Variability of the Standard Reference Height and Its Influence on the Stereometric Parameters of the Heidelberg Retina Tomograph 3

被引:10
作者
Breusegem, Christophe [1 ]
Fieuws, Steffen [2 ]
Stalmans, Ingeborg [1 ]
Zeyen, Thierry [1 ]
机构
[1] Katholieke Univ Leuven Hosp, Dept Ophthalmol, B-3000 Louvain, Belgium
[2] Catholic Univ Louvain, Ctr Biostat, B-3000 Louvain, Belgium
关键词
D O I
10.1167/iovs.08-2331
中图分类号
R77 [眼科学];
学科分类号
100212 ;
摘要
PURPOSE. To investigate the variability of the standard reference height (SRH) in consecutive Heidelberg Retina Tomograph 3 (HRT3) examinations and its influence on five main stereometric parameters. METHODS. HRT3 results of all patients attending our glaucoma center from August to October 2007 were retrospectively reviewed. Only one eye per patient with at least three consecutive HRTs and a quality control label of acceptable or better was selected. An SRH difference >= 10% compared with baseline was considered as excess SRH variability. RESULTS. A review of 641 patients yielded 150 patients (150 eyes) who met the inclusion criteria, representing a total of 556 examinations. The mean total number of HRTs per patient was 3.7 (range, 3-7), and the mean follow-up time was 3.5 years (range, 1.4-6.7). The mean proportion of consecutive HRT3 examinations with intraindividual excess SRH variability was 46% (95% CI, 40-52), whereas the mean intraindividual excess SRH variability was 24% (95% CI, 20-28). The median absolute SRH difference was 8.6% (interquartile range, 3.9%-16.2%). Furthermore, 55.5%, 75.3%, 74.7%, 0.0%, and 19.4% of the variability in rim area, rim volume, retinal nerve fiber layer thickness, cup shape measure, and height variation contour, respectively, could be attributed to SRH variability. CONCLUSIONS. There is considerable SRH variability between HRT3 examinations of the same eye, and this could explain more than half the variability of the parameters RA, RV, and RNFL. These findings indicate that changes of HRT3 parameters should be considered with caution when excess SRH variability is present. (Invest Ophthalmol Vis Sci. 2008;49: 4881-4885) DOI: 10.1167/iovs.08-2331
引用
收藏
页码:4881 / 4885
页数:5
相关论文
共 24 条
[1]   Longitudinal changes in the visual field and optic disc in glaucoma [J].
Artes, PH ;
Chauhan, BC .
PROGRESS IN RETINAL AND EYE RESEARCH, 2005, 24 (03) :333-354
[2]   Development of the standard reference plane for the Heidelberg retina tomograph [J].
Burk, ROW ;
Vihanninjoki, K ;
Bartke, T ;
Tuulonen, A ;
Airaksinen, PJ ;
Völcker, HE ;
König, JM .
GRAEFES ARCHIVE FOR CLINICAL AND EXPERIMENTAL OPHTHALMOLOGY, 2000, 238 (05) :375-384
[3]  
Chauhan BC, 2001, ARCH OPHTHALMOL-CHIC, V119, P1492
[4]   Clinician change detection viewing longitudinal stereophotographs compared to confocal scanning laser tomography in the LSU Experimental Glaucoma (LEG) study [J].
Ervin, JC ;
Lemij, HG ;
Mills, RP ;
Quigley, HA ;
Thompson, HW ;
Burgoyne, CF .
OPHTHALMOLOGY, 2002, 109 (03) :467-481
[5]   Diagnosing glaucoma progression: current practice and promising technologies [J].
Giangiacomoa, Annette ;
Garway-Heath, David ;
Caprioli, Joseph .
CURRENT OPINION IN OPHTHALMOLOGY, 2006, 17 (02) :153-162
[6]  
*HEID ENG GMBH, 2005, OP INSTR SOFTW VERS
[7]   Agreement among 3 optical imaging methods for the assessment of optic disc topography [J].
Hoffmann, EM ;
Bowd, C ;
Medeiros, FA ;
Boden, C ;
Grus, FH ;
Bourne, RR ;
Zangwill, LM ;
Weinreb, RN .
OPHTHALMOLOGY, 2005, 112 (12) :2149-2156
[8]  
Kalaboukhova L, 2006, GRAEF ARCH CLIN EXP, V244, P654, DOI 10.1007/s00417-005-0107-3
[9]   ABC of eyes - Glaucoma - 1: Diagnosis [J].
Khaw, PT ;
Shah, P ;
Elkington, AR .
BRITISH MEDICAL JOURNAL, 2004, 328 (7431) :97-99
[10]   Comparison of glaucoma progression evaluated with Heidelberg retina tomograph II versus optic nerve head stereophotographs [J].
Kourkoutas, Dimitrios ;
Buys, Yvonne M. ;
Flanagan, John G. ;
Hatch, Wendy V. ;
Balian, Carmen ;
Trope, Graham E. .
CANADIAN JOURNAL OF OPHTHALMOLOGY-JOURNAL CANADIEN D OPHTALMOLOGIE, 2007, 42 (01) :82-88