A new smart wristband equipped with an artificial intelligence algorithm to detect atrial fibrillation

被引:74
作者
Chen, Erdong [1 ]
Jiang, Jie [1 ]
Su, Rui [2 ]
Gao, Meng [1 ]
Zhu, Sainan [3 ]
Zhou, Jing [1 ]
Huo, Yong [1 ]
机构
[1] Peking Univ First Hosp, Dept Cardiol, Beijing 100034, Peoples R China
[2] Huami Beijing Informat Technol Co Ltd, Beijing, Peoples R China
[3] Peking Univ First Hosp, Dept Stat, Beijing, Peoples R China
关键词
Artificial intelligence; Atrial fibrillation; Electrocardiogram; Photoplethysmography; Wearable electronic device;
D O I
10.1016/j.hrthm.2020.01.034
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
BACKGROUND Detection of atrial fibrillation (AF) occurrence over a long duration has been a challenge in the screening and follow-up of AF patients. Wearable devices may be an ideal solution. OBJECTIVE The purpose of this study was to measure the sensitivity, specificity, and accuracy of a recently developed smart wristband device that is equipped with both photoplethysmographic (PPG) and single-channel electrocardiogram (ECG) systems and an AF-identifying, artificial intelligence (AI) algorithm, used in the short term. METHODS Use of the Amazfit Health Band 1S, which records both PPG and single-channel ECG data, was assessed in 401 patients (251 normal individuals and 150 ECG-diagnosed AF patients). RESULTS ECG and PPG readings could not be judged in 15 and 18 subjects, respectively. Subjects who were unable to be judged were defined as either false negative or false positive. The sensitivity, specificity, and accuracy of wristband PPG readings were 88.00%, 96.41%, and 93.27%, respectively, and those of wristband ECG readings were 87.33%, 99.20%, and 94.76%, respectively. When the original wristband ECG records were judged by physicians, the sensitivity, specificity, and accuracy were 96.67%, 98.01%, and 97.51%, respectively. CONCLUSION This promising new combination of PPG, ECG, and AI algorithm has the potential to facilitate AF detection.
引用
收藏
页码:847 / 853
页数:7
相关论文
共 15 条
[1]   Cost-effectiveness of mass screening for untreated atrial fibrillation using intermittent ECG recording [J].
Aronsson, Mattias ;
Svennberg, Emma ;
Rosenqvist, Marten ;
Engdahl, Johan ;
Al-Khalili, Faris ;
Friberg, Leif ;
Frykman-Kull, Viveka ;
Levin, Lars-Ake .
EUROPACE, 2015, 17 (07) :1023-1029
[2]  
Camm AJ, 2010, EUROPACE, V12, P1360, DOI [10.1093/eurheartj/ehq278, 10.1093/europace/euq350]
[3]   A Comprehensive Evaluation of Rhythm Monitoring Strategies for the Detection of Atrial Fibrillation Recurrence Insights From 647 Continuously Monitored Patients and Implications for Monitoring After Therapeutic Interventions [J].
Charitos, Efstratios I. ;
Stierle, Ulrich ;
Ziegler, Paul D. ;
Baldewig, Malte ;
Robinson, Derek R. ;
Sievers, Hans-Hinrich ;
Hanke, Thorsten .
CIRCULATION, 2012, 126 (07) :806-814
[4]   STARD 2015 guidelines for reporting diagnostic accuracy studies: explanation and elaboration [J].
Cohen, Jeremie F. ;
Korevaar, Daniel A. ;
Altman, Douglas G. ;
Bruns, David E. ;
Gatsonis, Constantine A. ;
Hooft, Lotty ;
Irwig, Les ;
Levine, Deborah ;
Reitsma, Johannes B. ;
de Vet, Henrica C. W. ;
Bossuyt, Patrick M. M. .
BMJ OPEN, 2016, 6 (11)
[5]   Digital health: Present conundrum, future hope or hype? [J].
Deering, Thomas F. ;
Hindricks, Gerhard ;
Marrouche, Nassir F. .
HEART RHYTHM, 2019, 16 (09) :1303-1304
[6]   The WATCH AF Trial: SmartWATCHes for Detection of Atrial Fibrillation [J].
Doerr, Marcus ;
Nohturfft, Vivien ;
Brasier, Noe ;
Bosshard, Emil ;
Djurdjevic, Aleksandar ;
Gross, Stefan ;
Raichle, Christina J. ;
Rhinisperger, Mattias ;
Stoeckli, Raphael ;
Eckstein, Jens .
JACC-CLINICAL ELECTROPHYSIOLOGY, 2019, 5 (02) :199-208
[7]  
Guo Y, 2019, J AM COLL CARDIOL, V74, P2375
[8]  
Nemati S, 2016, IEEE ENG MED BIO, P3394, DOI 10.1109/EMBC.2016.7591456
[9]   Detection of Atrial Fibrillation with Intermittent Handheld Electrocardiogram in Patients with Ischemic Stroke and Transient Ischemic Attack [J].
Olsson, Ann-Sofie ;
Engdahl, Johan .
JOURNAL OF STROKE & CEREBROVASCULAR DISEASES, 2016, 25 (11) :2648-2652
[10]  