Gated Myocardial Perfusion SPECT: Algorithm-Specific Influence of Reorientation on Calculation of Left Ventricular Volumes and Ejection Fraction

被引:9
|
作者
Knollmann, Daniela [1 ]
Winz, Oliver H. [1 ]
Meyer, Philipp T. [1 ]
Raptis, Mardjan [1 ]
Krohn, Thomas [1 ]
Koch, Karl-Christian [2 ]
Schaefer, Wolfgang M. [1 ]
机构
[1] Rhein Westfal TH Aachen, Univ Hosp Aachen, Dept Nucl Med, D-52074 Aachen, Germany
[2] Rhein Westfal TH Aachen, Univ Hosp Aachen, Med Clin Cardiol 1, D-52074 Aachen, Germany
关键词
gated SPECT; QGS; 4D-MSPECT; CARE heart; reorientation;
D O I
10.2967/jnumed.108.050484
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Gated myocardial perfusion SPECT allows calculation of end-diastolic and end-systolic volumes (EDV and ESV, respectively) and left ventricular ejection fraction (LVEF). The quantification algorithms QGS (quantitative gated SPECT), 4D-MSPECT, and CARE heart show a good correlation with cardiac MRI. Nevertheless, differences in contour finding suggest algorithm-specific effects if heart axes vary. The effect of tilting heart axes on gated SPECT was quantified as a possible source of error. Methods: Sixty men underwent gated SPECT (450 MBq of Tc-99m-tetrofosmin or sestamibi, 8 gates/cycle). After correct reorientation (R-0), datasets were tilted by 5, 10, 15, 20, 30, and 45 along both long axes (R-5, R-10, R-15, R-20, R-30, and R-45, respectively). EDV, ESV, and LVEF were calculated using QGS, 4D-MSPECT, and CARE heart. Because a 15 tilt could be a maximum possible misreorientation in routine, R-0 and R-15 results were analyzed in detail. Absolute-difference values between results of tilted and correctly reoriented datasets were calculated for all tilts and algorithms. Results: QGS and CARE heart succeeded for R-0 and R-15 in all cases, whereas 4D-MSPECT failed to find the basal plane in 1 case (patient B). R-2 values between paired R-15/R-0 results were 0.992 (QGS), 0.796 (4D-MSPECT; R-2 = 0.919 inn = 59 after exclusion of the failed case), and 0.916 (CARE heart) for EDV; 0.994 (QGS), 0.852 (4D-MSPECT; R-2 = 0.906 in n = 59), and 0.899 (CARE heart) for ESV; and 0.988 (QGS), 0.814 (4D-MSPECT; R-2 = 0.810 in n = 59), and 0.746 (CARE heart) for LVEF Concerning all levels of misreorientation, 1 patient was excluded for all algorithms because of multiple problems in contour finding; additionally for 4D-MSPECT patient B was excluded. In the 45 group, QGS succeeded in 58 of 59 cases, 4D-MSPECT in 58 of 58, and CARE heart in 33 of 59. Mean absolute differences for EDV ranged from 5.1 +/- 4.1 to 12.8 +/- 10.5 mL for QGS, from 6.7 +/- 6.3 to 34.2 +/- 20.7 mL for 4D-MSPECT, and from 5.4 +/- 5.6 to 25.2 +/- 16.1 mL for CARE heart (tilts between 5 and 45). Mean absolute differences for ESV ranged from 4.1 +/- 3.7 to 8.0 +/- 9.4 mL for QGS, from 5.6 +/- 8.0 to 10.0 +/- 10.5 mL for 4D-MSPECT, and from 5.4 +/- 5.6 to 25.5 +/- 16.1 mLfor CARE heart. Mean absolute differences for LVEF ranged from 1.1% +/- 1.0% to 2.2% +/- 1.8% for QGS, from 4.0% +/- 3.5% +/- to 8.0% +/- 7.1 for 4D-MSPECT, and from 3.4% +/- 2.9% to 9.2% +/- 6.0% for CARE heart. Conclusion: Despite tilted heart axes, QGS showed stable results even when using tilts up to 45. 4D-MSPECT and CARE heart results varied with reorientation of the heart axis, implying that published validation results apply to correctly reoriented data only.
引用
收藏
页码:1636 / 1642
页数:7
相关论文
共 50 条
  • [1] Quantitative Myocardial Perfusion-SPECT Algorithm-Specific Influence of Reorientation on Calculation of Summed Stress Score
    Knollmann, Daniela
    Raptis, Mardjan
    Meyer, Philipp T.
    Winz, Oliver H.
    Krohn, Thomas
    Schaefer, Wolfgang M.
    CLINICAL NUCLEAR MEDICINE, 2012, 37 (11) : 1089 - 1093
  • [2] Comparison of totally automated and manual processing for calculation of left ventricular volumes and ejection fraction using gated myocardial perfusion SPECT
    Al-Shammari, Amirah Fahad
    Afzal, Uzma
    CARDIOMETRY, 2022, (22): : 136 - 142
  • [3] Impact of imaging protocol on left ventricular ejection fraction using gated-SPECT myocardial perfusion imaging
    C. Marcassa
    R. Giubbini
    W. Acampa
    C. Cittanti
    O. Djepaxhija
    A. Gimelli
    A. Kokomani
    G. Medolago
    E. Milan
    R. Sciagrà
    Journal of Nuclear Cardiology, 2017, 24 : 1292 - 1301
  • [4] Impact of imaging protocol on left ventricular ejection fraction using gated-SPECT myocardial perfusion imaging
    Marcassa, C.
    Giubbini, R.
    Acampa, W.
    Cittanti, C.
    Djepaxhija, O.
    Gimelli, A.
    Kokomani, A.
    Medolago, G.
    Milan, E.
    Sciagra, R.
    JOURNAL OF NUCLEAR CARDIOLOGY, 2017, 24 (04) : 1292 - 1301
  • [5] Rest Versus Stress Ejection Fraction on Gated Myocardial Perfusion SPECT
    Wheat, Janelle M.
    Currie, Geoffrey M.
    JOURNAL OF NUCLEAR MEDICINE TECHNOLOGY, 2005, 33 (04) : 218 - 223
  • [6] Evaluation of left ventricular volumes and ejection fraction by gated SPECT and cardiac MRI in patients with dilated cardiomyopathy
    Feng Wang
    Jian Zhang
    Wei Fang
    Shi-Hua Zhao
    Min-Jie Lu
    Zuo-Xiang He
    European Journal of Nuclear Medicine and Molecular Imaging, 2009, 36 : 1611 - 1621
  • [7] Evaluation of left ventricular volumes and ejection fraction by gated SPECT and cardiac MRI in patients with dilated cardiomyopathy
    Wang, Feng
    Zhang, Jian
    Fang, Wei
    Zhao, Shi-Hua
    Lu, Min-Jie
    He, Zuo-Xiang
    EUROPEAN JOURNAL OF NUCLEAR MEDICINE AND MOLECULAR IMAGING, 2009, 36 (10) : 1611 - 1621
  • [8] Comparison between Tc-99m-sestamibi gated myocardial perfusion SPECT and echocardiography in assessment of left ventricular volumes and ejection fraction - effect of perfusion defect and small heart
    Danesh-Sani, Seyed Hashem
    Zakavi, Seyed Rasoul
    Oskoueian, Leyla
    Kakhki, Vahid-Reza Dabbagh
    NUCLEAR MEDICINE REVIEW, 2014, 17 (02) : 70 - 74
  • [9] Left ventricular volumes and ejection fraction calculated from quantitative electrocardiographic-gated 99mTc-tetrofosmin myocardial SPECT
    Yoshioka, J
    Hasegawa, S
    Yamaguchi, H
    Tokita, N
    Paul, AK
    Mu, XL
    Maruyama, A
    Hori, M
    Nishimura, T
    JOURNAL OF NUCLEAR MEDICINE, 1999, 40 (10) : 1693 - 1698
  • [10] A comparison of 8 and 16 frames gated SPECT imaging for determination of left ventricular volumes and ejection fraction: effects of gender and myocardial counts
    Mazinani, Mahdi
    Tajik-Mansoury, Mohammad Ali
    Sabour, Mahsa
    Jadidi, Majid
    INTERNATIONAL JOURNAL OF CARDIOVASCULAR IMAGING, 2021, 37 (06) : 2079 - 2084