Comparison of activated clotting times measured using the Hemochron Jr. Signature and Medtronic ACT Plus during cardiopulmonary bypass with acute normovolemic haemodilution

被引:15
作者
Lee, Jung Min [1 ]
Park, Eun Young [1 ]
Kim, Kyung Mi [1 ]
Won, Jong Chan [1 ]
Jung, Tack Koon [2 ]
Lee, Soo Kyung [1 ]
机构
[1] Hallym Univ, Coll Med, Sacred Heart Hosp, Dept Anaesthesiol & Pain Med, Anyang, South Korea
[2] CHA Univ, Dept Cardiovasc Surg, Bundang CHA Med Ctr, CHA Bundang Med Ctr, Seongnam, South Korea
关键词
Activated clotting time; cardiopulmonary bypass; haemodilution; kaolin; heparin; cardiac surgery; COAGULATION; HEPARIN; THERAPY;
D O I
10.1177/0300060517731952
中图分类号
R-3 [医学研究方法]; R3 [基础医学];
学科分类号
1001 ;
摘要
Objective This study compared the activated clotting time (ACT) measured using the Hemochron Jr. Signature (HACT) with the ACT measured using the Medtronic ACT Plus (MACT) during cardiopulmonary bypass (CPB) with acute normovolemic haemodilution (ANH) in patients undergoing cardiac surgery. Methods The ACT was checked at baseline with both devices after inducing anaesthesia, and 400 to 800 mL of whole blood was withdrawn to induce moderate ANH. Before initiating CPB, a 300-IU/kg bolus dose of heparin was administered to maintain the HACT at >400 s; protamine was later given to reverse the anticoagulation. The ACT was checked using both devices at baseline, during heparinisation, and after protamine administration. Results In total, 106 pairs of samples from 29 patients were analysed. The ACT showed a good correlation between the two devices (r=0.956). However, Bland-Altman analysis showed that the MACT was higher, particularly at baseline and during heparinisation. Multiple regression analysis showed that the blood glucose concentration significantly influenced the differences between the two ACT devices. Conclusions The HACT was lower than the MACT during CPB with ANH in patients undergoing cardiac surgery. Clinicians should be cautious when using each ACT device within generally accepted reference ACT values.
引用
收藏
页码:873 / 882
页数:10
相关论文
共 17 条
[1]   COMPARISON OF HEMOCHRON AND HEMOTEC ACTIVATED COAGULATION TIME TARGET VALUES DURING PERCUTANEOUS TRANSLUMINAL CORONARY ANGIOPLASTY [J].
AVENDANO, A ;
FERGUSON, JJ .
JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 1994, 23 (04) :907-910
[2]   THE USE OF ACTIVATED CLOTTING TIMES TO MONITOR HEPARIN-THERAPY DURING AND AFTER INTERVENTIONAL PROCEDURES [J].
BOWERS, J ;
FERGUSON, JJ .
CLINICAL CARDIOLOGY, 1994, 17 (07) :357-361
[3]  
Brouwer Monique E, 2012, J Extra Corpor Technol, V44, P194
[4]   Comparison of activated clotting times obtained using Hemochron and Medtronic analysers in patients receiving anti-thrombin therapy during cardiac catheterisation [J].
Chia, Stanley ;
Van Cott, Elizabeth M. ;
Raffell, O. Christopher ;
Jang, Ik-Kyung .
THROMBOSIS AND HAEMOSTASIS, 2009, 101 (03) :535-540
[5]  
Dietrich W, 2001, ANESTH ANALG, V92, P66
[6]   Measurement of activated clotting time in children -: comparison of the Celite i-STAT® ACT with the Medtronic ACT® II [J].
Hug, MI ;
Di Bernardo, S ;
Berger, F ;
Bauersfeld, U ;
Weiss, M .
ACTA ANAESTHESIOLOGICA SCANDINAVICA, 2004, 48 (02) :211-217
[7]   Point-of-Care Testing for Anticoagulation Monitoring in Neuroendovascular Procedures [J].
Hussein, H. M. ;
Georgiadis, A. L. ;
Qureshi, A. I. .
AMERICAN JOURNAL OF NEURORADIOLOGY, 2012, 33 (07) :1211-1220
[8]  
Kaplan JA., 2008, ESSENTIALS CARDIAC A, P267
[9]   Acute normovolemic hemodilution does not reduce antithrombin concentration for cardiac surgery [J].
Linden, MD ;
Schneider, M ;
Erber, WN .
JOURNAL OF THROMBOSIS AND THROMBOLYSIS, 2004, 17 (03) :173-176
[10]   In vivo effect of haemodilution with saline on coagulation:: a randomized controlled trial [J].
Ng, KFJ ;
Lam, CCK ;
Chan, LC .
BRITISH JOURNAL OF ANAESTHESIA, 2002, 88 (04) :475-480