An Analog Integrated Circuit Beamformer for High-Frequency Medical Ultrasound Imaging

被引:38
作者
Gurun, Gokce [1 ]
Zahorian, Jaime S. [1 ]
Sisman, Alper [2 ]
Karaman, Mustafa [3 ]
Hasler, Paul E. [1 ]
Degertekin, F. Levent [1 ]
机构
[1] Georgia Inst Technol, Dept Elect & Comp Engn, Atlanta, GA 30332 USA
[2] Marmara Univ, Dept Elect & Elect Engn, TR-34722 Istanbul, Turkey
[3] Isik Univ, Dept Elect Elect Engn, TR-34980 Istanbul, Turkey
关键词
All-pass filter; analog delay; beam forming; biquad current mirror; delay line; ultrasound imaging; PRML READ/WRITE CHANNEL; MATRIX TRANSDUCER; CMUT ARRAYS; LOW-POWER; DESIGN; CHIP;
D O I
10.1109/TBCAS.2012.2219532
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
We designed and fabricated a dynamic receive beamformer integrated circuit (IC) in 0.35-mu m CMOS technology. This beamformer IC is suitable for integration with an annular array transducer for high-frequency (30-50 MHz) intravascular ultrasound (IVUS) imaging. The beamformer IC consists of receive preamplifiers, an analog dynamic delay-and-sum beamformer, and buffers for 8 receive channels. To form an analog dynamic delay line we designed an analog delay cell based on the current-mode first-order all-pass filter topology, as the basic building block. To increase the bandwidth of the delay cell, we explored an enhancement technique on the current mirrors. This technique improved the overall bandwidth of the delay line by a factor of 6. Each delay cell consumes 2.1-mW of power and is capable of generating a tunable time delay between 1.75 ns to 2.5 ns. We successfully integrated the fabricated beamformer IC with an 8-element annular array. Experimental test results demonstrated the desired buffering, preamplification and delaying capabilities of the beamformer.
引用
收藏
页码:454 / 467
页数:14
相关论文
共 55 条
[11]  
Chiang A. M., 1998, INTEGRATED BEAM FORM
[12]   A 6-b 1.3-Gsample/s A/D converter in 0.35-μm CMOS [J].
Choi, M ;
Abidi, AA .
IEEE JOURNAL OF SOLID-STATE CIRCUITS, 2001, 36 (12) :1847-1858
[13]   A matrix transducer design with improved image quality and acquisition rate [J].
Daft, Chris ;
Brueske, Dan ;
Wagner, Paul ;
Liu, Donald .
2007 IEEE ULTRASONICS SYMPOSIUM PROCEEDINGS, VOLS 1-6, 2007, :411-+
[14]  
de Ana FJ, 2002, ULTRASON, P1773
[15]   Advances in ultrasound biomicroscopy [J].
Foster, FS ;
Pavlin, CJ ;
Harasiewicz, KA ;
Christopher, DA ;
Turnbull, DH .
ULTRASOUND IN MEDICINE AND BIOLOGY, 2000, 26 (01) :1-27
[16]   Imaging of coronary atherosclerosis: intravascular ultrasound [J].
Garcia-Garcia, Hector M. ;
Costa, Marco A. ;
Serruys, Patrick W. .
EUROPEAN HEART JOURNAL, 2010, 31 (20) :2456-2469C
[17]  
Gurun Gokce, 2009, 2009 IEEE International Ultrasonics Symposium, P345, DOI 10.1109/ULTSYM.2009.5441823
[18]  
Gurun G, 2011, BIOMED CIRC SYST C, P65, DOI 10.1109/BioCAS.2011.6107728
[19]   A log-domain μbeamformer for medical ultrasound imaging systems [J].
Halvorsrod, T ;
Luzi, W ;
Lande, TS .
IEEE TRANSACTIONS ON CIRCUITS AND SYSTEMS I-REGULAR PAPERS, 2005, 52 (12) :2563-2575
[20]  
Halvorsrod T, 2005, ULTRASON, P500