How the measurement depth influences the liver stiffness assessment using real-time shear wave elastography (SWE)

被引:8
作者
Wang, Cong-zhi [1 ]
Xiao, Yang [1 ]
Song, Dan [1 ]
Ming, Yan [1 ]
Zheng, Hai-rong [1 ]
Zheng, Jian [2 ]
Zeng, Jie [2 ]
Huang, Ze-ping [2 ]
Zheng, Rong-qin [2 ]
机构
[1] Chinese Acad Sci, Inst Biomed & Hlth Engn, Shenzhen Inst Adv Technol, Paul C Lauterbur Res Ctr Biomed Imaging, Shenzhen, Peoples R China
[2] Sun Yat Sen Univ, Third Affiliated Hosp, Dept Ultrasound, Guangzhou, Guangdong, Peoples R China
来源
2013 IEEE INTERNATIONAL ULTRASONICS SYMPOSIUM (IUS) | 2013年
基金
美国国家科学基金会;
关键词
Measurement depth; Real-time shear wave elastography; Liver stiffness; Young's modulus; Liver capsule; ACOUSTIC RADIATION FORCE; TOUCH TISSUE QUANTIFICATION; RELIABLE MEASUREMENT; FIBROSIS;
D O I
10.1109/ULTSYM.2013.0004
中图分类号
O42 [声学];
学科分类号
070206 ; 082403 ;
摘要
The purpose of this study was to evaluate the influence of measurement depth on the results of real-time shear wave elastography (SWE) measurements and to find a reliable measurement depth range for liver stiffness assessment using SWE. SWE measurements were performed on an Elasticity Quality Assurance phantom and healthy volunteers' livers. The different depths from probe surface and from liver capsule were studied in 30 and 88 healthy volunteers, separately. The success rate of measurements (SRoM), success rate of subjects (SRoS), and coefficients of variation (CV) of repeated measurements were calculated and compared. In phantom study, the CV at 2-6cm depth (from 0.6 % to 0.9 %) was larger than those at the other depths (from 2.8 % to 9.8 %). For the first group of healthy volunteers, the SRoMs at 3-7cm depths were significantly higher than the others, and the measured results were found to be more reliable at 2-5cm (P<0.001). In the second group of subjects, the SRoSs at 1-4cm depths below liver capsule were all 100 % but at 0-1 cm depth was zero. According to our results, the most reliable depth range of liver stiffness assessment using SWE should be 3-5cm from the probe surface and simultaneously 1-2cm below the liver capsule.
引用
收藏
页码:13 / 16
页数:4
相关论文
共 12 条
[1]   Non-invasive evaluation of liver fibrosis using transient elastography [J].
Castera, Laurent ;
Forns, Xavier ;
Alberti, Alfredo .
JOURNAL OF HEPATOLOGY, 2008, 48 (05) :835-847
[2]   Reproducibility of real-time shear wave elastography in the evaluation of liver elasticity [J].
Ferraioli, Giovanna ;
Tinelli, Carmine ;
Zicchetti, Mabel ;
Above, Elisabetta ;
Poma, Gianluigi ;
Di Gregorio, Marta ;
Filice, Carlo .
EUROPEAN JOURNAL OF RADIOLOGY, 2012, 81 (11) :3102-3106
[3]   Prevalence and factors associated with failure of liver stiffness measurement using FibroScan in a prospective study of 2114 examinations [J].
Foucher, Juliette ;
Castera, Laurent ;
Bernard, Pierre-Henri ;
Adhoute, Xavier ;
Laharie, David ;
Bertet, Julien ;
Couzigou, Patrice ;
de Ledinghen, Victor .
EUROPEAN JOURNAL OF GASTROENTEROLOGY & HEPATOLOGY, 2006, 18 (04) :411-412
[4]  
Garra Brian Stephen, 2007, Ultrasound Q, V23, P255, DOI 10.1097/ruq.0b013e31815b7ed6
[5]   Reliable Measurement Procedure of Virtual Touch Tissue Quantification With Acoustic Radiation Force Impulse Imaging [J].
Kaminuma, Chie ;
Tsushima, Yoshito ;
Matsumoto, Noriko ;
Kurabayashi, Takemi ;
Taketomi-Takahashi, Ayako ;
Endo, Keigo .
JOURNAL OF ULTRASOUND IN MEDICINE, 2011, 30 (06) :745-751
[6]   Diagnosis and quantitation of fibrosis [J].
Manning, Diarmuid S. ;
Afdhal, Nezam H. .
GASTROENTEROLOGY, 2008, 134 (06) :1670-1681
[7]   QUANTITATIVE VISCOELASTICITY MAPPING OF HUMAN LIVER USING SUPERSONIC SHEAR IMAGING: PRELIMINARY IN VIVO FEASABILITY STUDY [J].
Muller, Marie ;
Gennisson, Jean-Luc ;
Deffieux, Thomas ;
Tanter, Mickael ;
Fink, Mathias .
ULTRASOUND IN MEDICINE AND BIOLOGY, 2009, 35 (02) :219-229
[8]   Quantifying hepatic shear modulus in vivo using acoustic radiation force [J].
Palmeri, M. L. ;
Wang, M. H. ;
Dahl, J. J. ;
Frinkley, K. D. ;
Nightingale, K. R. .
ULTRASOUND IN MEDICINE AND BIOLOGY, 2008, 34 (04) :546-558
[9]   Liver fibrosis evaluation using real-time shear wave elastography: Applicability and diagnostic performance using methods without a gold standard [J].
Poynard, Thierry ;
Munteanu, Mona ;
Luckina, Elena ;
Perazzo, Hugo ;
Ngo, Yen ;
Royer, Luca ;
Fedchuk, Larysa ;
Sattonnet, Florence ;
Pais, Raluca ;
Lebray, Pascal ;
Rudler, Marika ;
Thabut, Dominique ;
Ratziu, Vlad .
JOURNAL OF HEPATOLOGY, 2013, 58 (05) :928-935
[10]  
Sporea I, 2011, Ultraschall Med, V32 Suppl 1, pS46, DOI 10.1055/s-0029-1245360