Measurement of mini-implant stability using resonance frequency analysis

被引:40
作者
Nienkemper, Manuel [1 ]
Wilmes, Benedict [1 ]
Panayotidis, Agamemnon [1 ]
Pauls, Alexander [1 ]
Golubovic, Vladimir [2 ]
Schwarz, Frank [2 ]
Drescher, Dieter [1 ]
机构
[1] Univ Dusseldorf, Dept Orthodont, D-40225 Dusseldorf, Germany
[2] Univ Dusseldorf, Dept Oral Surg & Admittance, D-40225 Dusseldorf, Germany
关键词
Orthodontics; Skeletal anchorage; Implant stability; Resonance frequency analysis; DAMPING CAPACITY ASSESSMENT; ORTHODONTIC ANCHORAGE; IN-VITRO; BONE THICKNESS; MICRO-IMPLANTS; INTERFACE; FAILURE; TISSUE;
D O I
10.2319/043012-354.1
中图分类号
R78 [口腔科学];
学科分类号
1003 ;
摘要
Objective: To investigate whether resonance frequency analysis (RFA) is suitable to measure orthodontic mini-implant stability. Implant size significantly affects the level of resonance frequency. Regarding the operating mode of RFA, it has to be proven whether the resonance frequency of mini-implants in bone fits the range of frequency emitted by the Osstell ISO device. Material and Methods: For this purpose the SmartPegs in the Osstell ISO device were modified to fit with the inner screw thread of orthodontic mini-implants, and 110 mini-implants were inserted into porcine pelvic bone. RFA was performed parallel and perpendicular to the run of superficial bone fibers. A suitability test, Periotest, was also performed in the same directions. Compacta thickness was measured using cone-beam computed tomography. Correlation tests and linear regression analysis were carried out between the three methods. Results: The RFA showed a mean Implant Stability Quotient value of 36.36 +/- 2.67, and the Periotest mean value was -2.10 +/- 1.17. The differences between the two directions of measurement were statistically significant (P > .001) for RFA and the Periotest. There was a high correlation between RFA and the Periotest (r = -0.90) and between RFA and compacta thickness (r = 0.71). The comparison between the Periotest and compacta thickness showed a correlation coefficient of r = -0.64. Conclusion: The present results suggest that RFA is feasible as a measurement method for orthodontic mini-implant stability. As a consequence, it could be used for clinical evaluation of current stability and allow stability-related loading of mini-implants to reduce the failure rate. (Angle Orthod. 2013;83:230-238.)
引用
收藏
页码:230 / 238
页数:9
相关论文
共 38 条
[1]   Mini- and micro-screws for temporary skeletal anchorage in orthodontic therapy; [Mini- und Mikroschrauben zur temporären skelettalen Verankerung in der Kieferorthopädie] [J].
Berens A. ;
Wiechmann D. ;
Dempf R. .
Journal of Orofacial Orthopedics / Fortschritte der Kieferorthopädie, 2006, 67 (6) :450-458
[2]   Correlations between radiographic, clinical and mobility parameters after loading of oral implants with fixed partial dentures - A 2-year longitudinal study [J].
Bragger, U ;
HugelPisoni, C ;
Burgin, W ;
Buser, D ;
Lang, NP .
CLINICAL ORAL IMPLANTS RESEARCH, 1996, 7 (03) :230-239
[3]   Tierexperimentelle Evaluation des periimplantären Knochens bei zylindrischen gegenüber konischen ImplantattypenPeri-implant bone formation around cylindrical and conical implant systems [J].
A. Büchter ;
J. Kleinheinz ;
H.-P. Wiesmann ;
L. Seper ;
U. Joos ;
U. Meyer .
Mund-, Kiefer- und Gesichtschirurgie, 2004, 8 (5) :282-288
[4]   Interface reaction at dental implants inserted in condensed bone [J].
Büchter, A ;
Kleinheinz, J ;
Wiesmann, HP ;
Jayaranan, M ;
Joos, U ;
Meyer, U .
CLINICAL ORAL IMPLANTS RESEARCH, 2005, 16 (05) :509-517
[5]   Biological and biomechanical evaluation of bone remodelling and implant stability after using an osteotome technique [J].
Büchter, A ;
Kleinheinz, J ;
Wiesmann, HP ;
Kersken, J ;
Nienkemper, M ;
von Weyhrother, H ;
Joos, U ;
Meyer, U .
CLINICAL ORAL IMPLANTS RESEARCH, 2005, 16 (01) :1-8
[6]   Load-related bone modelling at the interface of orthodontic micro-implants [J].
Buechter, Andre ;
Wiechmann, Dirk ;
Gaertner, Christoph ;
Hendrik, Marc ;
Vogeler, Martin ;
Wiesmann, Hans Peter ;
Piffko, Josef ;
Meyer, Ulrich .
CLINICAL ORAL IMPLANTS RESEARCH, 2006, 17 (06) :714-722
[7]   The use of microimplants in orthodontic anchorage [J].
Chen, Chung-Ho ;
Chang, Chao-San ;
Hsieh, Chi-Hsin ;
Tseng, Yu-Chuan ;
Shen, Yee-Shyong ;
Huang, I-Yueh ;
Yang, Chia-Fu ;
Chen, Chun-Ming .
JOURNAL OF ORAL AND MAXILLOFACIAL SURGERY, 2006, 64 (08) :1209-1213
[8]  
Costa A, 1998, Int J Adult Orthodon Orthognath Surg, V13, P201
[9]  
Gedrange T., 2006, Folia Morphologica, V65, P70
[10]   A comparative study on the initial stability of different implants placed above the bone level using resonance frequency analysis [J].
Kang, In-Ho ;
Kim, Chang-Whe ;
Lim, Young-Jun ;
Kim, Myung-Joo .
JOURNAL OF ADVANCED PROSTHODONTICS, 2011, 3 (04) :190-195