Effect of Surface Pretreatment on Corrosion Resistance of Anodically Oxidized Ti6Al7Nb Alloy

被引:0
作者
Szewczenko, Janusz [1 ]
Marciniak, Jan [1 ]
Tyrlik-Held, Jadwiga [1 ]
Nowinska, Katarzyna [2 ]
机构
[1] Silesian Tech Univ, Fac Biomed Engn, Dept Biomat & Med Devices Engn, PL-44100 Gliwice, Poland
[2] Silesian Tech Univ, Fac Min & Geol, Inst Appl Geol, PL-44100 Gliwice, Poland
来源
INFORMATION TECHNOLOGIES IN BIOMEDICINE, ITIB 2012 | 2012年 / 7339卷
关键词
ELECTROCHEMICAL IMPEDANCE SPECTROSCOPY; INTERVENTIONAL CARDIOLOGY; TI-6AL-4V; TI; COATINGS;
D O I
暂无
中图分类号
Q5 [生物化学];
学科分类号
071010 ; 081704 ;
摘要
Results presented in the paper are an attempt of defining the influence of selected preliminary treatments of surface on the final biocompatibility of anodized Ti6Al7Nb alloy. The anodization process was preceded by combination of different surface modification treatments including grinding, vibration machining, mechanical polishing, sandblasting and electropolishing. Steam sterilization was the final surface modification treatment. Range of research included: surface roughness tests, crevice and pitting corrosion investigations in Ringer's solution, determination of concentration of metallic elements which penetrated to the solution after 28 days exposition and microscopic observations of the specimens surface. The results of research have shown significant influence of preliminary methods of surface preparing which precede anodization. Preliminary preparing of the surface to anodization, as a multistage process, can not be understood solely as a single treatment which directly preceded anodization process. In fact, all the earlier treatments of surface modification before anodization have influence on corrosion resistance.
引用
收藏
页码:398 / 411
页数:14
相关论文
共 23 条
  • [11] Morphological evolution of Ti surfaces during oxidation treatment
    Nagy, PM
    Ferencz, B
    Kálmán, E
    Djuricic, B
    Sonnleitner, R
    [J]. MATERIALS AND MANUFACTURING PROCESSES, 2005, 20 (01) : 105 - 114
  • [12] Phosphoric acid anodization of Ti-6Al-4V - Structural and corrosion aspects
    Narayanan, R.
    Seshadri, S. K.
    [J]. CORROSION SCIENCE, 2007, 49 (02) : 542 - 558
  • [13] Usefulness of passive-carbon layer for implants applied in interventional cardiology
    Paszenda, Z
    Tyrlik-Held, J
    Nawrat, Z
    Zak, J
    Wilczek, K
    [J]. JOURNAL OF MATERIALS PROCESSING TECHNOLOGY, 2004, 157 : 399 - 404
  • [14] Paszenda Z, 2008, ADV INTEL SOFT COMPU, V47, P15
  • [15] Electropolishing and passivation of NiTi shape memory alloy
    Simka, Wojciech
    Kaczmarek, Marcin
    Baron-Wiechec, Aleksandra
    Nawrat, Ginter
    Marciniak, Jan
    Zak, Jerzy
    [J]. ELECTROCHIMICA ACTA, 2010, 55 (07) : 2437 - 2441
  • [16] Sittig C, 1999, J MATER SCI-MATER M, V10, P35, DOI 10.1023/A:1008840026907
  • [17] The effects of spark anodizing treatment of pure titanium metals and titanium alloys on corrosion characteristics
    Song, Ho-Jun
    Kim, Min-Kang
    Jung, Gyoung-Chun
    Vang, Mong-Sook
    Park, Yeong-Joon
    [J]. SURFACE & COATINGS TECHNOLOGY, 2007, 201 (21) : 8738 - 8745
  • [18] Corrosion resistance of Ti-6Al-4V alloy after diverse surface treatments
    Szewczenko, J.
    Walke, W.
    Nowinska, K.
    Marciniak, J.
    [J]. MATERIALWISSENSCHAFT UND WERKSTOFFTECHNIK, 2010, 41 (05) : 360 - 371
  • [19] Szewczenko J, 2011, PRZ ELEKTROTECHNICZN, V87, P228
  • [20] Corrosion behaviour of Ti-6Al-7Nb and Ti-6Al-4V ELI alloys in the simulated body fluid solution by electrochemical impedance spectroscopy
    Tamilselvi, S.
    Raman, V.
    Rajendran, N.
    [J]. ELECTROCHIMICA ACTA, 2006, 52 (03) : 839 - 846