A Novel Method for Implementation of Frameless StereoEEG in Epilepsy Surgery

被引:64
作者
Nowell, Mark [1 ,2 ,3 ]
Rodionov, Roman [1 ,2 ]
Diehl, Beate [1 ,2 ]
Wehner, Tim [1 ,2 ]
Zombori, Gergely [4 ]
Kinghorn, Jane [1 ]
Ourselin, Sebastien [4 ]
Duncan, John [1 ,2 ]
Miserocchi, Anna [1 ,3 ]
McEvoy, Andrew [1 ,3 ]
机构
[1] UCL Inst Neurol, Dept Clin & Expt Epilepsy, London, England
[2] Epilepsy Soc, MRI Unit, Gerrards Cross, England
[3] Natl Hosp Neurol & Neurosurg, Dept Neurosurg, London WC1N 3BG, England
[4] UCL, Ctr Med Imaging & Comp, London, England
关键词
Epilepsy; Frameless stereotaxy; StereoEEG;
D O I
10.1227/NEU.0000000000000544
中图分类号
R74 [神经病学与精神病学];
学科分类号
摘要
BACKGROUND: Stereoelectroencephalography (SEEG) is an invasive diagnostic procedure in epilepsy surgery that is usually implemented with frame-based methods. OBJECTIVE: To describe a new technique of frameless SEEG and report a prospective case series at a single center. METHODS: Image integration and planning of electrode trajectories were performed preoperatively on specialized software and exported to a Medtronic S7 StealthStation. Trajectories were implemented by frameless stereotaxy using percutaneous drilling and bolt insertion. RESULTS: Twenty-two patients went this technique, with the insertion of 187 intracerebral electrodes. Of 187 electrodes, 175 accurately reached their neurophysiological target, as measured by postoperative computed tomography reconstruction and multimodal image integration with preoperative magnetic resonance imaging. Four electrodes failed to hit their target due to extradural deflection, and 3 were subsequently resited satisfactorily. Eight electrodes were off target by a mean of 3.6 mm (range, 0.9-6.8 mm) due to a combination of errors in bolt trajectory implementation and bending of the electrode. There was 1 postoperative hemorrhage that was clinically asymptomatic and no postoperative infections. Sixteen patients were offered definitive cortical resections, and 6 patients were excluded from resective surgery. CONCLUSION: Frameless SEEG is a novel and safe method for implementing SEEG and is easily translated into clinical practice.
引用
收藏
页码:525 / 533
页数:9
相关论文
共 11 条
[1]   Use of robot-guided stereotactic placement of intracerebral electrodes for investigation of focal epilepsy: initial experience in the UK [J].
Abhinav, Kumar ;
Prakash, Savithru ;
Sandeman, David R. .
BRITISH JOURNAL OF NEUROSURGERY, 2013, 27 (05) :704-705
[2]   Stereoelectroencephalography: Surgical Methodology, Safety, and Stereotactic Application Accuracy in 500 Procedures [J].
Cardinale, Francesco ;
Cossu, Massimo ;
Castana, Laura ;
Casaceli, Giuseppe ;
Schiariti, Marco Paolo ;
Miserocchi, Anna ;
Fuschillo, Dalila ;
Moscato, Alessio ;
Caborni, Chiara ;
Arnulfo, Gabriele ;
Lo Russo, Giorgio .
NEUROSURGERY, 2013, 72 (03) :353-366
[3]   Stereotactic placement of depth electrodes in medically intractable epilepsy [J].
Gonzalez-Martinez, Jorge ;
Mullin, Jeffrey ;
Vadera, Sumeet ;
Bulacio, Juan ;
Hughes, Gwyneth ;
Jones, Stephen ;
Enatsu, Re ;
Najm, Imad .
JOURNAL OF NEUROSURGERY, 2014, 120 (03) :639-644
[4]   Classification and Analysis of the Errors in Neuronavigation [J].
Man Ning Wang ;
Zhi Jian Song .
NEUROSURGERY, 2011, 68 (04) :1131-1143
[5]   Frameless stereotactic placement of depth electrodes in epilepsy surgery [J].
Mehta, AD ;
Labar, D ;
Dean, A ;
Harden, C ;
Hosain, S ;
Pak, J ;
Marks, D ;
Schwartz, TH .
JOURNAL OF NEUROSURGERY, 2005, 102 (06) :1040-1045
[6]  
Pfisterer WK, 2008, NEUROSURGERY, V63, P201, DOI [10.1227/01.NEU.0000297007.04975.C6, 10.1227/01.neu.0000317394.14303.99]
[7]   The advantages of frameless stereotactic biopsy over frame-based biopsy [J].
Price, R .
BRITISH JOURNAL OF NEUROSURGERY, 2003, 17 (01) :90-91
[8]   Feasibility of multimodal 3D neuroimaging to guide implantation of intracranial EEG electrodes [J].
Rodionov, Roman ;
Vollmar, Christian ;
Nowell, Mark ;
Miserocchi, Anna ;
Wehner, Tim ;
Micallef, Caroline ;
Zombori, Gergely ;
Ourselin, Sebastien ;
Diehl, Beate ;
McEvoy, Andrew W. ;
Duncan, John S. .
EPILEPSY RESEARCH, 2013, 107 (1-2) :91-100
[9]   Localization and registration accuracy in image guided neurosurgery: a clinical study [J].
Shamir, Reuben R. ;
Joskowicz, Leo ;
Spektor, Sergey ;
Shoshan, Yigal .
INTERNATIONAL JOURNAL OF COMPUTER ASSISTED RADIOLOGY AND SURGERY, 2009, 4 (01) :45-52
[10]  
Szikla G., 1977, ANGIOGRAPHY HUMAN BR