A simple dilute-and-shoot procedure for Si determination in diesel and biodiesel by microwave-induced plasma optical emission spectrometry

被引:61
作者
Amais, Renata S. [1 ]
Donati, George L. [1 ]
Schiavo, Daniela [2 ]
Nobrega, Joaquim A. [1 ]
机构
[1] Univ Fed Sao Carlos, Dept Chem, Grp Appl Instrumental Anal, BR-13560 Sao Carlos, SP, Brazil
[2] Agilent Technol, Sao Paulo, Brazil
基金
巴西圣保罗研究基金会;
关键词
Silicon determination; Liquid fuel; MIP OES; Ethanol dilution; SAMPLE PREPARATION; METALS; OIL; NA; CA; MG;
D O I
10.1016/j.microc.2012.09.001
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
In the present work, microwave-induced plasma optical emission spectrometry (MIP OES) and the flow blurring nebulizer (FBN) technology are used to determine silicon in diesel and biodiesel samples. A simple dilute-and-shoot procedure with ethanol is presented. Two additional sample preparation procedures are also evaluated for comparison: closed-vessel microwave-assisted acid digestion and microemulsion preparation in n-propanol. Limits of detection (LOD) vary from 5 to 20 mu g L-1 and relative standard deviations (RSD) are lower than 2% in all cases. Accuracy is evaluated by spike experiments, with recoveries between 80 and 103%. The influence of Ca, K, Mg and Na as concomitant ions on silicon determination by MIP OES is also investigated. Significant effects on Si emission signals are observed neither in single nor multiple concomitant experiments. Although good results are obtained with all sample preparation procedures evaluated, sample dilution in ethanol represents a simpler and faster strategy, which allows external calibration with inorganic standards and aqueous solutions. (C) 2012 Elsevier B.V. All rights reserved.
引用
收藏
页码:318 / 322
页数:5
相关论文
共 25 条
[21]   Effect of the silicon chemical form on the emission intensity in inductively coupled plasma atomic emission spectrometry for xylene matrices [J].
Sanchez, Raquel ;
Todoli, Jose-Luis ;
Lienemann, Charles-Philippe ;
Mermet, Jean-Michel .
JOURNAL OF ANALYTICAL ATOMIC SPECTROMETRY, 2009, 24 (04) :391-401
[22]   Sample preparation in the determination of metals in oil and petroleum products by ICP MS [J].
Soin, A. V. ;
Maryutina, T. A. ;
Arbuzova, T. V. ;
Spivakov, B. Ya. .
JOURNAL OF ANALYTICAL CHEMISTRY, 2010, 65 (06) :571-576
[23]  
Thomsen V, 2000, SPECTROSCOPY, V15, P33
[24]  
Wu C., 2012, 59909791EN AG TECHN
[25]   Determination of sulfur in biodiesel microemulsions using the summation of the intensities of multiple emission lines [J].
Young, Carl G. ;
Amais, Renata S. ;
Schiavo, Daniela ;
Garcia, Edivaldo E. ;
Nobrega, Joaquim A. ;
Jones, Bradley T. .
TALANTA, 2011, 84 (03) :995-999