The NAC proton treatment planning system

被引:4
作者
Schreuder, AN
Jones, DTL
Symons, JE
De Kock, EA
Hough, JK
Wilson, J
Vernimmen, FJA
Schlegel, W
Höss, A
Lee, M
机构
[1] Natl Accelerator Ctr, ZA-7131 Faure, South Africa
[2] Groote Schuur Hosp, Dept Radiat Oncol, ZA-7925 Cape Town, South Africa
[3] Tygerberg Hosp, ZA-7505 Tygerberg, South Africa
[4] Deutsch Krebsforschungszentrum, D-69120 Heidelberg, Germany
[5] Inst Canc Res, Sutton SM2 5PT, Surrey, England
[6] Royal Marsden Hosp, Sutton SM2 5PT, Surrey, England
关键词
proton therapy; dose calculation; treatment planning; rayline-tracing;
D O I
10.1007/BF03038876
中图分类号
R73 [肿瘤学];
学科分类号
100214 ;
摘要
A three-dimensional proton treatment planning system called PROXELPLAN has been used at the National Accelerator Centre (NAC) since October 1994. This system is entirely based on the VOXELPLAN planning system, developed at the Deutches Krebsforschungszentrum (DKFZ), Heidelberg, Germany. The VOXELPLAN system provides the treatment planning infrastructure while the proton dose distributions are calculated using a software module that was initially developed at the Royal Marsden Hospital, UK. The proton module has been extensively modified and refined. It uses a rayline-tracing algorithm which is suitable for planning current treatments but is not sufficiently dynamic to accommodate the use of compensators. A sophisticated pencil, beam algorithm is currently under development.
引用
收藏
页码:10 / 12
页数:3
相关论文
共 12 条
[1]  
BENNETT KF, 1994, S AFR J SCI, V90, P370
[2]  
GADEMAN G, 1993, RADIOTHER ONCOL, V29, P203
[3]  
Gademann G., 1993, Strahlentherapie und Onkologie, V169, P159
[4]  
JONES DTL, 1994, INT CONGR SER, V1077, P307
[5]   A COMPARISON OF PROTON AND MEGAVOLTAGE X-RAY TREATMENT PLANNING FOR PROSTATE-CANCER [J].
LEE, M ;
WYNNE, C ;
WEBB, S ;
NAHUM, AE ;
DEARNALEY, D .
RADIOTHERAPY AND ONCOLOGY, 1994, 33 (03) :239-253
[6]   AN EMPIRICAL-METHOD TO BUILD UP A MODEL OF PROTON DOSE DISTRIBUTION FOR A RADIOTHERAPY TREATMENT-PLANNING PACKAGE [J].
LEE, M ;
NAHUM, AE ;
WEBB, S .
PHYSICS IN MEDICINE AND BIOLOGY, 1993, 38 (07) :989-998
[7]   DIFFERENTIAL-PENCIL-BEAM DOSE CALCULATIONS FOR CHARGED-PARTICLES [J].
PETTI, PL .
MEDICAL PHYSICS, 1992, 19 (01) :137-149
[8]   The calibration of CT Hounsfield units for radiotherapy treatment planning [J].
Schneider, U ;
Pedroni, E ;
Lomax, A .
PHYSICS IN MEDICINE AND BIOLOGY, 1996, 41 (01) :111-124
[9]  
Schreuder AN, 1997, INT CONGR SER, V1144, P251
[10]  
SCHREUDER AN, 1996, P 14 INT C CYCL THEI, P523