Optimization of the Contrast Mixture Ratio for Simultaneous Direct MR and CT Arthrography: an in Vitro Study

被引:17
作者
Choi, Ja-Young [2 ,3 ]
Kang, Heung Sik [1 ]
Hong, Sung Hwan [2 ,3 ]
Lee, Joon Woo [1 ]
Kim, Na Ra [2 ,3 ]
Jun, Woo Sun [2 ,3 ]
Moon, Sung Gyu [2 ,3 ]
Choi, Jung-Ah [1 ]
机构
[1] Seoul Natl Univ, Bundang Hosp, Dept Radiol, Songnam 463707, Gyeonggi Do, South Korea
[2] Seoul Natl Univ, Coll Med, Dept Radiol, Seoul 110744, South Korea
[3] Seoul Natl Univ, Coll Med, Inst Radiat Med, Seoul 110744, South Korea
关键词
Magnetic resonance (MR); arthrography; Computed tomography (CT); Gadolinium; contrast media;
D O I
10.3348/kjr.2008.9.6.520
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Objective: This study was designed to determine the optimal mixture ratio of gadolinium and iodinated contrast agent for simultaneous direct MR arthrography and CT arthrography. Materials and Methods: An in vitro study was performed utilizing mixtures of gadolinium at six different concentrations (0.625, 1.25, 2.5, 5.0, 10 and 20 mmol/L) and iodinated contrast agent at seven different concentrations (0, 12.5, 25, 37.5, 50, 75 and 92-99.9%). These mixtures were placed in tissue culture plates, and were then imaged with CT and MR (with T1-weighted sequences, proton-density sequences and T2-weighted sequences). CT numbers and signal intensities were measured. Pearson's correlation coefficients were used to assess the correlations between the gadolinium/iodinated contrast agent mixtures and the CT numbers/MR signal intensities. Scatter diagrams were plotted for all gadolinium/iodinated contrast agent combinations and two radiologists in consensus identified the mixtures that yielded the optimal CT numbers and MR signal intensities. Results: The CT numbers showed significant correlation with iodinated contrast concentrations (r = 0.976, p < 0.001), whereas the signal intensities as measured on MR images showed a significant correlation with both gadolinium and iodinated contrast agent concentrations (r = -484 to -0.719, p < 0.001). A review of the CT and MR images, graphs, and scatter diagram of 42 combinations of the contrast agent showed that a concentration of 1.25 mmol/L gadolinium and 25% iodinated contrast agent was the best combination for simultaneous CT and MR imaging. Conclusion: A mixture of 1.25 mmol/L gadolinium and 25% iodinated contrast agent was found to be optimal for simultaneous direct MR arthrography and CT arthrorahgpy.
引用
收藏
页码:520 / 525
页数:6
相关论文
共 15 条
[1]   Comparison of double-contrast CT arthrography image quality with nonionic contrast agents - Isotonic dimeric iodixanol 270 mg I/mL and monomeric iohexol 300 mg I/mL [J].
Blum, AG ;
Simon, JM ;
Cotten, A ;
Quirin-Cosmidis, I ;
Boyer, B ;
Boutry, N ;
Antonini, JP .
INVESTIGATIVE RADIOLOGY, 2000, 35 (05) :304-310
[2]   Lesions of the acetabular labrum: Accuracy of MR imaging and MR arthrography in detection and staging [J].
Czerny, C ;
Hofmann, S ;
Neuhold, A ;
Tschauner, C ;
Engel, A ;
Recht, MP ;
Kramer, J .
RADIOLOGY, 1996, 200 (01) :225-230
[3]  
Engel A, 1990, Acta Orthop Scand Suppl, V240, P1
[4]   CT arthrography and postoperative musculoskeletal imaging with multichannel computed tomography [J].
Farber, JM .
SEMINARS IN MUSCULOSKELETAL RADIOLOGY, 2004, 8 (02) :157-166
[5]   MR ARTHROGRAPHY - ANATOMIC-PATHOLOGICAL INVESTIGATION [J].
HAJEK, PC ;
BAKER, LL ;
SARTORIS, DJ ;
NEUMANN, CH ;
RESNICK, D .
RADIOLOGY, 1987, 163 (01) :141-147
[6]  
JINKINS JR, 1992, AM J NEURORADIOL, V13, P19
[7]   MR ARTHROGRAPHY OF THE SHOULDER WITH GADOPENTETATE DIMEGLUMINE - INFLUENCE OF CONCENTRATION, IODINATED CONTRAST MATERIAL, AND TIME ON SIGNAL INTENSITY [J].
KOPKA, L ;
FUNKE, M ;
FISCHER, U ;
KEATING, D ;
OESTMANN, J ;
GRABBE, E .
AMERICAN JOURNAL OF ROENTGENOLOGY, 1994, 163 (03) :621-623
[8]   Multidetector spiral CT arthrography of the shoulder Clinical applications and limits, with MR arthrography and arthroscopic correlations [J].
Lecouvet, Frederic E. ;
Simoni, Paolo ;
Koutaissoff, Sophie ;
Berg, Bruno C. Vande ;
Malghem, Jacques ;
Dubuc, Jean-Emile .
EUROPEAN JOURNAL OF RADIOLOGY, 2008, 68 (01) :120-136
[9]   Optimization of gadodiamide concentration for MR arthrography at 3 T [J].
Masi, JN ;
Newitt, D ;
Sell, CA ;
Daldrup-Link, H ;
Steinbach, L ;
Majumdar, S ;
Link, TM .
AMERICAN JOURNAL OF ROENTGENOLOGY, 2005, 184 (06) :1754-1761
[10]   Effects of iodinated contrast and field strength on gadolinium enhancement: Implications for direct MR arthrography [J].
Montgomery, DD ;
Morrison, WB ;
Schweitzer, ME ;
Weishaupt, D ;
Dougherty, L .
JOURNAL OF MAGNETIC RESONANCE IMAGING, 2002, 15 (03) :334-343