Measurement of mini-implant stability using resonance frequency analysis

被引:39
|
作者
Nienkemper, Manuel [1 ]
Wilmes, Benedict [1 ]
Panayotidis, Agamemnon [1 ]
Pauls, Alexander [1 ]
Golubovic, Vladimir [2 ]
Schwarz, Frank [2 ]
Drescher, Dieter [1 ]
机构
[1] Univ Dusseldorf, Dept Orthodont, D-40225 Dusseldorf, Germany
[2] Univ Dusseldorf, Dept Oral Surg & Admittance, D-40225 Dusseldorf, Germany
关键词
Orthodontics; Skeletal anchorage; Implant stability; Resonance frequency analysis; DAMPING CAPACITY ASSESSMENT; ORTHODONTIC ANCHORAGE; IN-VITRO; BONE THICKNESS; MICRO-IMPLANTS; INTERFACE; FAILURE; TISSUE;
D O I
10.2319/043012-354.1
中图分类号
R78 [口腔科学];
学科分类号
1003 ;
摘要
Objective: To investigate whether resonance frequency analysis (RFA) is suitable to measure orthodontic mini-implant stability. Implant size significantly affects the level of resonance frequency. Regarding the operating mode of RFA, it has to be proven whether the resonance frequency of mini-implants in bone fits the range of frequency emitted by the Osstell ISO device. Material and Methods: For this purpose the SmartPegs in the Osstell ISO device were modified to fit with the inner screw thread of orthodontic mini-implants, and 110 mini-implants were inserted into porcine pelvic bone. RFA was performed parallel and perpendicular to the run of superficial bone fibers. A suitability test, Periotest, was also performed in the same directions. Compacta thickness was measured using cone-beam computed tomography. Correlation tests and linear regression analysis were carried out between the three methods. Results: The RFA showed a mean Implant Stability Quotient value of 36.36 +/- 2.67, and the Periotest mean value was -2.10 +/- 1.17. The differences between the two directions of measurement were statistically significant (P > .001) for RFA and the Periotest. There was a high correlation between RFA and the Periotest (r = -0.90) and between RFA and compacta thickness (r = 0.71). The comparison between the Periotest and compacta thickness showed a correlation coefficient of r = -0.64. Conclusion: The present results suggest that RFA is feasible as a measurement method for orthodontic mini-implant stability. As a consequence, it could be used for clinical evaluation of current stability and allow stability-related loading of mini-implants to reduce the failure rate. (Angle Orthod. 2013;83:230-238.)
引用
收藏
页码:230 / 238
页数:9
相关论文
共 50 条
  • [1] Orthodontic mini-implant stability at different insertion depths Sensitivity of three stability measurement methods
    Nienkemper, Manuel
    Santel, Natascha
    Hoenscheid, Ralf
    Drescher, Dieter
    JOURNAL OF OROFACIAL ORTHOPEDICS-FORTSCHRITTE DER KIEFERORTHOPADIE, 2016, 77 (04): : 296 - 303
  • [2] Mini-implant stability at the initial healing period A clinical pilot study
    Nienkemper, Manuel
    Wilmes, Benedict
    Pauls, Alexander
    Drescher, Dieter
    ANGLE ORTHODONTIST, 2014, 84 (01) : 127 - 133
  • [3] A novel technique for measurement of orthodontic mini-implant stability using the Osstell ISQ device
    Hosein, Yara K.
    Dixon, S. Jeffrey
    Rizkalla, Amin S.
    Tassi, Ali
    ANGLE ORTHODONTIST, 2019, 89 (02) : 284 - 291
  • [4] Influence of the orientation of the Osstell® transducer during measurement of dental implant stability using resonance frequency analysis: A numerical approach
    Capek, Lukas
    Simunek, Antonin
    Slezak, Radovan
    Dzan, Ladislav
    MEDICAL ENGINEERING & PHYSICS, 2009, 31 (07) : 764 - 769
  • [5] Measurement of Primary Stability of Mini Implants Using Resonance Frequency Analysis
    Singh, Ashish Kumar
    Kannan, Sridhar
    Arora, Nitin
    Bajaj, Yatin
    Revankar, Ameet V.
    APOS TRENDS IN ORTHODONTICS, 2018, 8 (03) : 139 - 145
  • [6] Reliability of the Osstell Implant Stability Quotient and Penguin Resonance Frequency Analysis to Evaluate Implant Stability
    Buyukguclu, Gul
    Ozkurt-Kayahan, Zeynep
    Kazazoglu, Ender
    IMPLANT DENTISTRY, 2018, 27 (04) : 429 - 433
  • [7] Resonance frequency analysis of miniscrew implant stability
    Tseng, Yu-Chuan
    Pan, Chin-Yun
    Liu, Pao-Hsin
    Yang, Yi-Hsin
    Chang, Hong-Po
    Chen, Chun-Ming
    JOURNAL OF ORAL SCIENCE, 2018, 60 (01) : 64 - 69
  • [8] In vitro comparative analysis of two resonance frequency measurement devices: Osstell implant stability coefficient and Penguin resonance frequency analysis
    Herrero-Climent, Mariano
    Falcao, Artur
    Lopez-Jarana, Paula
    Diaz-Castro, Carmen M.
    Rios-Carrasco, Blanca
    Vicente Rios-Santos, Jose
    CLINICAL IMPLANT DENTISTRY AND RELATED RESEARCH, 2019, 21 (06) : 1124 - 1131
  • [9] Measurement of dental implant stability by resonance frequency analysis: A review of the literature
    Pilar Quesada-Garcia, Ma
    Prados-Sanchez, Estrella
    Victoria Olmedo-Gaya, Ma
    Munoz-Soto, Esther
    Paloma Gonzalez-Rodriguez, Ma
    Vallecillo-Capilla, Manuel
    MEDICINA ORAL PATOLOGIA ORAL Y CIRUGIA BUCAL, 2009, 14 (10): : E538 - E546
  • [10] Mechanical influence of thread pitch on orthodontic mini-implant stability
    da Cunha, Amanda Carneiro
    Ayres de Freitas, Amanda Osorio
    Marquezan, Mariana
    Nojima, Lincoln Issamu
    BRAZILIAN ORAL RESEARCH, 2015, 29 (01): : 1 - 6