Lock-in detection using a cryogenic low noise current preamplifier for the readout of resistive bolometers

被引:9
作者
Yvon, D
Sushkov, V
Bernard, R
Bret, JL
Cahan, B
Cloue, O
Maillard, O
Mazeau, B
Passerieux, JP
Paul, B
Veyssiere, C
机构
[1] CEA, Ctr Etud Saclay, DAPNIA, Serv Phys Particules, F-91191 Gif Sur Yvette, France
[2] Ctr Etud Saclay, CEA, DAPNIA, Serv Instrumentat Gen, F-91191 Gif Sur Yvette, France
关键词
D O I
10.1016/S0168-9002(01)01332-8
中图分类号
TH7 [仪器、仪表];
学科分类号
0804 ; 080401 ; 081102 ;
摘要
We implemented a low noise current preamplifier for the readout of resistive bolometers, We tested the apparatus on thermometer resistances ranging from 10 to 500 MOmega. The use of current preamplifier overcomes constraints introduced by the readout time constant due to the thermometer resistance and the input capacitance, Using cold JFETs, this preamplifier board is shown to have very low noise: the Johnson noise of the source resistor (1 fA/Hz(1 2)) dominated in our noise measurements. We also implemented a lock-in chain using this preamplifier. Due to fast risetime, compensation of the phase shift may be unnecessary. If implemented, no tuning is necessary when the sensor impedance changes. Transients are very short, and thus low-passing or sampling of the signal is simplified. In case of spurious noise, the modulation frequency can be chosen in a much wider frequency range. without requiring a new calibration of the apparatus. (C) 2002 Elsevier Science B.V. All rights reserved.