Comparing the cup deformation following implantation of a novel ceramic-on-ceramic hip resurfacing bearing to a metal standard in a cadaveric model

被引:4
作者
Farrier, Adam James [1 ]
Moore, Lauren [2 ]
Manning, Will [1 ]
Avila, Carolina [3 ]
Collins, Simon N. [3 ]
Holland, James [1 ]
机构
[1] Newcastle Upon Tyne Hosp NHS Fdn Trust, Newcastle Upon Tyne, Tyne & Wear, England
[2] Queens Univ Belfast, Sch Med Dent & Biomed Sci, Belfast, Antrim, North Ireland
[3] MatOrtho Ltd, R&D, Leatherhead, Surrey, England
关键词
Deformation; resurfacing; ceramics; hip prosthesis; metallic materials; orthopaedic materials; PRESS-FIT; BONE;
D O I
10.1177/0954411919845721
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Hip resurfacing is an attractive alternative to total hip replacement preserving bone and reducing dislocation risk. Recent metal-on-metal designs have caused failure due to metal wear debris. Ceramic implants may mitigate this risk. Deformation of the acetabular cup can affect the lubrication, producing high friction torques between the femoral head and the cup that would increase wear and/or lead to cup loosening due to femoral head clamping. Our objective was to quantify the deformation of a novel monobloc ceramic hip resurfacing cup component compared to a metal standard, in a fresh frozen cadaveric model using a press-fit technique representative of standard surgical conditions. For this study eight acetabula were prepared from four fresh frozen cadavers. One surgeon with extensive experience in hip resurfacing surgery (J.H.) prepared the acetabulum by sequential reaming. The implants were then impacted into the acetabulum. Four ceramic and four metal implants were used of equal and varying size. Deformation was measured peri-implantation, and at 30 min, using an optical high-precision deformation sensor (GOM GmbH, Braunschweig, Germany). The maximum inscribed circle and the measurement of radial segment techniques were used. Deformation was greater in the metal implants (mean: 34-22mm) immediately after implantation. At 30 min after implantation, the deformation increased to 36mm in the metal and 26mm in the ceramic cup. Greater diameter changes were observed in larger cups. Metal and ceramic implants did not return to the initial diameter. We conclude the ceramic resurfacing acetabular implants undergo similar deformation to existing metal-on-metal implants. The deformation observed was significantly less in the ceramic component at 30 min on one measure. Less deformation may result in better surface conditions and wear characteristics. Deformation change did not resolve after 30 min for both implants.
引用
收藏
页码:603 / 610
页数:8
相关论文
共 20 条
[1]  
[Anonymous], 1036022009 ISO
[2]  
[Anonymous], 2014, TOT HIP REPL RES ART
[3]   The influence of the strength of bone on the deformation of acetabular shells A LABORATORY EXPERIMENT IN CADAVERS [J].
Bone, M. C. ;
Dold, P. ;
Flohr, M. ;
Preuss, R. ;
Joyce, T. J. ;
Aspden, R. M. ;
Holland, J. ;
Deehan, D. .
BONE & JOINT JOURNAL, 2015, 97B (04) :473-477
[4]   A novel method for measuring acetabular cup deformation in cadavers [J].
Bone, Martin C. ;
Dold, Philipp ;
Flohr, Markus ;
Preuss, Roman ;
Joyce, Tom J. ;
Deehan, David ;
Holland, James .
PROCEEDINGS OF THE INSTITUTION OF MECHANICAL ENGINEERS PART H-JOURNAL OF ENGINEERING IN MEDICINE, 2013, 227 (12) :1341-1344
[5]   Validation of an optical system to measure acetabular shell deformation in cadavers [J].
Dold, Philipp ;
Bone, Martin C. ;
Flohr, Markus ;
Preuss, Roman ;
Joyce, Tom J. ;
Deehan, David ;
Holland, James .
PROCEEDINGS OF THE INSTITUTION OF MECHANICAL ENGINEERS PART H-JOURNAL OF ENGINEERING IN MEDICINE, 2014, 228 (08) :781-786
[6]   Metal-on-Metal Hip Arthroplasty: A Review of Adverse Reactions and Patient Management [J].
Drummond, James ;
Phong Tran ;
Fary, Camdon .
JOURNAL OF FUNCTIONAL BIOMATERIALS, 2015, 6 (03) :486-499
[7]   Cup Press Fit in Uncemented THA Depends on Sex, Acetabular Shape, and Surgical Technique [J].
Garcia-Rey, Eduardo ;
Garcia-Cimbrelo, Eduardo ;
Cruz-Pardos, Ana .
CLINICAL ORTHOPAEDICS AND RELATED RESEARCH, 2012, 470 (11) :3014-3023
[8]   The influence of head diameter and wall thickness on deformations of metallic acetabular press-fit cups and UHMWPE liners: a finite element analysis [J].
Goebel, Paul ;
Kluess, Daniel ;
Wieding, Jan ;
Souffrant, Robert ;
Heyer, Horst ;
Sander, Manuela ;
Bader, Rainer .
JOURNAL OF ORTHOPAEDIC SCIENCE, 2013, 18 (02) :264-270
[9]   Biological effects of metal degradation in hip arthroplasties [J].
Granchi, Donatella ;
Savarino, Lucia Maria ;
Ciapetti, Gabriela ;
Baldini, Nicola .
CRITICAL REVIEWS IN TOXICOLOGY, 2018, 48 (02) :170-193
[10]  
Guo X., 2001, BONE MECH HDB, V2nd, P1