Quality assurance/quality control of foot and mouth disease solid phase competition enzyme-linked immunosorbent assay - Part II. Quality control: comparison of two charting methods to monitor assay performance

被引:14
作者
Goris, N [1 ]
De Clercq, K [1 ]
机构
[1] Vet & Agrochem Res Ctr, Dept Virol, B-1180 Brussels, Belgium
来源
REVUE SCIENTIFIQUE ET TECHNIQUE-OFFICE INTERNATIONAL DES EPIZOOTIES | 2005年 / 24卷 / 03期
关键词
accreditation; control chart; enzyme-linked immunosorbent assay; exponentially weighted moving average; foot and mouth disease; quality control; Shewhart; validation; westgard rule;
D O I
10.20506/rst.24.3.1630
中图分类号
S85 [动物医学(兽医学)];
学科分类号
0906 ;
摘要
Diagnostic laboratories are increasingly required to meet stringent quality standards, and validated assays are needed to achieve formal accreditation. Validation of test methods is often considered to be finalised when the assay parameters and characteristics have been established. However, like any process, diagnostic assays are subject to random variation resulting in shifts in the mean test values. Continuous monitoring of assays using control charts will alert the interpreter of changes in performance. For this purpose, several charting methods have been developed and implemented. This paper compares the Shewhart and the exponentially weighted moving average (EWMA) control charts with respect to the day to day monitoring of internal quality control samples for the foot and mouth disease solid phase competition enzyme-linked immunosorbent assay. Both chart types are equally sensitive to shifts, but the EWMA method seems to provide the best balance between false rejection and false acceptance.
引用
收藏
页码:1005 / 1016
页数:12
相关论文
共 33 条