An Analog Integrated Circuit Beamformer for High-Frequency Medical Ultrasound Imaging

被引:38
作者
Gurun, Gokce [1 ]
Zahorian, Jaime S. [1 ]
Sisman, Alper [2 ]
Karaman, Mustafa [3 ]
Hasler, Paul E. [1 ]
Degertekin, F. Levent [1 ]
机构
[1] Georgia Inst Technol, Dept Elect & Comp Engn, Atlanta, GA 30332 USA
[2] Marmara Univ, Dept Elect & Elect Engn, TR-34722 Istanbul, Turkey
[3] Isik Univ, Dept Elect Elect Engn, TR-34980 Istanbul, Turkey
关键词
All-pass filter; analog delay; beam forming; biquad current mirror; delay line; ultrasound imaging; PRML READ/WRITE CHANNEL; MATRIX TRANSDUCER; CMUT ARRAYS; LOW-POWER; DESIGN; CHIP;
D O I
10.1109/TBCAS.2012.2219532
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
We designed and fabricated a dynamic receive beamformer integrated circuit (IC) in 0.35-mu m CMOS technology. This beamformer IC is suitable for integration with an annular array transducer for high-frequency (30-50 MHz) intravascular ultrasound (IVUS) imaging. The beamformer IC consists of receive preamplifiers, an analog dynamic delay-and-sum beamformer, and buffers for 8 receive channels. To form an analog dynamic delay line we designed an analog delay cell based on the current-mode first-order all-pass filter topology, as the basic building block. To increase the bandwidth of the delay cell, we explored an enhancement technique on the current mirrors. This technique improved the overall bandwidth of the delay line by a factor of 6. Each delay cell consumes 2.1-mW of power and is capable of generating a tunable time delay between 1.75 ns to 2.5 ns. We successfully integrated the fabricated beamformer IC with an 8-element annular array. Experimental test results demonstrated the desired buffering, preamplification and delaying capabilities of the beamformer.
引用
收藏
页码:454 / 467
页数:14
相关论文
共 55 条
[1]   A 200-MSample/s trellis-coded PRML read/write channel with analog adaptive equalizer and digital servo [J].
Alini, R ;
Betti, G ;
Bietti, I ;
Bollati, G ;
Brianti, F ;
Dati, A ;
Demicheli, M ;
Gadducci, P ;
Marchese, S ;
Marconetti, E ;
Pisati, V ;
Zuffada, M ;
Castello, R ;
Heydati, F ;
Gillen, P ;
Maguire, G ;
Marrow, M ;
McDonagh, S ;
OBrien, F ;
OBrien, J ;
OhEarcain, N ;
Reddy, D ;
Fredrickson, L ;
Stone, D ;
Volz, LR .
IEEE JOURNAL OF SOLID-STATE CIRCUITS, 1997, 32 (11) :1824-1838
[2]  
[Anonymous], 2014, Diagnostic Ultrasound Imaging
[3]   A Floating-Gate-Based Field-Programmable Analog Array [J].
Basu, Arindam ;
Brink, Stephen ;
Schlottmann, Craig ;
Ramakrishnan, Shubha ;
Petre, Csaba ;
Koziol, Scott ;
Baskaya, Faik ;
Twigg, Christopher M. ;
Hasler, Paul .
IEEE JOURNAL OF SOLID-STATE CIRCUITS, 2010, 45 (09) :1781-1794
[4]   CMOS CHIP FOR INVASIVE ULTRASOUND IMAGING [J].
BLACK, WC ;
STEPHENS, DN .
IEEE JOURNAL OF SOLID-STATE CIRCUITS, 1994, 29 (11) :1381-1387
[5]   A digital beamformer for high-frequency annular arrays [J].
Brown, JA ;
Lockwood, GR .
IEEE TRANSACTIONS ON ULTRASONICS FERROELECTRICS AND FREQUENCY CONTROL, 2005, 52 (08) :1262-1269
[6]   Design and fabrication of annular arrays for high-frequency ultrasound [J].
Brown, JA ;
Démoré, CEM ;
Lockwood, GR .
IEEE TRANSACTIONS ON ULTRASONICS FERROELECTRICS AND FREQUENCY CONTROL, 2004, 51 (08) :1010-1017
[7]  
Brunner E., 2002, Analog Devices
[8]   A CMOS ANALOG CONTINUOUS-TIME DELAY-LINE WITH ADAPTIVE DELAY-TIME CONTROL [J].
BULT, K ;
WALLINGA, H .
IEEE JOURNAL OF SOLID-STATE CIRCUITS, 1988, 23 (03) :759-766
[9]  
Burlingame E, 2000, P 26 EUROPEAN SOLID, P288
[10]  
Chang YW, 2010, ASIA PACIF MICROWAVE, P426