Octavius 4D characterization for flattened and flattening filter free rotational deliveries

被引:42
作者
McGarry, Conor K. [1 ,2 ]
O'Connell, Barry F. [1 ]
Grattan, Mark W. D. [1 ]
Agnew, Christina E. [1 ]
Irvine, Denise M. [1 ]
Hounsell, Alan R. [1 ,2 ]
机构
[1] Belfast Hlth & Social Care Trust, Northern Ireland Canc Ctr, Belfast BT9 7AB, Antrim, North Ireland
[2] Queens Univ Belfast, Ctr Canc Res & Cell Biol, Belfast BT9 7BL, Antrim, North Ireland
关键词
Octavius; 4D; flattening filter free; VMAT; rotational therapy; arrays; patient specific QA; radiation therapy; COMMERCIAL DETECTOR ARRAYS; IONIZATION-CHAMBER ARRAYS; IMRT QUALITY-ASSURANCE; ION-CHAMBER; DOSE VERIFICATION; CLINICAL ROUTINE; RAPIDARC; PHANTOM; BEAMS; QA;
D O I
10.1118/1.4817482
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Purpose: In this study the Octavius detector 729 ionization chamber (IC) array with the Octavius 4D phantom was characterized for flattening filter (FF) and flattening filter free (FFF) static and rotational beams. The device was assessed for verification with FF and FFF RapidArc treatment plans. Methods: The response of the detectors to field size, dose linearity, and dose rate were assessed for 6 MV FF beams and also 6 and 10 MV FFF beams. Dosimetric and mechanical accuracy of the detector array within the Octavius 4D rotational phantom was evaluated against measurements made using semiflex and pinpoint ionization chambers, and radiochromic film. Verification FF and FFF RapidArc plans were assessed using a gamma function with 3%/3 mm tolerances and 2%/2 mm tolerances and further analysis of these plans was undertaken using film and a second detector array with higher spatial resolution. Results: A warm-up dose of >6 Gy was required for detector stability. Dose-rate measurements were stable across a range from 0.26 to 15 Gy/min and dose response was linear, although the device overestimated small doses compared with pinpoint ionization chamber measurements. Output factors agreed with ionization chamber measurements to within 0.6% for square fields of side between 3 and 25 cm and within 1.2% for 2 x 2 cm(2) fields. The Octavius 4D phantom was found to be consistent with measurements made with radiochromic film, where the gantry angle was found to be within 0.4. of that expected during rotational deliveries. RapidArc FF and FFF beams were found to have an accuracy of >97.9% and >90% of pixels passing 3%/3 mm and 2%/2 mm, respectively. Detector spatial resolution was observed to be a factor in determining the accurate delivery of each plan, particularly at steep dose gradients. This was confirmed using data from a second detector array with higher spatial resolution and with radiochromic film. Conclusions: The Octavius 4D phantom with associated Octavius detector 729 ionization chamber array is a dosimetrically and mechanically stable device for pretreatment verification of FF and FFF RapidArc treatments. Further improvements may be possible through use of a detector array with higher spatial resolution (detector size and/or detector spacing). (C) 2013 American Association of Physicists in Medicine.
引用
收藏
页数:11
相关论文
共 23 条
  • [1] Response to high-energy photons of PTW31014 PinPoint ion chamber with a central aluminum electrode
    Agostinelli, S.
    Garelli, S.
    Piergentili, M.
    Foppiano, F.
    [J]. MEDICAL PHYSICS, 2008, 35 (07) : 3293 - 3301
  • [2] Using an EPID for patient-specific VMAT quality assurance
    Bakhtiari, M.
    Kumaraswamy, L.
    Bailey, D. W.
    de Boer, S.
    Malhotra, H. K.
    Podgorsak, M. B.
    [J]. MEDICAL PHYSICS, 2011, 38 (03) : 1366 - 1373
  • [3] LOWERING WHOLE-BODY RADIATION DOSES IN PEDIATRIC INTENSITY-MODULATED RADIOTHERAPY THROUGH THE USE OF UNFLATTENED PHOTON BEAMS
    Cashmore, Jason
    Ramtohul, Mark
    Ford, Dan
    [J]. INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2011, 80 (04): : 1220 - 1227
  • [4] Comparison of four commercial devices for RapidArc and sliding window IMRT QA
    Chandraraj, Varatharaj
    Stathakis, Sotirios
    Manickam, Ravikumar
    Esquivel, Carlos
    Supe, Sanjay S.
    Papanikolaou, Nikos
    [J]. JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2011, 12 (02): : 338 - 349
  • [5] Evaluation of a new VMAT QA device, or the "X" and "O" array geometries
    Feygelman, Vladimir
    Zhang, Geoffrey
    Stevens, Craig
    Nelms, Benjamin E.
    [J]. JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2011, 12 (02): : 146 - 168
  • [6] Patient QA systems for rotational radiation therapy: A comparative experimental study with intentional errors
    Fredh, Anna
    Scherman, Jonas Bengtsson
    Fog, Lotte S.
    Af Rosenschold, Per Munck
    [J]. MEDICAL PHYSICS, 2013, 40 (03)
  • [7] Evaluation of chamber response function influence on IMRT verification using 2D commercial detector arrays
    Gago-Arias, A.
    Brualla-Gonzalez, L.
    Gonzalez-Castano, D. M.
    Gomez, F.
    Sanchez Garcia, M.
    Luna Vega, V.
    Mosquera Sueiro, J.
    Pardo-Montero, J.
    [J]. PHYSICS IN MEDICINE AND BIOLOGY, 2012, 57 (07) : 2005 - 2020
  • [8] Current status and future perspective of flattening filter free photon beams
    Georg, Dietmar
    Knoos, Tommy
    McClean, Brendan
    [J]. MEDICAL PHYSICS, 2011, 38 (03) : 1280 - 1293
  • [9] On the sensitivity of common gamma-index evaluation methods to MLC misalignments in Rapidarc quality assurance
    Heilemann, G.
    Poppe, B.
    Laub, W.
    [J]. MEDICAL PHYSICS, 2013, 40 (03)
  • [10] Dosimetric evaluation of a 2D pixel ionization chamber for implementation in clinical routine
    Herzen, J.
    Todorovic, M.
    Cremers, F.
    Platz, V.
    Albers, D.
    Bartels, A.
    Schmidt, R.
    [J]. PHYSICS IN MEDICINE AND BIOLOGY, 2007, 52 (04) : 1197 - 1208