A Simple Isotope Dilution Electrospray Ionization Tandem Mass Spectrometry Method for the Determination of Free Phenytoin

被引:7
作者
Garg, Uttam [1 ]
Peat, Judy
Frazee, Clinton
Trang Nguyen
Ferguson, Angela M.
机构
[1] Childrens Mercy Hosp & Clin, Dept Pathol & Lab Med, Kansas City, MO 64108 USA
关键词
phenytoin; free phenytoin; immunoassay; tandem mass spectrometry; IMMUNOASSAY MEASUREMENTS; UREMIC PATIENTS; HUMAN PLASMA; ASSAY; CHROMATOGRAPHY; INTERFERENCE; PREDICTION;
D O I
10.1097/FTD.0b013e31828f5337
中图分类号
R446 [实验室诊断]; R-33 [实验医学、医学实验];
学科分类号
1001 ;
摘要
Background:Phenytoin (diphenylhydantoin) is an anticonvulsant drug frequently prescribed for the treatment of many types of seizures. Because the drug is highly protein bound (90%-95%) and many conditions can displace the drug from proteins, the measurement of free phenytoin is warranted. Due to the unavailability of free phenytoin assays in many chemistry analyzers or limitations of immunoassays, chromatographic methods such as liquid chromatography-tandem mass spectrometry (LC-MS-MS) are preferred for the assay of free phenytoin.Methods:The sample preparation involved ultrafiltration of serum or plasma to separate free phenytoin. Acetonitrile containing internal standard, phenytoin-d10, was added to the ultrafiltrate. The samples were centrifuged, and supernatants were injected into an LC-MS-MS involving reverse phase Ultra BiPh 5-m x 50 x 2.1-mm analytical column, and mobile phases, water and methanol containing 0.1% formic acid. The mass/charge (m/z) transitions were as follows: phenytoin -253.0 > 182.2 and 253.0 > 104.00; phenytoin-d10 -263.2 > 192.12.Results:Linearity of the method ranged from 0.1 to 4.0 g/mL. Within-run and between-run imprecision values were <5% and <10%, respectively. The samples were stable for 2 weeks at 4 degrees C and 4 weeks at -20 degrees C. The method compared well with the laborious liquid-liquid extraction method and did not show any significant ion suppression or enhancement.Conclusions:A simple LC-MS-MS method was developed for the assay of free phenytoin. The method does not require laborious liquid-liquid or solid-phase extraction. The method has high analytical sensitivity, low imprecision, and a wide analytical measurement range.
引用
收藏
页码:831 / 835
页数:5
相关论文
共 20 条
[1]   Determination of free levels of phenytoin in human plasma by liquid chromatography/tandem mass spectrometry [J].
Bardin, S ;
Ottinger, JC ;
Breau, AP ;
O'Shea, TJ .
JOURNAL OF PHARMACEUTICAL AND BIOMEDICAL ANALYSIS, 2000, 23 (2-3) :573-579
[2]  
BECK DE, 1987, CLIN PHARMACY, V6, P888
[3]   Unexpected suppression of free phenytoin concentration by salicylate in uremic sera due to the presence of inhibitors: Maldi mass spectrometric determination of molecular weight range of inhibitors [J].
Biddle, DA ;
Wells, A ;
Dasgupta, A .
LIFE SCIENCES, 1999, 66 (02) :143-151
[4]   Undetectable phenytoin serum levels by an automated particle-enhanced turbidimetric inhibition immunoassay in a patient with monoclonal IgMλ [J].
Brauchli, Yolanda B. ;
Scholer, Andre ;
Schwietert, Martin ;
Kraehenbuehl, Stephan .
CLINICA CHIMICA ACTA, 2008, 389 (1-2) :174-176
[5]  
Broussard LA, 2007, THERAPEUTIC DRUG MON, V3rd, P41
[6]   PREDICTION OF FREE PHENYTOIN LEVELS BASED ON [TOTAL PHENYTOIN] [ALBUMIN] RATIOS - POTENTIAL ERRORS WITH HYPOALBUMINEMIA [J].
DASGUPTA, A ;
DENNEN, DA ;
DEAN, R ;
MCLAWHON, RW .
AMERICAN JOURNAL OF CLINICAL PATHOLOGY, 1991, 95 (02) :253-256
[7]  
Datta P, 1997, CLIN CHEM, V43, P1468
[8]  
Garg U, 2010, CLIN APPL MASS SPECT
[9]  
GODOLPHIN W, 1984, THER DRUG MONIT, V6, P374
[10]  
JOERN WA, 1981, CLIN CHEM, V27, P417