY;
D O I
10.1016/j.tsf.2013.02.098
中图分类号
T [工业技术];
学科分类号
08 ;
摘要
Graphene ultrathin films were synthesized by means of solid-state carbon diffusion from amorphous carbon (a-C) thin layers deposited on silicon substrates, which was catalyzed by nickel layers coated on the top of the a-C layers. The graphene films were used as working electrodes that were modified by a polyaniline (PANI) porous layer together with in-situ deposited bismuth (Bi) nanoparticles for the detection of trace heavy metal ions (Pb2+ and Cd2+) in acetate buffer solutions (pH 5.3) with square wave anodic stripping voltammetry. The graphene electrodes modified with PANI porous layers and Bi nanoparticles had excellent repeatability, ultrahigh sensitivity (as low as 0.33 nM) and good resistance to passivation caused by the surface active species adsorbed on the electrode surfaces. (C) 2013 Elsevier B. V. All rights reserved.
引用
收藏
页码:362 / 367
页数:6
相关论文