Intraoperative computed tomography and automated registration for image-guided cranial surgery

被引:24
作者
Eggers, G. [1 ]
Kress, B. [2 ]
Rohde, S. [2 ]
Muehling, J. [1 ]
机构
[1] Univ Heidelberg Hosp, Dept Oral & Craniomaxillofacial Surg, D-69120 Heidelberg, Germany
[2] Univ Heidelberg Hosp, Dept Neuroradiol, D-69120 Heidelberg, Germany
关键词
computed tomography; X-ray; neuronavigation; NAVIGATION; ACCURACY;
D O I
10.1259/dmfr/26098099
中图分类号
R78 [口腔科学];
学科分类号
1003 ;
摘要
Objectives: Two key problems for the use of navigation systems in image-guided surgery are accurate patient-to-image registration and the fact that with ongoing surgery the patient's anatomy is altered while the image data remains unchanged. A system for intraoperative CT imaging and fully automated registration of this image addresses both problems. It had been evaluated successfully in phantom studies. In this clinical study, we assessed the impact of the system on intraoperative workflow and registration accuracy in everyday patient care. Methods: In ten patients who underwent Image-guided surgery, CT image data were acquired intraoperatively and were automatically registered in the navigation system. Registration accuracy and surgical outcome were assessed clinically. In six of these patients, a maxillary splint with markers had been inserted to cross-check registration accuracy. The target registration error of these markers was measured. Results: In all cases, registration accuracy was clinically Sufficient and the surgical task could be performed successfully. In those cases where a maxillary template with target markers was attached for additional control of the registration accuracy, the target registration error was always better than 2 mm. Automated registration reduced the intraoperative registration time considerably and partially compensated for the time needed to perform the image data acquisition. Conclusions: Intraoperative CT imaging and automated registration Successfully address the two key problems of image-guided surgery. The method is robust and accurate and proved its usability in everyday patient care.
引用
收藏
页码:28 / 33
页数:6
相关论文
共 18 条
[1]   Development and implementation of intraoperative magnetic resonance imaging and its neurosurgical applications [J].
Black, PM ;
Moriarty, T ;
Alexander, E ;
Stieg, P ;
Woodard, EJ ;
Gleason, PL ;
Martin, CH ;
Kikinis, R ;
Schwartz, RB ;
Jolesz, FA .
NEUROSURGERY, 1997, 41 (04) :831-842
[2]   Comparison of different 3D navigation systems by a clinical "user" [J].
Cartellieri, M ;
Kremser, J ;
Vorbeck, F .
EUROPEAN ARCHIVES OF OTO-RHINO-LARYNGOLOGY, 2001, 258 (01) :38-41
[3]   Endoscopic sinus surgery using intraoperative computed tomography imaging for updating a three-dimensional navigation system [J].
Cartellieri, M ;
Vorbeck, F .
LARYNGOSCOPE, 2000, 110 (02) :292-296
[4]   Frameless computer-aided surgery system for revision endoscopic sinus surgery [J].
Caversaccio, M ;
Bächler, R ;
Lädrach, K ;
Schroth, G ;
Nolte, LP ;
Häusler, R .
OTOLARYNGOLOGY-HEAD AND NECK SURGERY, 2000, 122 (06) :808-813
[5]   Automated registration of intraoperative CT image data for navigated skull base surgery [J].
Eggers, G. ;
Kress, B. ;
Muehling, J. .
MINIMALLY INVASIVE NEUROSURGERY, 2008, 51 (01) :15-20
[6]   Image-to-patient registration techniques in head surgery [J].
Eggers, G. ;
Muehling, J. ;
Marmulla, R. .
INTERNATIONAL JOURNAL OF ORAL AND MAXILLOFACIAL SURGERY, 2006, 35 (12) :1081-1095
[7]  
Grunert P, 1998, Comput Aided Surg, V3, P166, DOI 10.1002/(SICI)1097-0150(1998)3:4<166::AID-IGS6>3.3.CO
[8]  
2-5
[9]   Indications and limitations of Intraoperative navigation in maxillofacial surgery [J].
Heiland, M ;
Habermann, CR ;
Schmezle, R .
JOURNAL OF ORAL AND MAXILLOFACIAL SURGERY, 2004, 62 (09) :1059-1063
[10]   Intraoperative computed tomography with the mobile CT Tomoscan M during surgical treatment of orbital fractures [J].
Hoelzle, F ;
Klein, M ;
Schwerdtner, O ;
Lueth, T ;
Albrecht, J ;
Hosten, N ;
Felix, R ;
Bier, J .