Gated Myocardial Perfusion SPECT: Algorithm-Specific Influence of Reorientation on Calculation of Left Ventricular Volumes and Ejection Fraction

被引:9
作者
Knollmann, Daniela [1 ]
Winz, Oliver H. [1 ]
Meyer, Philipp T. [1 ]
Raptis, Mardjan [1 ]
Krohn, Thomas [1 ]
Koch, Karl-Christian [2 ]
Schaefer, Wolfgang M. [1 ]
机构
[1] Rhein Westfal TH Aachen, Univ Hosp Aachen, Dept Nucl Med, D-52074 Aachen, Germany
[2] Rhein Westfal TH Aachen, Univ Hosp Aachen, Med Clin Cardiol 1, D-52074 Aachen, Germany
关键词
gated SPECT; QGS; 4D-MSPECT; CARE heart; reorientation;
D O I
10.2967/jnumed.108.050484
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Gated myocardial perfusion SPECT allows calculation of end-diastolic and end-systolic volumes (EDV and ESV, respectively) and left ventricular ejection fraction (LVEF). The quantification algorithms QGS (quantitative gated SPECT), 4D-MSPECT, and CARE heart show a good correlation with cardiac MRI. Nevertheless, differences in contour finding suggest algorithm-specific effects if heart axes vary. The effect of tilting heart axes on gated SPECT was quantified as a possible source of error. Methods: Sixty men underwent gated SPECT (450 MBq of Tc-99m-tetrofosmin or sestamibi, 8 gates/cycle). After correct reorientation (R-0), datasets were tilted by 5, 10, 15, 20, 30, and 45 along both long axes (R-5, R-10, R-15, R-20, R-30, and R-45, respectively). EDV, ESV, and LVEF were calculated using QGS, 4D-MSPECT, and CARE heart. Because a 15 tilt could be a maximum possible misreorientation in routine, R-0 and R-15 results were analyzed in detail. Absolute-difference values between results of tilted and correctly reoriented datasets were calculated for all tilts and algorithms. Results: QGS and CARE heart succeeded for R-0 and R-15 in all cases, whereas 4D-MSPECT failed to find the basal plane in 1 case (patient B). R-2 values between paired R-15/R-0 results were 0.992 (QGS), 0.796 (4D-MSPECT; R-2 = 0.919 inn = 59 after exclusion of the failed case), and 0.916 (CARE heart) for EDV; 0.994 (QGS), 0.852 (4D-MSPECT; R-2 = 0.906 in n = 59), and 0.899 (CARE heart) for ESV; and 0.988 (QGS), 0.814 (4D-MSPECT; R-2 = 0.810 in n = 59), and 0.746 (CARE heart) for LVEF Concerning all levels of misreorientation, 1 patient was excluded for all algorithms because of multiple problems in contour finding; additionally for 4D-MSPECT patient B was excluded. In the 45 group, QGS succeeded in 58 of 59 cases, 4D-MSPECT in 58 of 58, and CARE heart in 33 of 59. Mean absolute differences for EDV ranged from 5.1 +/- 4.1 to 12.8 +/- 10.5 mL for QGS, from 6.7 +/- 6.3 to 34.2 +/- 20.7 mL for 4D-MSPECT, and from 5.4 +/- 5.6 to 25.2 +/- 16.1 mL for CARE heart (tilts between 5 and 45). Mean absolute differences for ESV ranged from 4.1 +/- 3.7 to 8.0 +/- 9.4 mL for QGS, from 5.6 +/- 8.0 to 10.0 +/- 10.5 mL for 4D-MSPECT, and from 5.4 +/- 5.6 to 25.5 +/- 16.1 mLfor CARE heart. Mean absolute differences for LVEF ranged from 1.1% +/- 1.0% to 2.2% +/- 1.8% for QGS, from 4.0% +/- 3.5% +/- to 8.0% +/- 7.1 for 4D-MSPECT, and from 3.4% +/- 2.9% to 9.2% +/- 6.0% for CARE heart. Conclusion: Despite tilted heart axes, QGS showed stable results even when using tilts up to 45. 4D-MSPECT and CARE heart results varied with reorientation of the heart axis, implying that published validation results apply to correctly reoriented data only.
引用
收藏
页码:1636 / 1642
页数:7
相关论文
共 17 条
[1]   Left ventricular function: Correlation of quantitative gated SPECT and MR imaging over a wide range of values [J].
Bavelaar-Croon, CDL ;
Kayser, HWM ;
van der Wall, EE ;
de Roos, A ;
Dibbets-Schneider, P ;
Pauwels, EKJ ;
Germano, G ;
Atsma, DE .
RADIOLOGY, 2000, 217 (02) :572-575
[2]   STATISTICAL METHODS FOR ASSESSING AGREEMENT BETWEEN TWO METHODS OF CLINICAL MEASUREMENT [J].
BLAND, JM ;
ALTMAN, DG .
LANCET, 1986, 1 (8476) :307-310
[3]   Mortality risk associated with ejection fraction differs across resting nuclear perfusion findings [J].
Bourque, Jamieson M. ;
Velazquez, Eric J. ;
Tuttle, Robert H. ;
Shaw, Linda K. ;
O'Connor, Christopher M. ;
Borges-Neto, Salvador .
JOURNAL OF NUCLEAR CARDIOLOGY, 2007, 14 (02) :165-173
[4]  
DEPUEY EG, 1993, J NUCL MED, V34, P1871
[5]  
Ficaro EP, 1999, CIRCULATION, V100, P26
[6]  
GERMANO G, 1995, J NUCL MED, V36, P2138
[7]   Worsening of left ventricular ejection fraction induced by dipyridamole on TI-201 gated myocardial perfusion imaging predicts significant coronary artery disease [J].
Hung, GU ;
Lee, KW ;
Chen, CP ;
Yang, KT ;
Lin, WY .
JOURNAL OF NUCLEAR CARDIOLOGY, 2006, 13 (02) :225-232
[8]   Electrocardiogram-gated single-photon emission computed tomography versus cardiac magnetic resonance imaging for the assessment of left ventricular volumes and ejection fraction - A meta-analysis [J].
Ioannidis, JPA ;
Trikalinos, TA ;
Danias, PG .
JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 2002, 39 (12) :2059-2068
[9]   Validation of left ventricular volume measurements by gated SPECT 99mTc-labeled sestamibi imaging [J].
Iskandrian, AE ;
Germano, G ;
VanDecker, W ;
Ogilby, JD ;
Wolf, N ;
Mintz, R ;
Berman, DS .
JOURNAL OF NUCLEAR CARDIOLOGY, 1998, 5 (06) :574-578
[10]   Validation of 4D-MSPECT and QGS for quantification of left ventricular volumes and ejection fraction from gated 99mTc-MIBI SPET:: comparison with cardiac magnetic resonance imaging [J].
Lipke, CSA ;
Kühl, HP ;
Nowak, B ;
Kaiser, HJ ;
Reinartz, P ;
Koch, KC ;
Buell, U ;
Schaefer, WM .
EUROPEAN JOURNAL OF NUCLEAR MEDICINE AND MOLECULAR IMAGING, 2004, 31 (04) :482-490