A Nitinol Based Flexor Tendon Fixation Device: Gapping and Tensile Strength Measurements in Cadaver Flexor Tendon

被引:2
作者
Reese, Shawn P. [1 ,2 ]
Kubiak, Erik N. [1 ,3 ]
机构
[1] CoNextions Inc, Salt Lake City, UT 84104 USA
[2] Univ Utah, Dept Bioengn, Salt Lake City, UT 84112 USA
[3] Univ Utah, Div Orthopaed Trauma, Dept Orthopaed Surg, Salt Lake City, UT 84112 USA
来源
JOURNAL OF BIOMECHANICAL ENGINEERING-TRANSACTIONS OF THE ASME | 2014年 / 136卷 / 01期
关键词
flexor tendon; suture; surgical repair; tendon fixation; gapping; nitinol; GAP FORMATION; SUTURE MATERIAL; REPAIR; COMPLICATIONS; SURGERY; HAND;
D O I
10.1115/1.4025779
中图分类号
Q6 [生物物理学];
学科分类号
071011 ;
摘要
In this study, a new nitinol based fixation device was investigated for use in repairing severed digital flexor tendons. The device, composed of superelastic nitinol, is tubular in shape with inward facing tines for gripping tissue. Its cellular structure was designed such that it has a large effective Poisson's ratio, which facilitates a "finger trap" effect. This allows for reduced tendon compression during a resting state (to permit vascular perfusion) and increased compression during loading (to drive the tines into the tissue for gripping). To test the feasibility of using this device for flexor tendon repair, it was tested on cadaver flexor digitorum profundus tendons. The tendons were excised, cut in the region corresponding to a zone II laceration, and repaired using the device. The device was easy to install and did not prevent the tendon from bending. Constant strain rate tensile testing revealed a mean tensile strength of 57.6 +/- 7.7 N, with a force of 53.2 +/- 7.8 N at a 2 mm gap. This exceeds the suggested primary repair strength of 45 N, which has been proposed as the necessary strength for enabling early mobilization. Although considerable future studies will be needed to determine the suitability of the new repair device for clinical use, this study demonstrates the feasibility of utilizing a tubular, nitinol repair device for flexor tendon fixation.
引用
收藏
页数:5
相关论文
共 42 条
[11]  
Griffin M, 2012, Open Orthop J, V6, P28, DOI 10.2174/1874325001206010028
[12]  
Gussous Yazeed Mazen, 2011, Hand (N Y), V6, P268, DOI 10.1007/s11552-011-9344-6
[13]   A new barbed device for repair of flexor tendons [J].
Hirpara, K. M. ;
Sullivan, P. J. ;
O'Sullivan, M. E. .
JOURNAL OF BONE AND JOINT SURGERY-BRITISH VOLUME, 2010, 92B (08) :1165-1170
[14]   Advances in the biology of zone II flexor tendon healing and adhesion formation [J].
Jaibaji, M .
ANNALS OF PLASTIC SURGERY, 2000, 45 (01) :83-92
[15]   Comparison of the holding capacity of round monofilament, round multifilament, and flat multifilament nitinol suture loops in human cadaveric flexor tendon [J].
Karjalainen, T. ;
He, M. ;
Chong, A. K. S. ;
Lim, A. Y. T. ;
Goransson, H. ;
Ryhanen, J. .
JOURNAL OF HAND SURGERY-EUROPEAN VOLUME, 2012, 37E (05) :459-463
[16]   NICKEL-TITANIUM WIRE AS A FLEXOR TENDON SUTURE MATERIAL: AN EX VIVO STUDY [J].
Karjalainen, T. ;
Goransson, H. ;
Viinikainen, A. ;
Jamsa, T. ;
Ryhanen, J. .
JOURNAL OF HAND SURGERY-EUROPEAN VOLUME, 2010, 35E (06) :469-474
[17]   Prevention of adhesions in surgery of the flexor tendons of the hand: what is the evidence? [J].
Khanna, Anil ;
Friel, Mike ;
Gougoulias, Nikolaos ;
Longo, Umile Giuseppe ;
Maffulli, Nicola .
BRITISH MEDICAL BULLETIN, 2009, 90 (01) :85-109
[18]   Biocompatibility and strength properties of nitinol shape memory alloy suture in rabbit tendon [J].
Kujala, S ;
Pajala, A ;
Kallioinen, M ;
Pramila, A ;
Tuukkanen, J ;
Ryhänen, J .
BIOMATERIALS, 2004, 25 (02) :353-358
[19]   Modern Tendon Repair Techniques [J].
Lee, Steve K. .
HAND CLINICS, 2012, 28 (04) :565-+
[20]   Complications after treatment of flexor tendon injuries [J].
Lilly, Soma I. ;
Messer, Terry M. .
JOURNAL OF THE AMERICAN ACADEMY OF ORTHOPAEDIC SURGEONS, 2006, 14 (07) :387-396