Temperature variations as a source of uncertainty in medical fiber-coupled organic plastic scintillator dosimetry

被引:51
作者
Buranurak, S. [1 ]
Andersen, C. E. [1 ]
Beierholm, A. R. [1 ]
Lindvold, L. R. [1 ]
机构
[1] Tech Univ Denmark, Ctr Nucl Technol, DK-4000 Roskilde, Denmark
关键词
Dosimetry; Scintillator; Optical fiber; Uncertainty; HDR BRACHYTHERAPY; DETECTORS; SYSTEM;
D O I
10.1016/j.radmeas.2013.01.049
中图分类号
TL [原子能技术]; O571 [原子核物理学];
学科分类号
0827 ; 082701 ;
摘要
Fiber-coupled organic plastic scintillators have potential applications in medical dosimetry related to, for example, brachytherapy and external beam radiotherapy with MV photons. As medical dosimetry generally strives for high accuracy, we designed a study to assess if the light yield from commonly used scintillating fibers would change with temperature in the clinical range (15-40 degrees C). The study showed that the light yield in the peak regions of the scintillators studied decreases linearly with increasing temperature. For the blue BCF-12 and the green BCF-60 from Saint-Gobain, France we found temperature coefficients of -0.15 +/- 0.01%/K and -0.55 +/- 0.04%/K, respectively. These values are sufficiently large to warrant careful consideration for clinical measurements. (c) 2013 Elsevier Ltd. All rights reserved.
引用
收藏
页码:307 / 311
页数:5
相关论文
共 10 条
[1]   Fiber-coupled Luminescence Dosimetry in Therapeutic and Diagnostic Radiology [J].
Andersen, Claus E. .
CONCEPTS AND TRENDS IN MEDICAL RADIATION DOSIMETRY, 2011, 1345 :100-119
[2]   Precision of RL/OSL medical dosimetry with fiber-coupled Al2O3:C: Influence of readout delay and temperature variations [J].
Andersen, Claus E. ;
Edmund, Jens M. ;
Damkjaer, Sidsel M. S. .
RADIATION MEASUREMENTS, 2010, 45 (3-6) :653-657
[3]   WATER-EQUIVALENT PLASTIC SCINTILLATION DETECTORS FOR HIGH-ENERGY BEAM DOSIMETRY .1. PHYSICAL CHARACTERISTICS AND THEORETICAL CONSIDERATIONS [J].
BEDDAR, AS ;
MACKIE, TR ;
ATTIX, FH .
PHYSICS IN MEDICINE AND BIOLOGY, 1992, 37 (10) :1883-1900
[4]   On possible temperature dependence of plastic scintillator response [J].
Beddar, Sam .
MEDICAL PHYSICS, 2012, 39 (10) :6522-6522
[5]   Characterizing a pulse-resolved dosimetry system for complex radiotherapy beams using organic scintillators [J].
Beierholm, Anders R. ;
Ottosson, Rickard O. ;
Lindvold, Lars R. ;
Behrens, Claus F. ;
Andersen, Claus E. .
PHYSICS IN MEDICINE AND BIOLOGY, 2011, 56 (10) :3033-3045
[6]   Dose mapping of the rectal wall during brachytherapy with an array of scintillation dosimeters [J].
Cartwright, L. E. ;
Suchowerska, N. ;
Yin, Y. ;
Lambert, J. ;
Haque, M. ;
McKenzie, D. R. .
MEDICAL PHYSICS, 2010, 37 (05) :2247-2255
[7]   Extraction of depth-dependent perturbation factors for parallel-plate chambers in electron beams using a plastic scintillation detector [J].
Lacroix, Frederic ;
Guillot, Mathieu ;
McEwen, Malcolm ;
Cojocaru, Claudiu ;
Gingras, Luc ;
Beddar, A. Sam ;
Beaulieu, Luc .
MEDICAL PHYSICS, 2010, 37 (08) :4331-4342
[8]   In vivo dosimeters for HDR brachytherapy:: A comparison of a diamond detector, MOSFET, TLD, and scintillation detector [J].
Lambert, Jamil ;
Nakano, Tatsuya ;
Law, Sue ;
Elsey, Justin ;
McKenzie, David R. ;
Suchowerska, Natalka .
MEDICAL PHYSICS, 2007, 34 (05) :1759-1765
[9]   Small field diode correction factors derived using an air core fibre optic scintillation dosimeter and EBT2 film [J].
Ralston, Anna ;
Liu, Paul ;
Warrener, Kirbie ;
McKenzie, David ;
Suchowerska, Natalka .
PHYSICS IN MEDICINE AND BIOLOGY, 2012, 57 (09) :2587-2602
[10]   A phantom study of an in vivo dosimetry system using plastic scintillation detectors for real-time verification of 192Ir HDR brachytherapy [J].
Therriault-Proulx, Francois ;
Briere, Tina M. ;
Mourtada, Firas ;
Aubin, Sylviane ;
Beddar, Sam ;
Beaulieu, Luc .
MEDICAL PHYSICS, 2011, 38 (05) :2542-2551