COMPARING DIGITAL TOMOSYNTHESIS TO CONE-BEAM CT FOR POSITION VERIFICATION IN PATIENTS UNDERGOING PARTIAL BREAST IRRADIATION

被引:32
作者
Zhang, Junan [1 ]
Wu, Jackie [1 ]
Godfrey, Devon J. [1 ]
Fatunase, Toyosi [1 ]
Marks, Lawrence B. [1 ]
Yin, Fang-Fang [1 ]
机构
[1] Duke Univ, Med Ctr, Dept Radiat Oncol, Durham, NC 27710 USA
来源
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS | 2009年 / 73卷 / 03期
关键词
On-board imaging; Digital tomosynthesis; Cone-beam CT; Patient localization; Breast cancer; Limited angle cone-beam CT; COMPUTED-TOMOGRAPHY; RADIATION-THERAPY; 2ND CANCERS; LOCALIZATION; FEASIBILITY; LUMPECTOMY; GUIDANCE; WOMEN; SETUP;
D O I
10.1016/j.ijrobp.2008.10.036
中图分类号
R73 [肿瘤学];
学科分类号
100214 ;
摘要
Purpose: To evaluate digital tomosynthesis (DTS) technology for daily positioning of patients receiving accelerated partial breast irradiation (APBI) and to compare the positioning accuracy of DTS to three-dimensional cone-beam computed tomography (CBCT). Methods and Materials: Ten patients who underwent APBI were scanned daily with on-board CBCT. A subset of the CBCT projections was used to reconstruct a stack of DTS image slices. To optimize soft-tissue visibility, the DTS images were reconstructed in oblique directions so that the tumor bed, breast tissue, ribs, and lungs were well separated. Coronal and sagittal DTS images were also reconstructed. Translational shifts of DTS images were obtained on different days from the same patients and were compared with the translational shifts of corresponding CBCT images. Seventy-seven CBCT scans and 291 DTS scans were obtained from nine evaluable patients. Results: Tumor beds were best visible in the oblique DTS scans. One-dimensional positioning differences between DTS and CBCT images were 0.8-1.7 mm for the six patients with clips present and 1.2-2.0 mm for the three patients without clips. Because of the limited DTS scan angle, the DTS registration accuracy along the off-plane direction is lower than the accuracy along the in-plane directions. Conclusions: For patients receiving APBI, DTS localization offers comparable accuracy to CBCT localization for daily patient positioning while reducing mechanical constraints and imaging dose. Published by Elsevier Inc.
引用
收藏
页码:952 / 957
页数:6
相关论文
共 21 条
  • [11] Second malignancies after treatment of early-stage breast cancer: Lumpectomy and radiation therapy versus mastectomy
    Obedian, E
    Fischer, DB
    Haffty, BG
    [J]. JOURNAL OF CLINICAL ONCOLOGY, 2000, 18 (12) : 2406 - 2412
  • [12] Digital breast tomosynthesis: Initial experience in 98 women with abnormal digital screening mammography
    Poplack, Steven P.
    Tosteson, Tor D.
    Kogel, Christine A.
    Nagy, Helene M.
    [J]. AMERICAN JOURNAL OF ROENTGENOLOGY, 2007, 189 (03) : 616 - 623
  • [13] Cone-beam computed tomography for on-line image guidance of lung stereotactic radiotherapy: Localization, verification, and intrafraction tumor position
    Purdie, Thomas G.
    Bissonnette, Jean-Pierre
    Franks, Kevin
    Bezjak, Andrea
    Payne, David
    Sie, Fanny
    Sharpe, Michael B.
    Jaffray, David A.
    [J]. INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2007, 68 (01): : 243 - 252
  • [14] Sone S, 1996, Radiat Med, V14, P53
  • [15] A dose comparison study between XVI® and OBI® CBCT systems
    Song, William Y.
    Kamath, Srijit
    Ozawal, Shuichi
    Al Ani, Shlomi
    Chvetsov, Alexei
    Lhandare, Niranjan
    Palta, Jatinder R.
    Liu, Chihray
    Li, Jonathan G.
    [J]. MEDICAL PHYSICS, 2008, 35 (02) : 480 - 486
  • [16] Tomosynthesis-based localization of radioactive seeds in prostate brachytherapy
    Tutar, IB
    Managuli, R
    Shamdasani, V
    Cho, PS
    Pathak, SD
    Kim, Y
    [J]. MEDICAL PHYSICS, 2003, 30 (12) : 3135 - 3142
  • [17] A phase I/II trial to evaluate three-dimensional conformal radiation therapy confined to the region of the lumpectomy cavity for stage I/II breast carcinoma: Initial report of feasibility and reproducibility of radiation therapy oncology group (RTOG) study 0319
    Vicini, F
    Winter, K
    Straube, W
    Wong, J
    Pass, H
    Rabinovitch, R
    Chafe, S
    Arthur, D
    Petersen, I
    McCormick, B
    [J]. INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2005, 63 (05): : 1531 - 1537
  • [18] Cone beam computed tomography guidance for setup of patients receiving accelerated partial breast irradiation
    White, Elizabeth A.
    Cho, John
    Vallis, Katherine A.
    Sharpe, Michael B.
    Lee, Grace
    Blackburn, Helen
    Nageeti, Tahani
    McGibney, Carol
    Jaffray, David A.
    [J]. INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2007, 68 (02): : 547 - 554
  • [19] On-board patient positioning for head-and-neck imrt: Comparing digital tomosynthesis to kilovoltage radiography and cone-beam computed tomography
    Wu, Q. Jackie
    Godfrey, Devon J.
    Wang, Zhiheng
    Zhang, Junan
    Zhou, Sumin
    Yoo, Sua
    Brizel, David M.
    Yin, Fang-Fang
    [J]. INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2007, 69 (02): : 598 - 606
  • [20] Fast reconstruction of digital tomosynthesis using on-board images
    Yan, Hui
    Godfrey, Devon J.
    Yin, Fang-Fang
    [J]. MEDICAL PHYSICS, 2008, 35 (05) : 2162 - 2169