A computational study on the biomechanical factors related to stent-graft models in the thoracic aorta

被引:61
作者
Lam, S. K. [1 ,3 ]
Fung, George S. K. [2 ]
Cheng, Stephen W. K. [1 ]
Chow, K. W. [3 ]
机构
[1] Univ Hong Kong, Div Vasc Surg, Dept Surg, Hong Kong, Hong Kong, Peoples R China
[2] Univ Hong Kong, Dept Elect & Elect Engn, Hong Kong, Hong Kong, Peoples R China
[3] Univ Hong Kong, Dept Mech Engn, Hong Kong, Hong Kong, Peoples R China
关键词
Thoracic aorta; Endovascular stent-graft; Modeling of blood flow; Drag force and computational fluid dynamics;
D O I
10.1007/s11517-008-0361-8
中图分类号
TP39 [计算机的应用];
学科分类号
081203 ; 0835 ;
摘要
Endovascular aortic stent-graft is a new, minimally invasive procedure for treating thoracic aortic diseases, and has quickly evolved to be one of the standard treatments subject to anatomic constraints. This procedure involves the placement of a self-expanding stent-graft system in a high-flow thoracic aorta. Stent-graft deployment in the thoracic aorta, especially close to the aortic arch, normally experiences a significant drag force which might lead to the risk of stent-graft failure. A comprehensive investigation on the biomechanical factors affecting the drag force on a stent-graft in the thoracic aorta is thus in order, and the goal is to perform an in-depth study on the contributing biomechanical factors. Three factors affecting the deployed stent-graft are considered, namely, the internal diameter of the vessel, the starting position of the graft and the diameter of curvature of the aortic arch. Computational fluid dynamic techniques are applied to model the blood flow. The inlet velocity and outlet pressure are assumed to be pulsatile. The three-dimensional continuity equation and the time-dependent Navier-Stokes equations for an incompressible fluid were solved numerically. The drag force due to the change of momentum within the stent-graft and the shear stress were calculated and analyzed. The drag force on a stent-graft will depend critically on the internal diameter and the starting position of stent-graft deployment. Larger internal diameter leads to larger drag force and the stent-graft deployed at the more distal position may be associated with significantly diminished drag force. Smaller diameter of curvature of the aortic arch probably results in a decline of the drag force on the stent-graft, even though this factor merely causes only a modest difference. These findings may have important implications for the choice and design of stent-grafts in the future.
引用
收藏
页码:1129 / 1138
页数:10
相关论文
共 59 条
[1]  
Bergeron P, 2003, J CARDIOVASC SURG, V44, P349
[2]   The origin and significance of secondary flows in the aortic arch [J].
Black, MM ;
Hose, DR ;
Lawford, PV .
JOURNAL OF MEDICAL ENGINEERING & TECHNOLOGY, 1995, 19 (06) :192-197
[3]   A procedure to simulate coronary artery bypass graft surgery [J].
Cacho, Fernando ;
Doblare, Manuel ;
Holzapfel, Gerhard A. .
MEDICAL & BIOLOGICAL ENGINEERING & COMPUTING, 2007, 45 (09) :819-827
[4]   FLOW DYNAMICS IN THE HUMAN AORTA [J].
CHANDRAN, KB .
JOURNAL OF BIOMECHANICAL ENGINEERING-TRANSACTIONS OF THE ASME, 1993, 115 (04) :611-616
[5]   Flow patterns in an endovascular stent-graft for abdominal aortic aneurysm repair [J].
Chong, CK ;
How, TV .
JOURNAL OF BIOMECHANICS, 2004, 37 (01) :89-97
[6]   Stent-graft design: the good, the bad and the ugly [J].
Chuter, TAM .
CARDIOVASCULAR SURGERY, 2002, 10 (01) :7-13
[7]   Endograft migration one to four years after endovascular abdominal aortic aneurysm repair with the AneuRx device: A cautionary note [J].
Conners, MS ;
Sternbergh, WC ;
Carter, G ;
Tonnessen, BH ;
Yoselevitz, M ;
Money, SR .
JOURNAL OF VASCULAR SURGERY, 2002, 36 (03) :476-482
[8]   Construction of realistic branched, three-dimensional arteries suitable for computational modelling of flow. [J].
Corney, S ;
Johnston, PR ;
Kilpatrick, D .
MEDICAL & BIOLOGICAL ENGINEERING & COMPUTING, 2004, 42 (05) :660-668
[9]   The "first generation" of endovascular stent-grafts for patients with aneurysms of the descending thoracic aorta [J].
Dake, MD ;
Miller, DC ;
Mitchell, RS ;
Semba, CP ;
Moore, KA ;
Sakai, T .
JOURNAL OF THORACIC AND CARDIOVASCULAR SURGERY, 1998, 116 (05) :689-703
[10]   Time dependent non-Newtonian numerical study of the flow field in a realistic model of aortic arch [J].
Del Gaudio, C. ;
Morbiducci, U. ;
Grigioni, M. .
INTERNATIONAL JOURNAL OF ARTIFICIAL ORGANS, 2006, 29 (07) :709-718