Automated preconcentration and analysis of organic compounds by on-line hollow fiber liquid-phase microextraction-high performance liquid chromatography

被引:52
作者
Esrafili, Ali [1 ,2 ]
Yamini, Yadollah [1 ]
Ghambarian, Mahnaz [3 ]
Ebrahimpour, Behnam [1 ]
机构
[1] Tarbiat Modares Univ, Fac Sci, Dept Chem, Tehran, Iran
[2] Univ Tehran Med Sci, Sch Publ Hlth, Dept Environm Hlth Engn, Tehran, Iran
[3] ACECR, Iranian Res & Dev Ctr Chem Ind, Tehran, Iran
关键词
On-line hollow fiber liquid-phase microextraction; High performance liquid chromatography; Pyridine derivatives; Cigarette smoke; GAS-CHROMATOGRAPHY; CAPILLARY-ELECTROPHORESIS; MASS SPECTROMETRY; MEMBRANE MICROEXTRACTION; SOLVENT MICROEXTRACTION; SAMPLE PREPARATION; WATER SAMPLES; SINGLE DROP; PESTICIDES; DRUGS;
D O I
10.1016/j.chroma.2012.09.003
中图分类号
Q5 [生物化学];
学科分类号
071010 ; 081704 ;
摘要
The present work describes the first automated instrument, based on on-line hollow fiber liquid-phase microextraction (HF-LPME)-high performance liquid chromatography (HPLC), for the preconcentration and determination of organic compounds in various matrices. Using an automated syringe pump for loading the supported liquid membrane and acceptor solvents, a platform lift for moving the sample vial, a sampling loop for on-line injection of the extract to HPLC, along with an electronic board with an AVR microcontroller for storage of data and instrument programs, a sample preparation-HPLC method was developed that allowed sample extraction and extract injection to be carried out completely automatically. Pyridine and pyridine derivatives were chosen for the development and for testing the applicability of the automated instrument. The limits of detection (3 times the S/N) ranged from 0.5 to 1.0 mu g L-1. Effective preconcentration of the analytes was also achieved (preconcentration factors of between 40 and 220). The main advantages of the method developed are minimum sample manipulation, full automation, suitable extraction time, low solvent consumption, and ease of use. The applicability of the on-line automated HF-LPME/HPLC-UV instrument was validated for quantitative extraction and determination of pyridines in cigarette smoke. (C) 2012 Elsevier B.V. All rights reserved.
引用
收藏
页码:27 / 33
页数:7
相关论文
共 22 条
[11]   Dynamic three-phase microextraction as a sample preparation technique prior to capillary electrophoresis [J].
Hou, L ;
Lee, HK .
ANALYTICAL CHEMISTRY, 2003, 75 (11) :2784-2789
[12]   Solvent microextraction into a single drop [J].
Jeannot, MA ;
Cantwell, FF .
ANALYTICAL CHEMISTRY, 1996, 68 (13) :2236-2240
[13]   Miniaturized hollow fiber assisted liquid-phase microextraction and gas chromatography-mass spectrometry for the measurement of progesterone in human serum [J].
Kawaguchi, Migaku ;
Takatsu, Akiko .
JOURNAL OF CHROMATOGRAPHY B-ANALYTICAL TECHNOLOGIES IN THE BIOMEDICAL AND LIFE SCIENCES, 2009, 877 (03) :343-346
[14]   Automation and optimization of liquid-phase microextraction by gas chromatography [J].
Ouyang, Gangfeng ;
Zhao, Wennan ;
Pawliszyn, Janusz .
JOURNAL OF CHROMATOGRAPHY A, 2007, 1138 (1-2) :47-54
[15]   Kinetic calibration for automated hollow fiber-protected liquid-phase microextraction [J].
Ouyang, Gangfeng ;
Pawliszyn, Janusz .
ANALYTICAL CHEMISTRY, 2006, 78 (16) :5783-5788
[16]   Liquid-liquid-liquid microextraction for sample preparation of biological fluids prior to capillary electrophoresis [J].
Pedersen-Bjergaard, S ;
Rasmussen, KE .
ANALYTICAL CHEMISTRY, 1999, 71 (14) :2650-2656
[17]   Development of an automatic multiple dynamic hollow fibre liquid-phase microextraction procedure for specific migration analysis of new active food packagings containing essential oils [J].
Pezo, Davinson ;
Salafranca, Jesus ;
Nerin, Cristina .
JOURNAL OF CHROMATOGRAPHY A, 2007, 1174 (1-2) :85-94
[18]   Rapid determination of pyridine derivatives by dispersive liquid-liquid microextraction coupled with gas chromatography/gas sensor based on nanostructured conducting polypyrrole [J].
Pirsa, Sajad ;
Alizadeh, Naader .
TALANTA, 2011, 87 :249-254
[19]   Determination of organic compounds in water using dispersive liquid-liquid microextraction [J].
Rezaee, Mohammad ;
Assadi, Yaghoub ;
Hosseinia, Mohammad-Reza Milani ;
Aghaee, Elham ;
Ahmadi, Fardin ;
Berijani, Sana .
JOURNAL OF CHROMATOGRAPHY A, 2006, 1116 (1-2) :1-9
[20]   Determination of pyridine, 2-picoline, 4-picoline and quinoline from mainstream cigarette smoke by solid-phase extraction liquid chromatography/electrospray ionization tandem mass spectrometry [J].
Saha, Subhrakanti ;
Mistri, Rajib ;
Ray, B. C. .
JOURNAL OF CHROMATOGRAPHY A, 2010, 1217 (03) :307-311