Thermo-mechanical phase-shift determination in Coriolis mass-flowmeters with added masses

被引:37
作者
Ghayesh, Mergen H. [1 ]
Amabili, Marco [1 ]
Paidoussis, Michael P. [1 ]
机构
[1] McGill Univ, Dept Mech Engn, Montreal, PQ H3A 0C3, Canada
基金
加拿大自然科学与工程研究理事会;
关键词
Coriolis flowmeter; Mass flowmeter; Perturbation technique; Sensor masses; Thermal effects; Analytical solution; Phase shift; Straight tube; PIPES CONVEYING FLUID; WEIGHT VECTOR; FLOW; DYNAMICS; VIBRATIONS; ERRORS;
D O I
10.1016/j.jfluidstructs.2012.05.003
中图分类号
TH [机械、仪表工业];
学科分类号
0802 ;
摘要
The aim of this paper is to develop an approximate analytical solution for phase-shift (and thus mass flow) prediction along the length of the measuring tube of a Coriolis mass-flowmeter. A single, straight measuring tube is considered; added masses at the sensor and excitation locations are included in the model, and thus in the equation of motion. The measuring tube is excited harmonically by an electromagnetic driver. Taking into account thermal effects, the equation of motion is derived through use of the extended Hamilton's principle and constitutive relations. The equation of motion is discretized into a set of ordinary differential equations via Galerkin's technique. The method of multiple timescales is applied to the set of resultant equations, and the equations of order one and epsilon are obtained analytically for the system at primary resonance. The solution of the equation of motion is obtained by satisfying the solvability condition (making the solution of order epsilon free of secular terms). The flow-related phase-shift in the driver-induced tube vibration is measured at two symmetrically located points on either side of the mid-length of the tube. The analytical results for the phase-shift are compared to those obtained numerically. The effect of system parameters on the measured phase-shift is discussed. It is shown that the measured phase-shift depends on the mass flow rate, of course, but it is also affected by the magnitude of the added sensor mass and location, and the temperature change; nevertheless, the factors investigated do not induce a zero phase-shift. (c) 2012 Elsevier Ltd. All rights reserved.
引用
收藏
页码:1 / 13
页数:13
相关论文
共 35 条
[1]   Coriolis mass flowmeters: Overview of the current state of the art and latest research [J].
Anklin, Martin ;
Drahm, Wolfgang ;
Rieder, Alfred .
FLOW MEASUREMENT AND INSTRUMENTATION, 2006, 17 (06) :317-323
[2]  
[Anonymous], J FLUIDS STRUCTURES
[3]  
[Anonymous], P 10 INT C COMP STRU
[4]   CORIOLIS FLOWMETERS - INDUSTRIAL PRACTICE AND PUBLISHED INFORMATION [J].
BAKER, RC .
FLOW MEASUREMENT AND INSTRUMENTATION, 1994, 5 (04) :229-246
[5]   The simulation of Coriolis meter response to pulsating flow using a general purpose FE code [J].
Belhadj, A ;
Cheesewright, R ;
Clark, C .
JOURNAL OF FLUIDS AND STRUCTURES, 2000, 14 (05) :613-634
[6]   Coupled finite-volume/finite-element modelling of the straight-tube Coriolis flowmeter [J].
Bobovnik, G ;
Mole, N ;
Kutin, J ;
Stok, B ;
Bajsic, I .
JOURNAL OF FLUIDS AND STRUCTURES, 2005, 20 (06) :785-800
[7]   The response of Coriolis flowmeters to pulsating flows [J].
Cheesewright, R ;
Clark, C ;
Hou, YY .
FLOW MEASUREMENT AND INSTRUMENTATION, 2004, 15 (02) :59-67
[8]   The dynamic response of Coriolis mass flow meters [J].
Cheesewright, R ;
Clark, C ;
Belhadj, A ;
Hou, YY .
JOURNAL OF FLUIDS AND STRUCTURES, 2003, 18 (02) :165-178
[9]   Effect of mechanical vibrations on Coriolis mass flow meters [J].
Cheesewright, R ;
Belhadj, A ;
Clark, C .
JOURNAL OF DYNAMIC SYSTEMS MEASUREMENT AND CONTROL-TRANSACTIONS OF THE ASME, 2003, 125 (01) :103-113
[10]   The effect of flow pulsations on coriolis mass flow meters [J].
Cheesewright, R ;
Clark, C .
JOURNAL OF FLUIDS AND STRUCTURES, 1998, 12 (08) :1025-1039